The Argus II Retinal Prosthesis System (Second Sight Medical Products, Inc, Sylmar, CA) was developed to restore some vision to patients blind as a result of retinitis pigmentosa (RP) or outer ...retinal degeneration. A clinical trial was initiated in 2006 to study the long-term safety and efficacy of the Argus II System in patients with bare or no light perception resulting from end-stage RP.
Prospective, multicenter, single-arm clinical trial. Within-patient controls included the nonimplanted fellow eye and patients' native residual vision compared with their vision with the Argus II.
Thirty participants in 10 centers in the United States and Europe.
The worse-seeing eye of blind patients was implanted with the Argus II. Patients wore glasses mounted with a small camera and a video processor that converted images into stimulation patterns sent to the electrode array on the retina.
The primary outcome measures were safety (the number, seriousness, and relatedness of adverse events) and visual function, as measured by 3 computer-based, objective tests. Secondary measures included functional vision performance on objectively scored real-world tasks.
Twenty-four of 30 patients remained implanted with functioning Argus II Systems at 5 years after implantation. Only 1 additional serious adverse event was experienced after the 3-year time point. Patients performed significantly better with the Argus II on than off on all visual function tests and functional vision tasks.
The 5-year results of the Argus II trial support the long-term safety profile and benefit of the Argus II System for patients blind as a result of RP. The Argus II is the first and only retinal implant to have market approval in the European Economic Area, the United States, and Canada.
Purpose
To evaluate the post‐approval long‐term outcomes of the Argus II Retinal Prosthesis, with a specific focus on its functional visual benefit in patients’ daily activities.
Methods
Eighteen ...patients with bare light perception due to end‐stage retinitis pigmentosa were included in a French prospective, multicentre, single‐arm study and followed for 2 years. Visual benefit in patients’ daily activities was monitored through the use of the Functional Low‐vision Observer Rated Assessment (FLORA), and the final score at 2 years was the primary effectiveness outcome. Standardized visual assessments were also performed. Device‐ or procedure‐related adverse events were recorded.
Results
Seventeen subjects completed the study. Positive impacts of the Argus II system on functional vision and well‐being were demonstrated for over 70% of subjects on the FLORA. Among the daily activities/tasks tested, finding doorways was one of the most statistically significantly improved tasks (p < 0.001), along with estimating the size of an obstacle (p < 0.001), visually locating a place setting on a dining table (p < 0.001) and visually locating people in a non‐crowded setting (p < 0.001). Visual function was improved on most standardized tests. Only two device‐ or procedure‐related serious adverse events were observed (one vitreous haemorrhage and one endophthalmitis, both resolved with treatment). No explantation was required.
Conclusion
This first report of a completed post‐approval study of Argus II with a two‐year follow‐up demonstrates the safety and effectiveness of the Argus II System in a real‐world cohort of patients and further highlights its real functional benefit in implanted patients’ daily activities.
To evaluate the incidence, evolution, clinical characteristics, possible risk factors or preventive factors, and visual outcomes of epiretinal membrane (ERM) recurrence.
Retrospective study of 440 ...consecutive patients (440 eyes) who underwent pars plana vitrectomy for ERM. The internal limiting membrane (ILM) was peeled in 266 cases, with the help of indocyanine green in 27 cases and brilliant blue in 45 cases. Cases of symptomatic ERM recurrence were reoperated.
The incidence of ERM recurrence was 5% (22/440), and 2% of the patients were reoperated (9/440). Epiretinal membrane recurrence was symptomatic in 9 cases (41%) and asymptomatic in 13 cases (59%). ILM peeling was the only factor preventing ERM recurrence (adjusted odds ratio = 0.33, P = 0.026). The use of staining dyes did not prevent recurrence (adjusted odds ratio = 0.35, P = 0.338). In the case of ERM reproliferation, the absence of ILM peeling, the existence of ERM on the fellow eye, and poor visual acuity before surgery seemed to be associated with a high risk of symptomatic recurrence and reoperation. The mean duration for follow-up was 3.5 ± 1.7 years.
ILM peeling not only reduces the likelihood of reproliferation of ERM but also seems to improve the visual prognosis of recurrent ERMs. The use of dyes did not reduce the rate of recurrence compared with when ILM was peeled without dyes.
The purpose of this analysis is to report the change in quality of life (QoL) after treatment with the Argus II Epiretinal Prosthesis in patients with end-stage retinitis pigmentosa.
