Purpose:
Silicon diode arrays are commonly implemented in radiation therapy quality assurance applications as they have a number of advantages including: real time operation (compared to the film) ...and high spatial resolution, large dynamic range and small size (compared to ionizing chambers). Most diode arrays have detector pitch that is too coarse for routine use in small field applications. The goal of this work is to characterize the two‐dimensional monolithic silicon diode array named “MagicPlate‐512” (MP512) designed for QA in stereotactic body radiation therapy (SBRT) and stereotactic radio surgery (SRS).
Methods:
MP512 is a silicon monolithic detector manufactured on ap‐type substrate. An array contains of 512 pixels with size 0.5 × 0.5 mm2 and pitch 2 mm with an overall dimension of 52 × 52 mm2. The MP512 monolithic detector is wire bonded on a printed circuit board 0.5 mm thick and covered by a thin layer of raisin to preserve the silicon detector from moisture and chemical contamination and to protect the bonding wires. Characterization of the silicon monolithic diode array response was performed, and included pixels response uniformity, dose linearity, percent depth dose, output factor, and beam profiling for beam sizes relevant to SBRT and SRS and depth dose response in comparison with ionization chamber.
Results:
MP512 shows a good dose linearity (R2 = 0.998) and repeatability within 0.2%. The measured depth dose response for field size of 10 × 10 cm2 agreed to within 1.3%, when compared to a CC13 ionization chamber for depths in PMMA up to 30 cm. The output factor of a 6 MV Varian 2100EX medical linac beam measured by MP512 at the isocenter agrees to within 2% when compared to PTW diamond, Scanditronix point EDD‐2 diode and MOSkin detectors for field sizes down to 1 × 1 cm2. An over response of 4% was observed for square beam size smaller than 1 cm when compared to EBT3 films, while the beam profiles (FWHM) of MP512 match to within 2% the data measured by radiochromic film.
Conclusions:
The response of the 2D detector array, MP512, has been evaluated. The properties of the array demonstrated suitability for use as in phantom dosimeter for QA in SRS and SBRT. Although MP512 matches film measurements down to 1 × 1 cm2 well, it showed a discrepancy of 4% in the determination of output factors of beams smaller than 0.5 × 0.5 cm2 due to the field perturbation generated by the large amount of silicon surrounding the central diode. MP512 is highly capable of measuring beam size (FWHM) and has a discrepancy of less than 1.3% when compared to EBT3 film. A reduction in the detector pitch to less than 2 mm would improve the penumbra reconstruction accuracy at the cost readout electronics complexity.
Full text
Available for:
BFBNIB, FZAB, GIS, IJS, KILJ, NLZOH, NUK, OILJ, SBCE, SBMB, UL, UM, UPUK
Proton therapy has a distinct dosimetric advantage over conventional photon therapy due to its Bragg peak profile. This allows greater accuracy in dose delivery and dose conformation to the target, ...however it requires greater precision in setup, delivery and for quality assurance (QA) procedures. The AAPM TG 224 report recommends daily range and spot position checks with tolerance on the order of a millimetre. Daily QA systems must therefore be efficient for daily use and be capable of sub-millimetre precision however few suitable commercial systems are available. In this work, a compact, real-time daily QA system is optimised and characterised for proton range verification using an ad-hoc Geant4 simulation. The system is comprised of a monolithic silicon diode array detector embedded in a perspex phantom. The detector is orientated at an angular offset to the incident proton beam to allow range in perspex to be determined for flat proton fields. The accuracy of the system for proton range in perspex measurements was experimentally evaluated over the full range of clinical proton energies. The mean
,
and
ranges measured with the system were accurate within ±0.6 mm of simulated ranges in a perspex phantom for all energies assessed. This system allows real-time read-out of individual detector channels also making it appropriate for temporal beam delivery diagnostics and for spot position monitoring along one axis. The system presented provides a suitable, economical and efficient alternative for daily QA in proton therapy.
Purpose:
This paper presents initial experimental results from a prototype of high dose rate (HDR) BrachyView, a novel in‐body source tracking system for HDR brachytherapy based on a multipinhole ...tungsten collimator and a high resolution pixellated silicon detector array. The probe and its associated position estimation algorithms are validated and a comprehensive evaluation of the accuracy of its position estimation capabilities is presented.
