To compare the diagnostic accuracy and radiation doses of two low-dose protocols for coronary artery imaging with second-generation, dual-source CT in comparison with catheter angiography (CA).
...Prospective, single-centre study conducted in a referral centre enrolling 100 patients with low-to-intermediate risk and suspicion of coronary artery disease. All patients underwent contrast-enhanced, 128-slice, dual-source CT coronary angiography and CA. Patients were randomly assigned to two different low-dose CT protocols (each 100 kV/320 mA): in group A (n=50), CT was performed using the prospectively electrocardiography (ECG)-gated step-and-shoot (SAS) mode; in group B (n=50), CT was performed using the prospectively ECG-gated high-pitch mode (pitch 3.4). The image quality and presence or absence of significant coronary stenosis in all coronary segments were evaluated by two blinded and independent observers. CA served as the standard of reference.
Sixty-one significant stenoses were found in group A, and 69 in group B. There was no significant difference in age (group A, 62+/-8 yrs; group B, 63+/-8 yrs; p=0.72), body mass index (group A, 26.4+/-3.1 kg/m2; group B, 25.9+/-2.8 kg/m2; p=0.41) and heart rate (HR) (group A, 58+/-8 bpm; group B, 56+/-10 bpm; p=0.66) between the groups. Diagnostic image quality was obtained in 98.6% (651/660) of segments in group A and in 98.9% (642/649) in group B, with no significant differences between groups. Sensitivity, specificity and positive and negative predictive values were 94%, 91%, 85% and 97% per-patient in group A, and 93%, 94%, 89% and 97% per-patient in group B (no significant differences). The effective radiation dose in group B (0.9+/-0.1 mSv) was significantly (p<0.01) lower than that in group A (1.4+/-0.4 mSv).
Both the high-pitch and the SAS mode for low-dose CT coronary angiography provide high accuracy for the assessment of significant coronary stenoses, while the high-pitch mode further significantly lowers the radiation dose.
Abstract
Purpose:
Percutaneous transhepatic primary biliary stenting (PS) is an alternative to the widely used staged procedure (secondary biliary stenting, SS) for treating obstructive jaundice in ...cancer patients. To evaluate the efficacy and safety of PS and SS, a retrospective analysis was carried out.
Materials and Methods:
The percutaneous biliary stenting procedures performed between January 2000 and December 2007 at one hepatobiliary center were retrospectively analyzed, comparing the technical success rates, complications, and length of hospitalization of the two procedures. Of 61 patients (mean age 65.5 ± 13.1 years; range 31.1-92.7 years) suffering from obstructive jaundice caused by primary or metastatic tumors, 30 received PS and 31 received SS. The groups were comparable in the anatomical level of the obstruction, stent configuration, or the concurrent presence of cholangitis. Placement of metallic stents required one session for patients in the PS group and an average of 2.6 ± 1.1 sessions for patients in the SS group.
Results:
The overall technical success rate was 98.4% with 1 (1/61) failed approach to transcend the occlusion in the SS group. The rate of minor complications was 10% in the PS group and 6.5% in the SS group. The corresponding rates of major complications were 23.3% and 54.8%, respectively. SS patients had a higher rate of complications in general (
P
< 0.05), as well as a higher rate of severe complications in particular (
P
< 0.05). Procedural mortality was 0% for both the groups. The mean overall length of hospitalization was 7.7 ± 9.6 days for PS and 20.6 ± 19.6 days for SS (
P
< 0.001).
Conclusion:
Primary percutaneous biliary stenting of malignant biliary obstructions is as efficacious and safer than a staged procedure with secondary stenting. By virtue of requiring shorter hospital stays, primary stenting is likely to be more cost-effective.
