We have designed a prospective observational study to analyze the incidence and predictive factors of atrial fibrillation (AF) during a long follow‐up, in a large population. Atrial fibrillation ...episodes were documented by the fallback mode switch (FMS) provided by implanted pacemakers. We have included 377 patients (61% men). The pacing indications were atrioventricular (AV) block (49%), sinus node disease (SND, 16%), bradycardia‐tachycardia syndrome (BTS, 5%), AV block + SND (19%), AV block + BTS (6%), and BTS + SND (5%). The mean age at implant was 75 ± 12 (range 28–95). Atrial fibrillation before inclusion was documented in 10% of patients. Drug therapy at first follow‐up included beta‐adrenergic blockers (17% of the patients), amiodarone (13%), and others (16%). The mean follow‐up was 30 ± 24 weeks. At least one AF episode was stored during follow‐up in the memory of 169 pacemakers (45%). Among patients without history of AF at implant, 46% had documented FMS during follow‐up. Patients with AF received more antiplatelet medications than patients without AF (P = 0.03). In patients with AF, New York Heart Association functional class was slightly higher, amiodarone and sotalol were more often prescribed, and the proportion of hypertension was higher than in patients without AF. However, these trends were not statistically significant. A significant higher incidence of premature atrial beats was observed in patients with AF than patients without AF (P < 0.0002). Patients with AF had a lower atrial percentage of paced events (55%) than patients without AF (63%, P < 0.02). These preliminary results confirm the high incidence of AF in paced patients and suggest a preventive effect of atrial pacing. The effects of other clinical variables may be confirmed with a longer follow‐up in a larger population.
It is possible to characterize some of the sensing functions of new multisite pacing systems by resurrecting the split format in the third position of the standard pacemaker code. This approach ...permits accurate representation of the horizontal and vertical triggering functions of multisite dual-chamber pacemakers without creating a new code.
The minute ventilation is known to be one of the most physiological indicators of exercise. A curvilinear relationship between VE and the normal sinus rhythm (NSR) has been demonstrated in healthy ...patients. The aim of this study is to show that a pacemaker based on a VE sensor can reproduce such a relationship. Eighty‐one patients received a Talent DR 213 (ELA Medical, Montrouge, France) pacemaker with a third‐generation rate responsive algorithm. At 1‐month follow‐up, the patients underwent a treadmill exercise test, after which three groups were defined: group 1 had 6 patients who were 100% paced throughout the exercise test; group 2 had 10 patients who maintained NSR throughout the test; and group 3 had 12 patients who had cardiopulmonary recording during the exercise test. In group 1 patients, the simulation function computed the simulated rate (sim‐rate), which was compared to the sensor‐driven rate (SDR). In group 2 patients, sim‐rate was compared to the NSR. In group 3 patients, cardiac and metabolic reserves were compared to determine the appropriateness of the rate response to exercise (HRR% vs MR%). The results showed that the mean correlation coefficient between sim‐rate and SDR was 0.983 ± 0.005 (P < 0.001); the mean correlation coefficient between NSR and SDR was 0.92 ± 0.07 (P < 0.001); and a linear relationship was found between HRR% and MR%, with a mean slope of 1.1 ± 0.2 that was significantly equal to the theoretical value of 1 (P = NS). In conclusion, combining an activity‐driven sensor with a physiological sensor allows the preservation of a physiological rate response during exercise.
Stored data in implantable pacemakers have rarely been used as a diagnostic tool because of the complexity. Our group bas developed software called AIDA, providing an automatic interpretation of data ...stored in memories of the Chorus (ELA medical) pacemaker. We com pared the results of AIDA analysis to surface ECG Holter interpretation in 59 patients (age 75 ± 9 years). In 33 cases, neither AIDA nor the Holter found any anomaly. Eleven patients demonstrated episodes of supraventricular tachycardia (SVT), confirmed by AIDA in ten patients; AIDA failure was due to nonsustained episodes of SVT not inducing mode switch. Loss of atrial sensing, pacemaker‐mediated tachycar dia, and ventricular extrasystoles were detected by AIDA in ten patients. Traditional Holter missed three cases. This initial study confirms that stored pacemaker data, automatically interpreted can provide reli able information over a 24‐hour period.
Special software allowing the memorization of 24‐hour minute ventilation can be loaded into the memoiy of the Chorus RM, a DDDR pacemaker driven by minute ventilation. This feature was tested in the ...postimplant period in 13 patients. Measurements of minute ventiiation, respiratory rate, and respiratory amplitude were analyzed according to prospectively defined diurnal and nocturnal time periods. Minute ventilation decreased by 39% (P < 0.001) from the diurnal to the nocturnal phase, while respiratory rate and amplitude decreased by 18% and 28%, respectively (P < 0.001 each). Thus, minute ventilation allowed discrimination between sleep and waking hours. This information could be utilized to modulate the backup rate of the pacemaker.
Mode switching algorithms are commonly used to protect the ventricles against high rates induced by atrial tachycardia. In the case of atrial fibrillation (AF), the response of these algorithms ...depends on the quality of atrial sensing. The Chorum 7234 DDDR pacemaker uses a new mode switching algorithm, based on a statistical analysis of the atrial rhythm. It includes two criteria of diagnosis: "high" if more than 28 of 32 cycles are abnormally accelerated; and "low" if more than 36 of 64 cycles are abnormally accelerated.
From a taped database of electrophysiological studies, episodes of AF lasting more than 2 minutes were selected. A tape recorder replayed the atrial signals into an external Chorum device. Each episode was replayed eight times with a programmed atrial sensitivity increasing from 0.4-2.0 mV. For each criterion of diagnosis and each programmed sensitivity, the percentage of atrial sensing, the time to switching, and the mean ventricular rate were measured. Ten episodes of AF from 10 patients (9 men and 1 woman; ages 62 +/- 16 years) were included: 1.95 +/- 0.97 mV and 196 +/- 64 ms. The sensitivity of the algorithm to diagnose atrial tachycardia reached 100%, for an atrial sensitivity set between 0.4 and 1.0 mV. The mean percentages of atrial sensed events were 74% +/- 18% and 46% +/- 9% for the "high" and "low" criteria, respectively. The mean diagnostic times were 28 +/- 26 seconds and 68 +/- 27 seconds, respectively. Sensing of < 23% of AF events resulted in failure to diagnose the arrhythmias by both algorithms. In the event of diagnostic failure, the mean ventricular pacing rate was 79 +/- 9 ppm.
Up to an atrial sensitivity of 1 mV, 100% of AF episodes were diagnosed. The Chorum mode switching algorithms are 100% reliable if > 45% of the AF waves are sensed. In the event of switching failure, the ventricle is protected by an average rate remaining below 80 ppm.
In contrast to testing of the capture threshold, the reliability of sensing tests has been little studied. This study was performed to test the automatic sensing algorithm included in the Biotronik ...pacing systems. The automatic measurements made by the devices were compared with the direct manual measurements made of 271 atrial (72%) or ventricular (28%) electrograms recorded in ten patients. A high correlation(r = 0.995, P < 0.0001) was found between the two types of measurements. The use of this function should facilitate the verification of accurate sensing during the long‐term follow‐up of pacemaker patients, and offers a tool to analyze the variations in amplitude of intracardiac signals.