  • Zero-dark-counting high-spe...
    Yamaguchi, Satoshi; Sato, Eiichi; Oda, Yasuyuki; Nakamura, Ryuji; Oikawa, Hirobumi; Yabuushi, Tomonori; Ariga, Hisanori; Ehara, Shigeru

    Japanese Journal of Applied Physics, 04/2014, Volume: 53, Issue: 4
    Journal Article

    X-ray photons are detected using a cerium-doped yttrium aluminum perovskite YAP(Ce) single-crystal scintillator with a decay time of 30 ns and a small-sized photomultiplier tube (SPMT). The negative output pulse from the SPMT is amplified by a high-speed inverse amplifier, and the event pulses are sent to a multichannel analyzer to measure X-ray spectra. The energy resolution of the spectrometer was 15% at 59.5 keV. We carried out photon-counting computed tomography using gadolinium media with a maximum rate of 650 kilo counts per second and confirmed the energy-dispersive effect with changes in the description voltage of event pulses using a high-speed comparator.