Full text
Available for:
IZUM, KILJ, NUK, PILJ, PNG, SAZU, UL, UM, UPUK
Background Identification and differentiation of coronary atherosclerotic plaques may improve risk stratification for incident coronary events. Objective We investigated the ability of dual-source ...computed tomography (CT) to depict and characterize atherosclerotic coronary plaques. Methods Contrast-enhanced CT was performed in 25 human heart specimens with a total of 322 histologically determined plaques. Coronary plaques were classified on CT as (1) noncalcified, mixed, or calcified and (2) by CT attenuation values. Atherosclerotic plaques were histopathologically characterized according to the Stary classification. Results CT detected 79% (245/322) of all plaques. Lesions missed by CT were generally early lesions, type I (n = 31), type II (n = 38), or type III (n = 8), according to Stary. CT detected 29% of early (Stary I–III) and 100% of advanced (Stary IV–VIII) plaques. Plaque classification as noncalcified was sensitive (100%) and specific (72%) for early, whereas classification as mixed/calcified was sensitive (92%, 89%) and specific (100%) for advanced plaques. Calcified plaques on CT were detected with high sensitivity (80%) and specificity (95%). Other subtypes were not distinguishable with CT according to the presence or absence of calcification. CT density was significantly higher for advanced (306 ± 470 HU) than for early (42 ± 14 HU; P < 0.01) plaques. The mean CT density value of type VII plaques (512 ± 349 HU) was significantly higher than those of other plaques (34–101 HU; P < 0.001). Conclusions CT reliably depicts advanced coronary plaques and allows for the differentiation between early and advanced plaques.
Full text
Available for:
GEOZS, IJS, IMTLJ, KILJ, KISLJ, NUK, OILJ, PNG, SAZU, SBCE, SBJE, UL, UM, UPCLJ, UPUK
The purpose of this study was to investigate the effect of low kilovoltage dual-source computed tomography coronary angiography (CTCA) on qualitative and quantitative image quality parameters and ...radiation dose. Dual-source CTCA with retrospective ECG gating was performed in 80 consecutive patients of normal weight. Forty were examined with a standard protocol (120 kV/330mAs), 20 were examined at 100 kV/330mAs, and 20 at 100 kV/220mAs. Two blinded observers independently assessed image quality of each coronary segment and measured the image parameters noise, attenuation, and contrast-to-noise ratio (CNR). The effective radiation dose was calculated using CT dose volume index and the dose-length product. Diagnostic image quality was obtained in 99% of all coronary segments (1,127/1,140) without significant differences among the protocols. Image noise, attenuation, and CNR were significantly higher for 100 kV/330mAs (26 ± 3 HU, 549 ± 62 HU, 25.5 ± 3.2; each
P
< 0.01) and 100 kV/220mAs (27 ± 2 HU, 560 ± 43 HU, 25.0 ± 2.2; each
P
< 0.01) when compared to the 120-kV protocol (21 ± 2 HU, 317 ± 28 HU, 20.6 ± 1.7). There was no significant difference between the two 100-kV protocols. Estimated effective radiation dose of the 120-kV protocol (8.9 ± 1.2 mSv) was significantly higher than the 100 kV/330mAs (6.7 ± 0.8 mSv,
P
< 0.01) or 100 kV/220mAs (4.4 ± 0.6 mSv,
P
< 0.001) protocols. Dual-source CTCA with 100 kV is feasible in patients of normal weight, results in a diagnostic image quality with a higher CNR, and at the same time significantly reduces the radiation dose.
Full text
Available for:
EMUNI, FIS, FZAB, GEOZS, GIS, IJS, IMTLJ, KILJ, KISLJ, MFDPS, NLZOH, NUK, OILJ, PNG, SAZU, SBCE, SBJE, SBMB, SBNM, UKNU, UL, UM, UPUK, VKSCE, VSZLJ, ZAGLJ
To assess the diagnostic performance of dual-energy dual-source computed tomography (CT) in the detection of endoleaks after endovascular abdominal aortic aneurysm (AAA) repair.
