The analysis of impurities in a uranium ore concentrate (UOC) could provide information regarding the source, production history, and potential intended use of the UOC. This study involves the ...analysis of UOC samples for phosphorus and sulfur. Concentrations were determined by triple quadrupole inductively coupled plasma – mass spectrometry and compared with results from a pyrohydrolysis method as well as previously reported results. The sulfur and phosphorus concentrations, determined by the mass spectrometer, were used to explore possible trends in a series of UOC material, and the uncertainties were calculated using GUM workbench software. The triple quadrupole inductively coupled plasma – mass spectrometer method allows for the removal of interferences in the analysis of species.
Display omitted
•A recent triple quadrupole ICP–MS method for determination of P and S in UOCs.•Ability to measure range of UOC material rapidly without reducing data quality.•Concentration of P and S were similar to traditional methods and/or referenced data.
The analysis of environmental swipe samples for ultra-trace uranium (U) and plutonium (Pu) determinations is essential in the nuclear safeguards community. While mass spectrometry techniques for U ...and Pu detection continually improve, established separation methods are seldom reevaluated. Currently, actinide separations within the forensics community predominantly employ either Eichrom TEVA® or UTEVA® resins. The direct optimization of U and Pu separations utilizing both resins has not been widely reported. Here, several methods were explored with goals of increasing analyte recovery, acquiring cleaner blanks, and improving the separation efficiency of ultra-trace levels of U and Pu from environmental swipe samples. The optimized separation methodology of U and Pu was examined using certified reference materials and archived environmental swipe samples.
Display omitted
•Greater than 98% recovery of U and Pu from TEVA and UTEVA cartridges.•Lower U and Pu process blank levels.•Improved separation of ultra-trace (fg-ng) U and Pu from environmental swipe samples.•Faster separation times with reduced reagent usa.
The ability to directly measure uranium isotope ratios on environmental swipes has been achieved through a solution-based microextraction process and represents a significant advancement toward the ...development of a rapid method to analyze international nuclear safeguard samples. Here, a microextraction probe is lowered and sealed onto the swipe surface, and analytes within the sampling site (∼8 mm2) are dissolved and extracted into a flowing solvent of 2% nitric acid (HNO3). The mobilized species are subsequently directed into an inductively coupled plasma-mass spectrometer (ICP-MS) for accurate and precise isotope ratio determination. This work highlights the novelty of the sampling mechanism, particularly with the direct coupling of the microextraction probe to the ICP-MS and measurement of uranium isotope ratios. The preliminary method detection limit for the microextraction-ICP-MS method, utilizing a quadrupole-based MS, was determined to be ∼50 pg of 238U. Additionally, precise and accurate isotope ratio measurements were achieved on uranium reference materials for both the major (235U/238U) and minor (234U/238U and 236U/238U) ratios. While the present work is focused on directly measuring uranium isotopic systems on swipe surfaces for nuclear safeguards and verification applications, the benefits would extend across many applications in which direct solid sampling is sought for elemental and isotopic analysis.
A fully automated method for the separation of low-concentration uranium from plutonium in environmental swipe samples has been developed. The offline chromatography system features renewable 1 mL ...Eichrom TEVA and UTEVA column generation from bulk resin slurry. Discrete fractions of the separated actinides are delivered into user defined vials for future analysis. Clean room background levels were achieved outside of a cleanroom environment with this method. Purification of uranium and plutonium from various sample matrixes and at various concentrations was successful. Major and minor isotope ratios for both elements were measured via multiple collector inductively coupled plasma mass spectrometry and were in good agreement with certified reference values. Validation of the separation method was conducted on archived environmental samples and agreed with values previously reported using standard column chemistry.
The collection of solid particulates and liquids from surfaces by the use of cloth swipes is fairly ubiquitous. In such methods, there is a continuous concern regarding the ability to locate and ...quantitatively sample the analyte species from the material. In this effort, we demonstrate the initial coupling of an Advion Plate Express plate reader to a liquid sampling-atmospheric pressure glow discharge (LS-APGD) microplasma ionization source with an Orbitrap mass spectrometer to perform uranium isotopic analyses of solution residues on cotton swipes. The Plate Express employs a sampling probe head to engage and seal against the swipe surface. Subsequentially, the analyte residues are desorbed and transported within a 2% HNO
electrolyte flow to the ionization source. Quantitative recoveries were observed following a single 30 s extraction step, with the absolute mass sampled per extraction being ∼100 ng. While the intrasample variability in the analytical responses for triplicate sampling of the same swipe yield ∼30% RSD, this lack of precision is offset by the ability to determine isotope ratios for enriched uranium specimens with a precision of better than 10% RSD. Pooled, intersample precision (
= 9) was found to be <5%RSD across the various sample compositions. Finally,
U/
U determinations (ranging from 0.053 to 1.806) were accurate with errors of <10%, absolute. The
U- and
U-inclusive ratios were determined with similar accuracy in enriched samples. While the driving force for the effort is in the realm of nuclear nonproliferation efforts, the ubiquitous use of cloth swipes across many application areas could benefit from this convenient approach, including the use of versatile, reduced-format mass spectrometer systems.