The Vision and ...Quality of Life Index (VisQoL) was used to assess changes in QoL dimensions and overall utility score in a prospective 30-patient single-arm clinical study. VisQoL is a multi-attribute instrument consisting of six dimensions (injury, life, roles, assistance, activity and friendship) that may be affected by visual impairment. Within each dimension, patients were divided into two groups based on how much their QoL was affected by their blindness at baseline (moderate/severe or minimal). Outcomes were compared within each dimension sub-group between baseline and the combined follow-up periods using the Friedman test. In addition, data from the six dimensions were combined into a single utility score, with baseline data compared to the combined follow-up periods.
Overall, 80 per cent of the patients reported difficulty in one or more dimensions pre-implant. Composite VisQoL utility scores at follow-up showed no statistically significant change from baseline; however, in three of the six VisQoL dimensions (injury, life and roles), patients with baseline deficits showed significant and lasting improvement after implantation with Argus II. In two of the three remaining dimensions (assistance and activity), data trended toward an improvement. In the final VisQoL dimension (friendship), none of the patients reported baseline deficits, suggesting that patients had largely adjusted to this attribute.
Patients whose vision negatively affected them with respect to three VisQoL dimensions (that is, getting injured, coping with the demands of their life and fulfilling their life roles) reported significant improvement in QoL after implantation of the Argus II retinal prosthesis. Furthermore, the benefit did not deteriorate at any point during the 36-month follow-up, suggesting a long-term, durable improvement.
To compare different clinical and Spectral-Domain Optical Coherence Tomography (SD-OCT) features of high myopic eyes with Stickler syndrome (STL) with matched controls.
Patients with genetically ...confirmed STL with axial length ≥ 26 mm and controls matched for axial length were included. The following data were obtained from SD-OCT scans and fundus photography: choroidal and retinal thickness (respectively, CT and RT), peripapillary atrophy area (PAA), presence of posterior staphyloma (PS).
Twenty-six eyes of 17 patients with STL and 25 eyes of 19 controls were evaluated. Compared with controls, patients with STL showed a greater CT subfoveally, at 1000 μm from the fovea at both nasal and temporal location, and at 2000 and 3000 μm from the fovea in nasal location (respectively, 188.7±72.8 vs 126.0±88.7 μm, 172.5±77.7 vs 119.3±80.6 μm, 190.1±71.9 vs 134.9±79.7 μm, 141.3±56.0 vs 98.1±68.5 μm, and 110.9±51.0 vs 67.6±50.7 μm, always P< 0.05). Furthermore, patients with STL showed a lower prevalence of PS (11.5% vs 68%, P< 0.001) and a lower PAA (2.2±2.1 vs 5.4±5.8 mm
, P=0.03), compared with controls.
This study shows that high myopic patients with STL show a greater CT, a lower PAA and a lower prevalence of PS, compared with controls matched for axial length. These findings could be relevant for the development and progression of myopic maculopathy in patients with STL.
Retinal detachment often leads to a severe and permanent loss of vision and its therapeutic management remains to this day exclusively surgical. We have used surgical specimens to perform a ...differential analysis of the transcriptome of human retinal tissues following detachment in order to identify new potential pharmacological targets that could be used in combination with surgery to further improve final outcome.
Statistical analysis reveals major involvement of the immune response in the disease. Interestingly, using a novel approach relying on coordinated expression, the interindividual variation was monitored to unravel a second crucial aspect of the pathological process: the death of photoreceptor cells. Within the genes identified, the expression of the major histocompatibility complex I gene HLA-C enables diagnosis of the disease, while PKD2L1 and SLCO4A1 -which are both down-regulated- act synergistically to provide an estimate of the duration of the retinal detachment process. Our analysis thus reveals the two complementary cellular and molecular aspects linked to retinal detachment: an immune response and the degeneration of photoreceptor cells. We also reveal that the human specimens have a higher clinical value as compared to artificial models that point to IL6 and oxidative stress, not implicated in the surgical specimens studied here.
This systematic analysis confirmed the occurrence of both neurodegeneration and inflammation during retinal detachment, and further identifies precisely the modification of expression of the different genes implicated in these two phenomena. Our data henceforth give a new insight into the disease process and provide a rationale for therapeutic strategies aimed at limiting inflammation and photoreceptor damage associated with retinal detachment and, in turn, improving visual prognosis after retinal surgery.