Methods:
The HDR brachytherapy source is moved through a sequence of positions in a prostate phantom, for various displacements in x, y, and z. For each position, multiple image acquisitions are performed, and source positions are reconstructed. Error estimates in each dimension are calculated at each source position and combined to calculate overall positioning errors. Gafchromic film is used to validate the accuracy of source placement within the phantom.
Results:
More than 90% of evaluated source positions were estimated with an error of less than one millimeter, with the worst‐case error being 1.3 mm. Experimental results were in close agreement with previously published Monte Carlo simulation results.
Conclusions:
The prototype of HDR BrachyView demonstrates a satisfactory level of accuracy in its source position estimation, and additional improvements are achievable with further refinement of HDR BrachyView's image processing algorithms.
Full text
Available for:
BFBNIB, FZAB, GIS, IJS, KILJ, NLZOH, NUK, OILJ, SBCE, SBMB, UL, UM, UPUK
Microdosimetry is a particularly powerful method to estimate the relative biological effectiveness (RBE) of any mixed radiation field. This is particularly convenient for therapeutic heavy ion ...therapy (HIT) beams, referring to ions larger than protons, where the RBE of the beam can vary significantly along the Bragg curve. Additionally, due to the sharp dose gradients at the end of the Bragg peak (BP), or spread out BP, to make accurate measurements and estimations of the biological properties of a beam a high spatial resolution is required, less than a millimetre. This requirement makes silicon microdosimetry particularly attractive due to the thicknesses of the sensitive volumes commonly being 10 m or less. Monte Carlo (MC) codes are widely used to study the complex mixed HIT radiation field as well as to model the response of novel microdosimeter detectors when irradiated with HIT beams. Therefore it is essential to validate MC codes against experimental measurements. This work compares measurements performed with a silicon microdosimeter in mono-energetic , and ion beams of therapeutic energies, against simulation results calculated with the Geant4 toolkit. Experimental and simulation results were compared in terms of microdosimetric spectra (dose lineal energy, ), the dose mean lineal energy, y D and the RBE10, as estimated by the microdosimetric kinetic model (MKM). Overall Geant4 showed reasonable agreement with experimental measurements. Before the distal edge of the BP, simulation and experiment agreed within 10% for y D and 2% for RBE10. Downstream of the BP less agreement was observed between simulation and experiment, particularly for the and beams. Simulation results downstream of the BP had lower values of y D and RBE10 compared to the experiment due to a higher contribution from lighter fragments compared to heavier fragments.
Microbeam radiation therapy (MRT) is a new radiation treatment modality in the pre-clinical stage of development at the ID17 Biomedical Beamline of the European synchrotron radiation facility (ESRF) ...in Grenoble, France. MRT exploits the dose volume effect that is made possible through the spatial fractionation of the high dose rate synchrotron-generated x-ray beam into an array of microbeams. As an important step towards the development of a dosimetry protocol for MRT, we have applied the International Atomic Energy Agency's TRS 398 absorbed dose-to-water protocol to the synchrotron x-ray beam in the case of the broad beam irradiation geometry (i.e. prior to spatial fractionation into microbeams). The very high dose rates observed here mean the ion recombination correction factor, ks, is the most challenging to quantify of all the necessary corrections to apply for ionization chamber based absolute dosimetry. In the course of this study, we have developed a new method, the so called 'current ramping' method, to determine ks for the specific irradiation and filtering conditions typically utilized throughout the development of MRT. Using the new approach we deduced an ion recombination correction factor of 1.047 for the maximum ESRF storage ring current (200 mA) under typical beam spectral filtering conditions in MRT. MRT trials are currently underway with veterinary patients at the ESRF that require additional filtering, and we have estimated a correction factor of 1.025 for these filtration conditions for the same ESRF storage ring current. The protocol described herein provides reference dosimetry data for the associated Treatment Planning System utilized in the current veterinary trials and anticipated future human clinical trials.