This study was local ...ethics board approved, and written informed consent was obtained from all patients. One hundred eighteen patients (21 women, 97 men; mean age, 74 years +/- 8 standard deviation) underwent follow-up dual-energy dual-source CT during the nonenhanced, arterial, and delayed phases after AAA repair. Delayed phase CT images were acquired in the dual-energy mode for reconstruction of virtual nonenhanced images. Two blinded and independent readers evaluated the data for the presence or absence of endoleaks during three reading sessions: Standard nonenhanced, arterial phase, and delayed phase images were read during session A; virtual nonenhanced, arterial phase, and delayed phase images, during session B; and virtual nonenhanced and delayed phase images, during session C. Sensitivity, specificity, negative predictive value (NPV), and positive predictive value (PPV) were calculated, with the session A image data set as the reference standard. Radiation dose parameters were estimated.
Reading session A revealed that 52 (44%) of 118 patients had endoleaks. Overall sensitivity, specificity, NPV, and PPV for CT endoleak detection during sessions B and C were identical: 100%, 97%, 100%, and 96%, respectively. The accuracy of the session B and session C readings was not significantly different from that of the session A reading (P = .50). The effective radiation dose in the image acquisition protocol involving one dual-energy scan was significantly (P < .001) lower than the effective doses in the protocols involving standard triple-phase scanning (mean difference, 61%) and standard nonenhanced and delayed phase scanning (mean difference, 41%).
Compared with standard protocols, one dual-energy dual-source CT scan performed during the delayed phase with reconstruction of virtual nonenhanced images enables detection of endoleaks after endovascular AAA repair with high accuracy and a considerably lower radiation dose.
To prospectively investigate computed tomographic (CT) image quality parameters by using different protocols and to calculate radiation dose estimates for noninvasive coronary angiography performed ...with dual-source CT in the step-and-shoot (SAS) mode.
This study was local ethics board approved; written informed consent was obtained from all patients. In the preliminary portion of the study, 40 patients underwent CT coronary angiography in the SAS mode: at 100 kV (protocol A) in 22 patients with a body mass index (BMI) of less than 25 kg/m(2) and at 120 kV (protocol B) in 18 patients with a BMI of 25-30 kg/m(2). Both protocols involved use of an attenuation-based tube current and 1 mL of contrast material per kilogram of body weight. The final portion of the study involved 50 additional patients: 21 patients with a BMI of 25-30 kg/cm(2) assigned to protocol B and 29 patients with a BMI of less than 25 kg/cm(2) assigned to protocol C, which was performed with 100 kV, an attenuation-based tube current, and a reduced contrast material dose of 0.8 mL/kg. Image quality was independently assessed. Attenuation in the aorta and coronary arteries and image noise were measured. Radiation dose was estimated.
Mean image noise was similar with protocols A and B. Mean attenuation in the aorta and coronary arteries with protocol A (444 HU) was significantly (P < .001) higher than that with protocol B (358 HU). The reduced contrast material dose in protocol C yielded attenuation similar to that with protocol B. Diagnostic image quality was achieved with all protocols in 1237 (97.9%) of 1264 coronary segments. No significant differences in image quality between the 100- and 120-kV protocols were found. Mean heart rate had a significant effect on motion artifacts (area under receiver operating characteristic curve AUC = 0.818; 95% confidence interval CI: 0.723, 0.892; P < .001), whereas heart rate variability had a significant effect on stair-step artifacts (AUC = 0.79; 95% CI: 0.687, 0.865; P < .001). The mean estimated effective dose was 1.2 mSv +/- 0.2 for protocols A and C and 2.6 mSv +/- 0.5 for protocol B.
Dual-source SAS-mode CT coronary angiography yielded diagnostic image quality for 97.9% of coronary segments at a low radiation dose.
Meeting abstracts - A single PDF containing all abstracts in this supplement is available at http://www.biomedcentral.com/content/files/pdf/1532-429X-13-S1-full.pdf.EMPTY Purpose To assess a late ...gadolinium enhancement (LGE) imaging single breath-hold 3D inversion recovery sequence for the quantification of myocardial scar mass and transmurality in comparison to a clinically established 2D acquisition sequence.
Full text
Available for:
DOBA, GEOZS, IJS, IMTLJ, IZUM, KILJ, KISLJ, NLZOH, NUK, OILJ, PILJ, PNG, SAZU, SBCE, SBJE, UILJ, UKNU, UL, UM, UPCLJ, UPUK, ZAGLJ, ZRSKP
The aim of this study was to assess the value of multislice computed tomography (CT) for the assessment of valvular abnormalities in patients with infective endocarditis (IE) in comparison with ...transesophageal echocardiography (TEE) and intraoperative findings.