Celotno besedilo
Dostopno za:
DOBA, IZUM, KILJ, NUK, PILJ, PNG, SAZU, SIK, UILJ, UKNU, UL, UM, UPUK
Purpose
To compare the safety and functional outcomes of 25-gauge and 23-gauge (G) micro-incision vitrectomy surgery (MIVS) instrumentation with the standard 20-G vitrectomy system in the treatment ...of epiretinal membranes (ERM).
Methods
A retrospective comparative study of 553 consecutive cases with epiretinal membrane who underwent pars plana vitrectomy. Twenty-gauge, 25-gauge and 23-gauge vitrectomy was performed respectively in 347, 91, and 115 eyes. Surgery duration, visual acuity improvement, intraocular pressure variation, intraoperative and postoperative complications were analyzed.
Results
The mean surgical time in the 23-G group and in the 25-G group was shorter than in the 20-G group (
P
< 0.001). Visual improvement was higher 8 days postoperatively in the 25-G group than in the 20-G and 23-G groups (
P
= 0.035), but not at 6 weeks postoperatively (
P
= 0.186). In the 20-G group, the IOP increased significantly on the first day postoperatively (
P
< 0.001), while in the 23-G group, the IOP decreased on the first day postoperatively (
P
= 0.073). In the 25-G group, the IOP did not change significantly (
P
= 0.807). The incidence of complications was not statistically significant between the three groups. Retinal breaks were significantly related to the induction of posterior vitreous detachment, independent of the system gauge.
Conclusion
In ERM surgery, 23-G and 25-G (MIVS) systems are as safe and effective as the 20-G system, and significantly reduce surgical time. Although the 25-G system provides an earlier visual improvement, the 23- and 25- gauge systems are comparable, and the selection will depend on the surgeon’s preference.
Purpose
To assess the impact of the Covid-19 pandemic and subsequent lockdown on the number and clinical characteristics of patients with retinal detachment (RD) in a French public university eye ...hospital.
Methods
Single-center, retrospective non-interventional study. Patients consulting at the emergency room (ER) of Quinze-Vingts Hospital (France) for rhegmatogenous RD before and after instauration of the lockdown were reviewed. We compared the characteristics of patients with RD between the containment period (March17th - April27th,2020) and the period preceding the lockdown (February18th - March16th,2020). We compared the number of RD surgeries performed between the first month of lockdown (March17th - April19th,2020) and the corresponding period of 2019. Number of cases, delay between diagnosis and surgery, visual acuity was measured.
Results
During the first month of lockdown, 59 RDs were operated on, compared to 107 in the corresponding period in 2019 (-44,8%). Mean time from first symptoms to surgery was significantly higher during the lockdown 12.7 (11.3) days vs 7.6 (7.8) days (p = 0.031) before. During the lockdown, the mean BCVA was lower albeit the difference did not reach statistical significance (1.16 (0.9) during pre-containment vs 1.5 (0.9) during containment; p = 0.09). Reasonsfor delayed consultation were: fear of Covid-19 (31%; p = 0.0001), absence of referral doctor (31%; p = 0.003) and difficulties in getting to public transport (10.3%;p = 0.859).
Conclusion
Despite maintaining accessto emergency eye care facilitiesin our hospital, the lockdown affected visual health. Should the lockdown be reinstated, we postulate that a better information about eye care access for non-Covid emergencies may attenuate its effect on visual health.
To analyze and provide an overview of the incidence, management, and prevention of conjunctival erosion in Argus II clinical trial subjects and postapproval patients.
This retrospective analysis ...followed the results of 274 patients treated with the Argus II Retinal Prosthesis System between June 2007 and November 2017, including 30 subjects from the US and European clinical trials, and 244 patients in the postapproval phase. Results were gathered for incidence of a serious adverse event, incidence of conjunctival erosion, occurrence sites, rates of erosion, and erosion timing.
Overall, 60% of subjects in the clinical trial subjects versus 83% of patients in the postapproval phase did not experience device- or surgery-related serious adverse events. In the postapproval phase, conjunctival erosion had an incidence rate of 6.2% over 5 years and 11 months. In 55% of conjunctival erosion cases, erosion occurred in the inferotemporal quadrant, 25% in the superotemporal quadrant, and 20% in both. Sixty percent of the erosion events occurred in the first 15 months after implantation, and 85% within the first 2.5 years.
Reducing occurrence of conjunctival erosion in patients with the Argus II Retinal Prosthesis requires identification and minimization of risk factors before and during implantation. Implementing inverted sutures at the implant tabs, use of graft material at these locations as well as Mersilene rather than nylon sutures, and accurate Tenon's and conjunctiva closure are recommended for consideration in all patients.