Automated introduction platforms integrated with inductively coupled plasma optical emission spectroscopy (ICP-OES) systems are continuously being improved. Expanding on the introduction systems, a ...newly developed automated ion chromatography system was explored for performing rapid in-line separations coupled to ICP-OES for the detection of trace elements in uranium. Trace elements are separated from a uranium material and the analytes are directed into the ICP-OES for subsequent detection. Detection parameters such as exposure time frequency, wavelength selection, and settling times were explored to gain insight on optimal detection schemes for in-line trace elemental analysis. The methodology was applied in the analysis of a uranium oxide (U3O8) certified reference material, CRM-124. It was found here that the sensitivity and uncertainty of the technique are greatly affected by how the ICP-OES is employed to collect data. Overall it was determined that faster exposure replicates can provide greater peak resolution with higher fidelity measurements but are limited with respect to the total analysis time (i.e., limited in detection timely separations). Zeta scores, which combine accuracy and uncertainty of certified values and experimental values, were used to validate the ICP-OES modes of operation.
The trace impurities of a uranium ore concentrate (UOC) can be examined to determine mine source, methods of production, and quality. This study presents a method to determine the concentration of ...halides and main group elements, specifically P, S, Br and I, utilizing triple quadrupole inductively coupled plasma–mass spectrometry. These analytes were measured in a uranium matrix to simulate a UOC sample. The concentrations determined with this method showed agreement with known values. Solutions with and without uranium were compared. A UOC certified reference material, CUP-2, was analyzed to further demonstrate the effectiveness of the method.
The present study documents an automated approach to performing elemental analysis on a large group uranium ore concentrate (UOC) samples. Here, 17 UOC samples, 2 quality control samples, and 26 ...process blanks were purified sequentially through a single 500 μL Uranium and TEtra Valent Actinides (UTEVA®) column. For each sample, the trace elemental impurities were separated from its dissolved uranium matrix on the UTEVA column and collected for analysis by inductively coupled plasma – optical emission spectroscopy / triple quadrupole mass spectrometry (ICP-OES/TQMS). The UTEVA column was subsequently regenerated prior to separation of the following sample. The column was efficiently regenerated, for each UOC, even after processing ~50 mg of uranium, cumulatively. The validity of the method was established by determining the trace impurities of two quality control uranium reference samples (CRM 124–1 and CUP-2). The current trace element measurements from the 17 UOC samples were compared to previously reported values from an interlaboratory comparison exercise, when available. The methodology employed here produces trace elemental analysis with excellent correlation to the previously reported data for many of the elements / samples, particularly when viewed through the context of existing geochemical comparisons tools (e.g. chondrite normalized variation plots).
Display omitted
•Automated platform for separation of bulk uranium from trace elemental impurities•Employment of ICP-OES and ICP-MS (specifically triple quadrupole technology) for elemental analysis•Analysis of 17 uranium ore concentrates (UOCs) for their elemental composition
This study demonstrates a method for determining the isotopic composition of low-level (sub-pg) plutonium (Pu) directly from a cotton swipe. Environmental sample (ES) swipes are routinely employed as ...a tool for monitoring activities in nuclear facilities. Traditional ES swipe analysis, as employed in nuclear safeguards, utilizes laborious ashing, digestion, and chemical separation procedures prior to mass spectrometric (MS) analysis. Here, an innovative sample introduction technique employing a microextraction probe to extract Pu directly from the swipe surface is described. The microextraction probe lowers onto the swipe surface, seals on a small area (8 mm
2
), and delivers solvent (2% HNO
3
) to extract actinide material that may be present. The extracted analyte is subsequently directed into a sector field inductively coupled plasma (ICP)-MS for isotope ratio determination. This microextraction-ICP-MS method successfully determined the isotopic composition (
240
Pu/
239
Pu and
242
Pu/
239
Pu) of three Pu certified reference materials (CRM 136, 137, and 138) that were deposited (1 pg) onto ES swipes. The percent relative difference from the certified value, uncorrected for instrumental fractionation, was <2% for the
240
Pu/
239
Pu ratio on all three CRMs and <10% for the
242
Pu/
239
Pu ratio on CRM 136 and 138. The percent relative standard deviation, an estimate of the sample-to-sample isotopic precision, was <4% for the
240
Pu/
239
Pu and <15% for the
242
Pu/
239
Pu. Method limits of detection were determined, based on measurements of an enriched
244
Pu material, to be ∼7 fg. Additionally, a mixed uranium (U) and Pu deposition was made to determine the method's ability to simultaneously extract U and Pu and determine the isotopic composition of both analytes.
This study demonstrates a method for determining the isotopic composition of low-level (sub-pg) plutonium (Pu) directly from a cotton swipe.
Information about elemental and isotopic systematics of ultra-trace level actinides (
e.g.
U and Pu) and main group elements (
e.g.
Ti) present within nuclear grade graphite is vital to the nuclear ...community for improved reactor operation and security. In support of this, extensive effort has been placed on improving analysis methods (
i.e.
, inductively coupled plasma-mass spectrometry). However, significantly less effort has been devoted to the optimization of chemical separation methods. Within the separation community, commercially available Eichrom™ resins are often employed, as their elution characteristics for various elements have been well studied, but the direct optimization of actinides and trace metal separations from a single sample have not been widely investigated. Here, methods using various Eichrom pre-packed cartridges were explored to achieve separation of ultra-trace levels of U, Pu, and Ti from a variety of graphite samples. Once the validity of the combined separation scheme was established using certified reference materials, the method was applied to historic, unirradiated and irradiated, graphite samples. For all samples investigated, precise isotope ratio measurements for the titanium isotope systems were made.
Purification of plutonium, uranium, and titanium from high-purity graphite for high precision isotopic analysis
via
MC-ICP-MS.