We compared the BD MAX enteric bacterial panel (EBP) to culture for the detection of Salmonella, Shigella, Campylobacter, and Shiga toxin-producing enterohemorrhagic Escherichia coli (EHEC) O157 in ...seeded stool samples. The EBP panel demonstrated superior sensitivity and reliably detected Salmonella, EHEC O157, Shigella, and Campylobacter at concentrations 1- to 2-log10 lower than those needed for culture detection.
Multiplex nucleic acid testing is increasingly used to diagnose childhood gastroenteritis. The advantages of this disruptive technology include rapidity, sensitivity, and ability to detect pathogenic ...viruses, bacteria, and parasites simultaneously. The drawbacks are its capacity to identify organisms of uncertain clinical significance in North American children, cost, and inability to provide viable bacteria for strain typing by public health authorities. However, this technology will certainly improve our knowledge of the causes of human gut infections. As data emerge, physicians should interpret results cautiously, and, most important, consider the context of the presentation before making clinical decisions based on the readouts.
Manual treponemal and nontreponemal serologic testing has historically been used for the diagnosis of syphilis. This approach is simple and reproducible but labor intensive. Recently, the FDA cleared ...the fully automated BioPlex 2200 Syphilis Total & RPR assay for the detection of treponemal and nontreponemal antibodies. We evaluated the clinical performance of this assay at a tertiary medical center with a high syphilis prevalence. Prospective consecutively collected (
= 400) and known RPR-positive (
= 100) specimens were compared using predicate manual rapid plasma reagin (RPR) and fluorescent treponemal antibody absorption (FTA) methods and the BioPlex 2200 Syphilis Total & RPR assay. Positive and negative percent agreements (PPA and NPA, respectively) between the assays were calculated. The PPA and NPA between the manual and BioPlex 2200 RPR results for the prospective population were 85% (17/20; 95% confidence interval CI, 69% to 100%) and 98% (373/380; 95% CI, 97% to 99%), respectively. The PPA for the manual RPR-positive population was 88% (88/100; 95% CI, 82% to 94%). Overall, the manual and BioPlex 2200 RPR titers demonstrated 78% (99/127) concordance within ±1 dilution and 94% (120/127) within ±2 dilutions. An interpretation of the syphilis serologic profile using the traditional algorithm showed a concordance of 99.5% in the prospective population and 85% in the manual RPR-positive cohort. The performance of the BioPlex 2200 Syphilis Total & RPR assay is comparable to those of manual methods. The high NPA of this assay combined with the ability to automate a historically labor-intensive assay is an appealing attribute for syphilis screening in a high-volume laboratory.
Diarrheal illness is a major cause of morbidity and mortality throughout the world, yet the etiologic agent of many cases of gastrointestinal illness remains unspecified, often due to the lack of ...convenient, timely, and sensitive diagnostic testing. Although bulk fecal specimens remain the recommended specimen type for enteric culture, rectal swabs may be an option preferred by clinicians and patients due to the convenience and timing of collection. However, the lack of data evaluating the sensitivity of rectal swabs compared to fecal specimens for detection of enteric pathogens precludes this specimen type from being recommended by national guidelines. In this study, we retrospectively reviewed 480 paired rectal swab and fecal specimens submitted for enteric culture to the Barnes-Jewish Hospital and St. Louis Children's Hospital microbiology laboratories in St. Louis, MO, from 2002 to 2017. We report 32% positivity of paired specimens with an overall agreement of 93% and Cohen's κ of 0.84 (95% confidence interval, 0.78 to 0.89). Additionally, we evaluated the time to result from the time of patient presentation to the health care setting and demonstrate that rectal swabs have a significantly shorter time to an actionable result than bulk fecal specimens (median, 67.4 h versus 78.4 h, respectively;
< 0.001). These findings indicate that rectal swabs facilitate on-demand culture-based testing with a sensitivity comparable to that of fecal specimens and thus should be recommended for enteric bacterial culture when bulk fecal specimens are unavailable.
Total laboratory automation (TLA) has the potential to reduce specimen processing time, improve standardization of cultures, and decrease turnaround time (TAT). The objective of this study was to ...perform a detailed interrogation of the impact of TLA implementation in all aspects of the workflow for routine culture of urine specimens. Using a detailed motion capture study, the time required for major steps of processing and result reporting were prospectively assessed for urine samples prior to (
n
= 215) and after (
n
= 203) implementation of the BD Kiestra TLA system. Specimens were plated on all shifts, but cultures were read only during the day shift for both time periods. Significant increases were noted in the time from receipt to inoculation (23.0 min versus 32.0 min,
p
< 0.001) and total processing time (28.0 min versus 66.0 min,
p
< 0.0001) for urine specimens post-TLA. Rates of positive (18.6% versus 16.3%) and negative (71.2% versus 79.3%) urine cultures remained stable through the pre- and post-TLA time periods (
p
= 0.58). There were no changes in TAT for organism identification or susceptibility results. The time to final report was decreased from 43.8 h pre-TLA to 42.0 h post-TLA, which was attributed to significant decreases in TAT for negative cultures (42.0 h versus 37.5 h,
p
= 0.01). These findings demonstrate that changes in laboratory workflow are necessary to maximize efficiency of TLA and optimize TAT.