Multislice CT has recently shown promising data regarding valvular imaging in a 4-dimensional fashion.
Thirty-seven consecutive patients with clinically suspected IE were examined with TEE and 64-slice CT or dual-source CT. Twenty-nine patients had definite IE and underwent surgery.
The diagnostic performance of CT for the detection of evident valvular abnormalities for IE compared with TEE was: sensitivity 97%, specificity 88%, positive predictive value (PPV) 97%, and negative predictive value (NPV) 88% on a per-patient basis (n = 37; excellent intermodality agreement kappa = 0.84). CT correctly identified 26 of 27 (96%) patients with valvular vegetations and 9 of 9 (100%) patients with abscesses/pseudoaneurysms compared with the intraoperative specimen. On a per-valve-based analysis, diagnostic accuracy for the detection of vegetations and abscesses/pseudoaneurysms compared with surgery was: sensitivity 96%, specificity 97%, PPV 96%, NPV 97%, and sensitivity 100%, specificity 100%, PPV 100%, NPV 100%, respectively, without significant differences as compared with TEE. Vegetation size measurements by CT correlated (r = 0.95; p <0.001) with TEE (mean 7.6 +/- 5.6 mm). The mobility of vegetations was accurately diagnosed in 21 of 22 (96%) patients with CT, but all of 4 leaflet perforations (<or=2 mm) were missed. CT provided more accurate anatomic information regarding perivalvular extent of abscess/pseudoaneurysms than TEE.
Multislice CT shows good results in detecting valvular abnormalities in IE and could be applied in pre-operative planning and exclusion of coronary artery disease before surgery.
Full text
Available for:
GEOZS, IJS, IMTLJ, KILJ, KISLJ, NLZOH, NUK, OILJ, PNG, SAZU, SBCE, SBJE, UILJ, UL, UM, UPCLJ, UPUK, ZAGLJ, ZRSKP
To evaluate the quality of stent lumen delineation using dual-source computed tomography (DSCT) in the standard-pitch mode (SP) as compared to the high-pitch mode (HP) in a phantom study.
Forty ...different coronary stents placed in plastic tubes filled with contrast agent were imaged with a second generation DSCT system in a SP (pitch 0.23) and HP (pitch 3.4) mode in orientations of 0°, 45°, and 90° relative to the z-axis. Two observers independently measured the in-stent lumen and the attenuation values in the center of the stents. The artificial lumen narrowing (ALN) was calculated using the measured in-stent lumen and the nominal diameter of the plastic tube.
Interobserver correlation was excellent for in-stent lumen (0.86) and attenuation measurements (0.91). There was no significant difference neither for ALN (SP: 54.7-62.8%; HP: 55.8-64.0%) nor attenuation (SP: 356-395 Hounsfield units HU; HP: 352-384 HU) between SP and HP mode. For both modes, the orientation of the stent relative to the z-axis significantly affected ALN and attenuation (each P < .001). CT volume dose index was significantly lower using HP mode as compared to SP mode (P < .001).
The HP mode in DSCT provides visualization of the coronary in-stent lumen comparable to that measured in SP mode while reducing applied radiation dose in a stationary phantom model.
ABSTRACTIpilimumab is a monoclonal antibody against the inhibitory CTLA-4 receptor expressed on T cells. It provokes an upregulation of the immune system. This substance was approved by the US Food ...and Drug Administration in 2011 and is since increasingly used as a targeted therapeutic approach for metastasized melanoma. Ipilimumab is known to cause neuroendocrine disorders, such as hypophysitis and adrenal insufficiency. Our case of a 79-year-old patient represents an important imaging pitfall. Imaging findings of newly symmetrically and smoothly enlarged, hypermetabolic adrenal glands in the setting of previous ipilimumab therapy represent drug-induced adrenalitis and not metastatic disease.