  • Carbon-fiber microsensor for in vivo monitoring of trace zinc(II) based on electrochemical stripping analysis
    Wang, Joseph ...
    The optimization and characterization of a microsensor for trace zinc based on stripping voltammetry at a single carbon-fiber microelectrode are reported. Despite the absence of a mercury coverage, ... the carbon-fiber stripping sensor displays a significant hydrogen overvoltage in the physiological pH and a well-defined zinc stripping peak (at ca. -1.17 V). 5 s deposition periods allow convenient quantitation of zinc over the 100-1000 ng/ml (ppb) range. Detection limits of 0.8 ng/ml (1.2 x 10(-8) M) and 20 ng/ml (3.0 x 10(-7) M) zinc were estimated in connection to 60 and 1 s deposition, respectively. The response is reversible with the stripping/measurement step completely removing the accumulated zinc. Measurement frequencies as high as 30 runs per min can thus be realized (in connection to 1 s deposition and measurement steps). The influence of relevant variables of the deposition and stripping steps is examined. The attractive behavior of the stripping-voltammetric microsensor offers great promise for in vivo monitoring of trace levels of zinc
    Source: Electrochemistry communications. - ISSN 1388-2481 (Vol. 3, no. 7, 2001, str. 352-356)
    Type of material - article, component part
    Publish date - 2001
    Language - english