Purpose:
High dose rate (HDR) brachytherapy is a radiation treatment technique capable of delivering large dose rates to the tumor. Radiation is delivered using remote afterloaders to drive highly ...active sources (commonly192Ir with an air KERMA strength range between 20 000 and 40 000 U, where 1 U = 1 μGy m2/h in air) through applicators directly into the patient's prescribed region of treatment. Due to the obvious ramifications of incorrect treatment while using such an active source, it is essential that there are methods for quality assurance (QA) that can directly and accurately verify the treatment plan and the functionality of the remote afterloader. This paper describes the feasibility study of a QA system for HDR brachytherapy using a phantom based two-dimensional 11 × 11 epitaxial diode array, named “magic phantom.”
Methods:
The HDR brachytherapy treatment plan is translated to the phantom with two rows of 10 (20 in total) HDR source flexible catheters, arranged above and below the diode array “magic plate” (MP). Four-dimensional source tracking in each catheter is based upon a developed fast iterative algorithm, utilizing the response of the diodes in close proximity to the192Ir source, sampled at 100 ms intervals by a fast data acquisition (DAQ) system. Using a 192Ir source in a solid water phantom, the angular response of the developed epitaxial diodes utilized in the MP and also the variation of the MP response as a function of the source-to-detector distance (SDD) were investigated. These response data are then used by an iterative algorithm for source dwelling position determination. A measurement of the average transit speed between dwell positions was performed using the diodes and a fast DAQ.
Results:
The angular response of the epitaxial diode showed a variation of 15% within 360°, with two flat regions above and below the detector face with less than 5% variation. For SDD distances of between 5 and 30 mm the relative response of the epitaxial diodes used in the MP is in good agreement (within 8%) with radial dose function measurements found within the TG-43 protocol, with SDD of up to 70 mm showing a 40% over response. A method for four-dimensional localization of the HDR source was developed, allowing the source dwell position to be derived within 0.50 mm of the expected position. An estimation of the average transit speed for varying step sizes was determined and was found to increase from (12.8 ± 0.3) up to (38.6 ± 0.4) cm/s for a step size of 2.5 and 50 mm, respectively.
Conclusions:
Our characterization of the designed QA “magic phantom” with MP in realistic HDR photon fields demonstrates the promising performance for real-time source position tracking in four dimensions and measurements of transit times. Further development of this system will allow a full suite for QA in HDR brachytherapy and analysis, and for futurein vivo tracking.
Full text
Available for:
BFBNIB, FZAB, GIS, IJS, KILJ, NLZOH, NUK, OILJ, SBCE, SBMB, UL, UM, UPUK
A new batch of microdosimeters has been extensively studied for their charge collection efficiency (CCE) properties, as well as their radiation hardness for medical, space and accident applications. ...Silicon-on-insulator (SOI) microdosimeters with an active layer thickness of 10, 20 and 50 μm have been investigated and were characterized with a 24 MeV carbon ion beam as well as a Co-60 gamma source. A negative pulse was observed in addition to the positive pulses generated within the sensitive volumes (SVs) by incident ions which led to undesirable low energy events in the SOI microdosimeters response. To study this phenomenon, the microdosimeters were irradiated with gamma radiation from a Co-60 source with a total dose of 3 and 10 Mrad(Si). It was determined that the negative pulse was originating from the support wafer due to the displacement current phenomenon. Irradiation with the Co-60 source led to a disappearing of the negative pulse due to an increase in recombination within the support wafer while almost no changes in CCE were observed. A radiation hardness study was also performed on the 50 μm SOI microdosimeter with 16 SVs being irradiated with a fluence of ~ 10 8 12 C ions/cm 2 . A CCE deficit of approximately 2% was observed at an operation bias of 10V within the SVs. The findings of this work demonstrate that the SOI microdosimeters can be utilized in space and medical applications as they can handle typical levels of dose encountered in these applications. Additionally, evidence for SOI microdosimeter fabrication standards in terms of support wafer resistivity and buried oxide (BOX) thickness is shown.
