Objective: The treatment of choice for aortic valve insufficiency due to root dilatation has become root replacement with aortic valve sparing. However, root replacement with a synthetic graft may ...result in altered valve stresses. The purpose of this study was to compare the stress/strain patterns in the spared aortic valve in different root replacement procedures by means of finite element modeling.
Methods: Our finite element model of the normal human root and valve was modified to simulate and evaluate three surgical techniques: (1) “cylindrical” graft sutured below the valve at the anulus, (2) “tailored” graft sutured just above the valve, and (3) “pseudosinus” graft, tailored and sutured below the valve at the anulus. Simulated diastolic pressures were applied, and stresses and strains were calculated for the valve, root, and graft. Leaflet coaptation was also quantified.
Results: All three root replacement models demonstrated significantly altered leaflet stress patterns as compared with normal patterns. The cylindrical model showed the greatest increases in stress (16%-173%) and strain (10%-98%), followed by the tailored model (stress +10%-157%, strain +9%-36%). The pseudosinus model showed the smallest increase in stress (9%-28%) and strain (2%-31%), and leaflet coaptation was closest to normal.
Conclusion: Valve-sparing techniques that allow the potential for sinus space formation (tailored, pseudosinus) result in simulated leaflet stresses that are closer to normal than the cylindrical technique. Normalized leaflet stresses in the clinical setting may result in improved longevity of the spared valve. (J Thorac Cardiovasc Surg 2000;119:753-63)
Impaired renal function after coronary artery bypass graft (CABG) surgery is a key risk factor for in-hospital mortality. However, perioperative increases in serum creatinine and the association with ...mortality has not been well-studied. We assessed the hypothesis that perioperative increases in creatinine are associated with increased 90-day mortality.
We studied 1391 patients in northern New England undergoing CABG in 2001 and evaluated preoperative and postoperative creatinine. Patients with preoperative dialysis were excluded. Data were linked to the National Death Index to assess 90-day survival. Kaplan-Meier and log-rank techniques were used. Patients were stratified by percent increase in creatinine from baseline: <25%, 25% to 49%, 50% to 99%, > or =100%. We assessed 90-day survival and calculated adjusted hazard ratios (HR) and 95% confidence intervals (95% CI) for creatinine groups, adjusting for age and sex. Patients with the largest creatinine increases (50% to 99% or > or =100%) had significantly higher 90-day mortality compared with patients with a smaller increase (<50%; P<0.001). Adjusted HR and 95% CI confirmed patients in the higher 2 groups had an increased risk of mortality compared with the <25% (referent); however, the 25% to 49% group was not different from the referent: 1.80 (95% CI: 0.73 to 4.44), 6.57 (95% CI, 3.03 to 14.27), and 22.10 (95% CI, 11.25 to 43.39).
Patients with large creatinine increases (> or = 50%) after CABG surgery have a higher 90-day mortality compared with patients with small increases. Efforts to identify patients with impaired renal function and to preserve renal function before cardiac surgery may yield benefits for patients in the future.
The diversity of mitral valve (MV) geometries and multitude of surgical options for correction of MV diseases necessitates the use of computational modeling. Numerical simulations of the MV would ...allow surgeons and engineers to evaluate repairs, devices, procedures, and concepts before performing them and before moving on to more costly testing modalities. Constructing, tuning, and validating these models rely upon extensive in vitro characterization of valve structure, function, and response to change due to diseases. Micro-computed tomography (
μ
CT) allows for unmatched spatial resolution for soft tissue imaging. However, it is still technically challenging to obtain an accurate geometry of the diastolic MV. We discuss here the development of a novel technique for treating MV specimens with glutaraldehyde fixative in order to minimize geometric distortions in preparation for
μ
CT scanning. The technique provides a resulting MV geometry which is significantly more detailed in chordal structure, accurate in leaflet shape, and closer to its physiological diastolic geometry. In this paper, computational fluid–structure interaction (FSI) simulations are used to show the importance of more detailed subject-specific MV geometry with 3D chordal structure to simulate a proper closure validated against
μ
CT images of the closed valve. Two computational models, before and after use of the aforementioned technique, are used to simulate closure of the MV.
The objective was to determine the influence of post-fabrication aging (2, 14, 21, 42, and 63days) on beef quality characteristics and consumer sensory perception of gluteus medius (GM) and ...longissimus lumborum (LL) steaks. Lipid oxidation and aerobic plate counts increased (P<0.05) with longer aging periods and retail display times. An aging period by day of retail display interaction (P<0.05) was observed for a* and b* values for both muscles and L* values for the LL. Warner–Bratzler shear force values decreased (P<0.05) with longer aging for the LL, while no difference was observed for the GM. Consumer panel results demonstrated that longer aging periods increased (P<0.05) tenderness of both muscles. Our results indicate that extended aging reduces retail color stability yet has positive effects on consumer perception of tenderness of beef loin muscles.
