The high dose conformity and healthy tissue sparing achievable in Particle Therapy when using C ions calls for safety factors in treatment planning, to prevent the tumor under-dosage related to the ...possible occurrence of inter-fractional morphological changes during a treatment. This limitation could be overcome by a range monitor, still missing in clinical routine, capable of providing on-line feedback. The Dose Profiler (DP) is a detector developed within the INnovative Solution for In-beam Dosimetry in hadronthErapy (INSIDE) collaboration for the monitoring of carbon ion treatments at the CNAO facility (Centro Nazionale di Adroterapia Oncologica) exploiting the detection of charged secondary fragments that escape from the patient. The DP capability to detect inter-fractional changes is demonstrated by comparing the obtained fragment emission maps in different fractions of the treatments enrolled in the first ever clinical trial of such a monitoring system, performed at CNAO. The case of a CNAO patient that underwent a significant morphological change is presented in detail, focusing on the implications that can be drawn for the achievable inter-fractional monitoring DP sensitivity in real clinical conditions. The results have been cross-checked against a simulation study.
Full text
Available for:
IZUM, KILJ, NUK, PILJ, PNG, SAZU, UL, UM, UPUK
Positron emission tomography (PET) is the most diffuse method to image molecular processes within the human body. This paper is meant as a brief overview of the physics related to its imaging ...modality and of its numerous applications. Several references are provided where more details on the topic can be found. The first section describes the working principle of PET, the type of detectors used in a PET scanner, and the most important figures of merit to determine its performance, namely sensitivity, spatial resolution and time resolution, the latter revelant in particular for the so-called time-of-flight PET. The second section summarizes the evolution and trends for PET in clinical applications, focusing on hadrontherapy monitoring, total-body scanners, multimodal and dedicated PET systems, high spatial and temporal resolution detectors, and innovative PET radiopharmaceuticals.
Positron emission tomography is one of the most mature techniques for monitoring the particles range in hadron therapy, aiming to reduce treatment uncertainties and therefore the extent of safety ...margins in the treatment plan. In-beam PET monitoring has been already performed using inter-spill and post-irradiation data, i.e. while the particle beam is off or paused. The full beam acquisition procedure is commonly discarded because the particle spills abruptly increase the random coincidence rates and therefore the image noise. This is because random coincidences cannot be separated by annihilation photons originating from radioactive decays and cannot be corrected with standard random coincidence techniques due to the time correlation of the beam-induced background with the ion beam microstructure. The aim of this paper is to provide a new method to recover in-spill data to improve the images obtained with full-beam PET acquisitions. This is done by estimating the temporal microstructure of the beam and thus selecting input PET events that are less likely to be random ones. The PET detector we used was the one developed within the INSIDE project and tested at the CNAO synchrotron-based facility. The data were taken on a PMMA phantom irradiated with 72 MeV proton pencil beams. The obtained results confirm the possibility of improving the acquired PET data without any external signal coming from the synchrotron or ad hoc detectors.
Treatment quality assessment is a crucial feature for both present and next-generation ion therapy facilities. Several approaches are being explored, based on prompt radiation emission or on PET ...signals by β+-decaying isotopes generated by beam interactions with the body. In-beam PET monitoring at synchrotron-based ion therapy facilities has already been performed, either based on inter-spill data only, to avoid the influence of the prompt radiation, or including both in-spill and inter-spill data. However, the PET images either suffer of poor statistics (inter-spill) or are more influenced by the background induced by prompt radiation (in-spill). Both those problems are expected to worsen for accelerators with improved duty cycle where the inter-spill interval is reduced to shorten the treatment time. With the aim of assessing the detector performance and developing techniques for background reduction, a test of an in-beam PET detector prototype was performed at the CNAO synchrotron-based ion therapy facility in full-beam acquisition modality. Data taken with proton beams impinging on PMMA phantoms showed the system acquisition capability and the resulting activity distribution, separately reconstructed for the in-spill and the inter-spill data. The coincidence time resolution for in-spill and inter-spill data shows a good agreement, with a slight deterioration during the spill. The data selection technique allows the identification and rejection of most of the background originated during the beam delivery. The activity range difference between two different proton beam energies (68 and 72 MeV) was measured and found to be in sub-millimeter agreement with the expected result. However, a slightly longer (2 mm) absolute profile length is obtained for in-spill data when compared to inter-spill data.
