Akademska digitalna zbirka SLovenije - logo
ALL libraries (COBIB.SI union bibliographic/catalogue database)
  • The effects of eluent mixing on TLS detection in gradient elution HPLC
    Kožar Logar, Jasmina ...
    The effects of changing solvent composition on the LOD of TLS detection in gradient elution HPLC have been studied from the perspective of thermo-optical properties of the solvent. Hyphenated ... gradient high-performance liquid chromatography (HPLC)-thermal lens spectrometry (TLS), was used to separate and detect 13 carotenoid compounds and two chlorophylls. Utilization of mixing coils into the system reduces the inhomogeneities during eluent changes and therefore enables the application of thermal lens detection in the gradient HPLC method. For gradient chromatographic conditions in which the thermo-optical properties and related enhancement factor change as much as 50% over 10 min, the LODs for the TLS detector were enhanced by as much as three times in comparison with UV-Vis detection. For the isocratic part of the chromatogram, up to a tenfold improvement of LODs was achieved with TLS detection.
    Source: Analytical and bioanalytical chemistry. - ISSN 1618-2642 (Vol. 374, no. 2, 2002, str. 323-328)
    Type of material - article, component part
    Publish date - 2002
    Language - english
    COBISS.SI-ID - 227067