    Saluveer, Ott; Drca, Nikola; Sahlgren, Bo; Jensen-Urstad, Mats; Braunschweig, Frieder

    JACC. Case reports, 11/2021, Volume: 3, Issue: 16
    Journal Article

    The authors stopped a case of left atrial appendage occlusion because of miscolored brownish devices. The investigation demonstrated that devices may show a range of colors from a typical blue to a brownish hue, depending on oxide layer thickness, and this does not appear to have any impact on the performance of the device. ( Level of Difficulty: Intermediate. )