The high dose conformity and healthy tissue sparing achievable in Particle Therapy when using C ions calls for safety factors in treatment planning, to prevent the tumor under-dosage related to the ...possible occurrence of inter-fractional morphological changes during a treatment. This limitation could be overcome by a range monitor, still missing in clinical routine, capable of providing on-line feedback. The Dose Profiler (DP) is a detector developed within the INnovative Solution for In-beam Dosimetry in hadronthErapy (INSIDE) collaboration for the monitoring of carbon ion treatments at the CNAO facility (Centro Nazionale di Adroterapia Oncologica) exploiting the detection of charged secondary fragments that escape from the patient. The DP capability to detect inter-fractional changes is demonstrated by comparing the obtained fragment emission maps in different fractions of the treatments enrolled in the first ever clinical trial of such a monitoring system, performed at CNAO. The case of a CNAO patient that underwent a significant morphological change is presented in detail, focusing on the implications that can be drawn for the achievable inter-fractional monitoring DP sensitivity in real clinical conditions. The results have been cross-checked against a simulation study.
Full text
Available for:
IZUM, KILJ, NUK, PILJ, PNG, SAZU, UL, UM, UPUK
This paper reports the development of dosimeters based on plastic scintillating fibers imaged by a charge-coupled device camera, and their performance evaluation through irradiations with the ...electron Flash research accelerator located at the Centro Pisano Flash Radiotherapy. The dosimeter prototypes were composed of a piece of plastic scintillating fiber optically coupled to a clear optical fiber which transported the scintillation signal to the readout systems (an imaging system and a photodiode). The following properties were tested: linearity, capability to reconstruct the percentage depth dose curve in solid water and to sample in time the single beam pulse. The stem effect contribution was evaluated with three methods, and a proof-of-concept one-dimensional array was developed and tested for online beam profiling. Results show linearity up to 10 Gy per pulse, and good capability to reconstruct both the timing and spatial profiles of the beam, thus suggesting that plastic scintillating fibers may be good candidates for low-energy electron Flash dosimetry.
•Flash radiotherapy may be groundbreaking for improving cancer treatment.•Conventional dosimeters do not function properly in the Flash regime.•Dosimeters based on plastic scintillating fibers were developed and characterized.•No quenching or energy dependence was observed in the Flash regime.•The beam time structure could be properly sampled.
Full text
Available for:
GEOZS, IJS, IMTLJ, KILJ, KISLJ, NLZOH, NUK, OILJ, PNG, SAZU, SBCE, SBJE, UILJ, UL, UM, UPCLJ, UPUK, ZAGLJ, ZRSKP
A PET detector featuring a pseudo-monolithic crystal is being developed as a more cost-effective alternative to a full monolithic crystal PET detector. This work evaluates different methods to ...localize the scintillation events in quartered monolithic crystals that are optically coupled. A semi-monolithic crystal assembly was formed using four 26 × 26 × 10 mm3 LYSO crystals optically coupled together using optical adhesive, to mimic a 52 × 52 × 10 mm3 monolithic crystal detector. The crystal assembly was coupled to a 64-channel multi-anode photomultiplier tube using silicon grease. The detector was calibrated using a 34 × 34 scan grid. Events were first filtered and depth separated using a multi-Lorentzian fit to the collected light distribution. Next, three different techniques were explored to generate the look up tables for the event positioning. The first technique was 'standard interpolation' across the interface. The second technique was 'central extrapolation', where a bin was placed at the midpoint of the interface and events positioned within the interface region were discarded. The third technique used a 'central overlap' method where an extended region was extrapolated at each interface. Events were then positioned using least-squares minimization and maximum likelihood methods. The least-squares minimization applied to the look up table generated with the standard interpolation technique had the best full width at half maximum (FWHM) intrinsic spatial resolution and the lowest bias. However, there were discontinuities in the event positioning that would most likely lead to artifacts in the reconstructed image. The central extrapolation technique also had discontinuities and a 30% sensitivity loss near the crystal-crystal interfaces. The central overlap technique had slightly degraded performance metrics, but it still provided ~2.1 mm intrinsic spatial resolution at the crystal-crystal interface and had a symmetric and continuously varying response function. Results using maximum likelihood positioning were similar to least-squares minimization for the central overlap data.
