Akademska digitalna zbirka SLovenije - logo
VSE knjižnice (vzajemna bibliografsko-kataložna baza podatkov COBIB.SI)
  • Preskušanje intramedularne armature pri premostitveni osteosintezi kostnega defekta z metilmetakrilatom = Testing of the intramedullary reinforcement used in osteosynthesis with bone grafting of the bone defect with methyl-metacrylate
    Smrkolj, Vladimir, upokojen ; Banovec, J.
    The purpose of this testing was to determine whether the routinely used intramedullary metallic plate can be replaced by our newly constructed metallic reinforcement ladder without interfering with ... the static properties of the bone alignment. Our results showed that the alignment obtained by the application of the new metallic reinforcement had the same static value as theformerly used technique. The new technique is superior to the former one asconcerns the bending loads due to the greater distance between the external osteosynthesis plate and the reinforcement, and because of the greater friction between the reinforcement and methyl-metacrylate than between the metallic plate and methyl-metacrylate.
    Vrsta gradiva - članek, sestavni del
    Leto - 1990
    Jezik - slovenski
    COBISS.SI-ID - 16152281