Microbeam radiation therapy (MRT) is a developing radiotherapy, based on the use of beams only a few tens of micrometres wide, generated by synchrotron X‐ray sources. The spatial fractionation of the ...homogeneous beam into an array of microbeams is possible using a multislit collimator (MSC), i.e. a machined metal block with regular apertures. Dosimetry in MRT is challenging and previous works still show differences between calculated and experimental dose profiles of 10–30%, which are not acceptable for a clinical implementation of treatment. The interaction of the X‐rays with the MSC may contribute to the observed discrepancies; the present study therefore investigates the dose contribution due to radiation interaction with the MSC inner walls and radiation leakage of the MSC. Dose distributions inside a water‐equivalent phantom were evaluated for different field sizes and three typical spectra used for MRT studies at the European Synchrotron Biomedical beamline ID17. Film dosimetry was utilized to determine the contribution of radiation interaction with the MSC inner walls; Monte Carlo simulations were implemented to calculate the radiation leakage contribution. Both factors turned out to be relevant for the dose deposition, especially for small fields. Photons interacting with the MSC walls may bring up to 16% more dose in the valley regions, between the microbeams. Depending on the chosen spectrum, the radiation leakage close to the phantom surface can contribute up to 50% of the valley dose for a 5 mm × 5 mm field. The current study underlines that a detailed characterization of the MSC must be performed systematically and accurate MRT dosimetry protocols must include the contribution of radiation leakage and radiation interaction with the MSC in order to avoid significant errors in the dose evaluation at the micrometric scale.
The study of the interaction of synchrotron X‐ray radiation with a multislit collimator is a critical factor for the creation of microbeams in radiation therapy. This study is a fundamental step forward in the understanding and definition of a reliable protocol for dosimetry at the micrometric scale to be used in microbeam radiation therapy.
Full text
Available for:
FZAB, GIS, IJS, IZUM, KILJ, NLZOH, NUK, OILJ, PILJ, PNG, SAZU, SBCE, SBMB, UL, UM, UPUK
Purpose:
High dose rate (HDR) brachytherapy is a treatment method that is used increasingly worldwide. The development of a sound quality assurance program for the verification of treatment ...deliveries can be challenging due to the high source activity utilized and the need for precise measurements of dwell positions and times. This paper describes the application of a novel phantom, based on a 2D 11 × 11 diode array detection system, named “magic phantom” (MPh), to accurately measure plan dwell positions and times, compare them directly to the treatment plan, determine errors in treatment delivery, and calculate absorbed dose.
Methods:
The magic phantom system was CT scanned and a 20 catheter plan was generated to simulate a nonspecific treatment scenario. This plan was delivered to the MPh and, using a custom developed software suite, the dwell positions and times were measured and compared to the plan. The original plan was also modified, with changes not disclosed to the primary authors, and measured again using the device and software to determine the modifications. A new metric, the “position–time gamma index,” was developed to quantify the quality of a treatment delivery when compared to the treatment plan. The MPh was evaluated to determine the minimum measurable dwell time and step size. The incorporation of the TG‐43U1 formalism directly into the software allows for dose calculations to be made based on the measured plan. The estimated dose distributions calculated by the software were compared to the treatment plan and to calibrated EBT3 film, using the 2D gamma analysis method.
Results:
For the original plan, the magic phantom system was capable of measuring all dwell points and dwell times and the majority were found to be within 0.93 mm and 0.25 s, respectively, from the plan. By measuring the altered plan and comparing it to the unmodified treatment plan, the use of the position–time gamma index showed that all modifications made could be readily detected. The MPh was able to measure dwell times down to 0.067 ± 0.001 s and planned dwell positions separated by 1 mm. The dose calculation carried out by the MPh software was found to be in agreement with values calculated by the treatment planning system within 0.75%. Using the 2D gamma index, the dose map of the MPh plane and measured EBT3 were found to have a pass rate of over 95% when compared to the original plan.
Conclusions:
The application of this magic phantom quality assurance system to HDR brachytherapy has demonstrated promising ability to perform the verification of treatment plans, based upon the measured dwell positions and times. The introduction of the quantitative position–time gamma index allows for direct comparison of measured parameters against the plan and could be used prior to patient treatment to ensure accurate delivery.
Full text
Available for:
BFBNIB, FZAB, GIS, IJS, KILJ, NLZOH, NUK, OILJ, SBCE, SBMB, UL, UM, UPUK
This article presents a study of the response of the silicon on insulator (SOI) microdosimeter with 3-D sensitive volumes (SVs) to 400 MeV/u 16 O and 500 MeV/u 56 Fe ions mimicking galactic cosmic ...rays outside and inside the International Space Station (ISS). An average quality factor (Q̅) and the dose equivalent (H) of the radiation field were obtained experimentally, and the results were compared with GEANT4 simulations.