Positron emission tomography (PET) is one of the most mature techniques for monitoring in ion beam therapy. PET allows to reconstruct the β+ activity generated in the patient by the nuclear ...interaction of the ions. Taking advantage of the spatial correlation between positron emitters created along the ions path and the dose distribution, it is possible to perform a quality control of the treatment. Usually, to reconstruct the activity generated within the irradiated volume, standard 3D PET reconstruction techniques are implemented. In this work, we explore a new reconstruction method (Straight Forward Reconstruction) particularly useful for reconstructing activity distribution generated by mono-energetic pencil beams. The method was validated by measuring the produced activity distribution with the DoPET system. Irradiations performed with mono-energetic pencil beams on phantoms mimicking human tissues were used for this study. Both reconstruction methods reach an accuracy in the reconstruction of the activity distribution width of the order of 1.5 mm for 2⋅108 primaries.
•PET is one of the most mature techniques for monitoring in ion beam therapy.•A new reconstruction method (Straight Forward Reconstruction) for PET is presented.•Activity distribution and accuracy by pencil beams irradiation are presented.
Full text
Available for:
GEOZS, IJS, IMTLJ, KILJ, KISLJ, NLZOH, NUK, OILJ, PNG, SAZU, SBCE, SBJE, UILJ, UL, UM, UPCLJ, UPUK, ZAGLJ, ZRSKP
Abstract
Background and purpose:
in-beam Positron Emission Tomography (PET) is one of the modalities that can be used for in-vivo non-invasive treatment monitoring in proton therapy. PET ...distributions obtained during various treatment sessions can be compared in order to identify regions that have anatomical changes. The purpose of this work is to test and compare different analysis methods in the context of inter-fractional PET image comparison for proton treatment verification.
Methods:
for our study we used the FLUKA Monte Carlo code and artificially generated CT scans to simulate in-beam PET distributions at different stages during proton therapy treatment. We compared the Beam-Eye-View method, the Most-Likely-Shift method, the Voxel-Based-Morphology method and the gamma evaluation method to compare PET images at the start of treatment, and after a few weeks of treatment. The results were compared to the CT scan.
Results and conclusions:
three-dimensional methods like VBM and gamma are preferred above two-dimensional methods like MLS and BEV if much statistics is available, since the these methods allow to identify the regions with anomalous activity. The VBM approach has as disadvantage that a larger number of MC simulations is needed. The gamma analysis has the disadvantage that no clinical indication exist on tolerance criteria. In terms of calculation time, the BEV and MLS method are preferred. We recommend to use the four methods together, in order to best identify the location and cause of the activity changes.
In this paper, we investigate the use of a positron emission tomography (PET) system to monitor the proton therapy. The monitoring procedure is based on the comparison between the <inline-formula> ...<tex-math notation="LaTeX">{ \beta }+ </tex-math></inline-formula> activity generated in the irradiated volume during the treatment, with the <inline-formula> <tex-math notation="LaTeX">{ \beta }+ </tex-math></inline-formula> activity distribution obtained with Monte Carlo (MC) simulation. The dedicated PET system is a dual head detection system; each head is composed of nine scintillating LYSO crystal matrices read out independently with a custom modularized acquisition system. Our experimental data were acquired at the Cyclotron Centre Bronowice, Institute Nuclear Physics in Kraków, Poland, and were simulated with the FLUKA MC code. Homogeneous and heterogeneous plastic phantoms were irradiated with monoenergetic 130 MeV protons. The capabilities of our PET system to distinguish different irradiated materials were investigated, and the proton pencil-beams were used as probes. Our focus was to analyze the activity width and the total activity event number in several cases. Irradiations were performed using either single pencil-beams one at a time, or two pencil-beams during the same data taking. The comparison of 1-D activity profile for experimental data and MC simulation were always in good agreement showing that, the treatment quality assessment in proton therapy can be based on <inline-formula> <tex-math notation="LaTeX">{ \beta }+ </tex-math></inline-formula> activity measurements.