•USDA Select beef strip loin and top sirloin butts were aged for up to 63days.•Consumer perception of strip loin tenderness improves up to 14days of aging.•Perception of top sirloin tenderness is optimized between 21 and 42days of aging.•Extended aging reduces retail shelf-life of beef strip loin and top sirloin butts.
Abstract Glioblastoma patients have a poor prognosis, even after surgery, radiotherapy, and chemotherapy with temozolomide or 1,3-bis(2-chloroethy)-1-nitrosourea. We developed an in vitro recovery ...model using neurosphere cultures to analyze the efficacy of chemotherapy treatments, and tested whether glioblastoma neurosphere-initiating cells are resistant. Concentrations of chemotherapy drugs that inhibit neurosphere formation are similar to clinically relevant doses. Some lines underwent a transient cell cycle arrest and a robust recovery of neurosphere formation. These results indicate that glioblastoma neurospheres can regrow after treatment with chemotherapy drugs. This neurosphere recovery assay will facilitate studies of chemo-resistant subpopulations and methods to enhance glioblastoma therapy.
The asymmetry of the aortic valve and aortic root may influence their biomechanics, yet was not considered in previous valve models. This study developed an anatomically representative model to ...evaluate the regional stresses of the valve within the root environment. A finite-element model was created from magnetic-resonance images of nine human valve-root specimens, carefully preserving their asymmetry. Regional thicknesses and anisotropic material properties were assigned to higher-order elastic shell elements representing the valve and root. After diastolic pressurization, peak principal stresses were evaluated for the right, left, and noncoronary leaflets and root walls. Valve stresses were highest in the noncoronary leaflet (538 kPa vs right 473 kPa vs left 410 kPa); peak stresses were located at the free margin and belly near the coaptation surfaces (averages 537 and 482 kPa for all leaflets, respectively). Right and noncoronary sinus stresses were 21% and 10% greater than the left sinus. In all sinuses, stresses near the annulus were higher than near the sinotubular junction. Stresses vary across the valve and root, likely due to their inherent morphologic asymmetry and stress sharing. These factors may influence bioprosthetic valve durability and the incidence of isolated sinus dilatation.
Concomitant aortic (AV) and mitral (MV) valve surgery accounts for 4% of all valve procedures in northern New England. We examined in-hospital and long-term mortality.
This is a report of a ...prospective study of 1057 patients undergoing concomitant AV and MV surgery from 1989 to 2007. The Social Security Administration Death Master File was used to assess long-term survival. Kaplan-Meier and log-rank tests were performed. In-hospital mortality was 15.5% (11.0% for patients <70 years, 18.0% for 70- to 79-year-olds, and 24% for those > or =80 years). Overall median survival was 7.3 years. Median survival without coronary artery bypass grafting was 9.5 years and with coronary artery bypass grafting was 5.7 years (P<0.001). Survival in women was worse than in men (7.3 versus 9.3, years, P=0.033). Median survival by age was 11.0 years for patients <70 years, 5.4 years for 70- to 79-year-olds, and 4.8 years for those > or =80 years. Median survival was not significantly different for patients > or =80 years compared with those who were 70 to 79 years old (P=0.245).
Double-valve surgery has a high in-hospital mortality rate and a median survival of 7.3 years. After patients have survived surgery, long-term survival is similar between men and women, smaller and larger patients, and those receiving MV repair or replacement. Survival continues to decline after surviving surgery for patients > or =70 years old and those who undergo concomitant coronary artery bypass grafting. In patients <70 years, either mechanical valves in both positions or a tissue AV and mitral repair have the lowest in-hospital mortality and the best long-term survival. In patients > or =70 years, tissue valves in both positions have the best in-hospital and long-term survival.
The objective was to determine the influence of post-fabrication aging (2, 14, 21, 42, and 63days) on beef quality characteristics and consumer sensory perception of biceps femoris (BF) and ...semimembranosus (SM) steaks. Lipid oxidation and aerobic plate counts increased (P<0.05) with longer aging periods and retail display times. An aging period by day of retail display interaction (P<0.05) was observed for a* and b* values of the BF and SM. Warner–Bratzler shear force values decreased (P<0.05) with longer aging for the SM, while no difference was observed for the BF. Consumer panel results revealed that longer aging periods increased (P<0.05) acceptability of the SM, tenderness of both muscles, and tended to increase (P=0.07) juiciness of the SM. Our results show that extended aging reduces retail color stability yet has positive effects on consumer perception of tenderness of both muscles and overall acceptability of the SM.
•USDA Select beef bottom round and top rounds were aged for up to 63days.•Aging for 14 to 21days optimizes consumer perception of bottom round tenderness.•Perception of top round tenderness is optimized between 21 and 42days of aging.•Extended aging reduces retail shelf-life of the biceps femoris and semimembranosus.
Background. Idiopathic root dilatation often results in dysfunction of an otherwise normal aortic valve. To examine the effect of root dilatation on leaflet stress, strain, and coaptation, we ...utilized a finite element model.
Methods. The normal model incorporated the geometry, tissue thickness, stiffness, and collagen fiber alignment of normal human roots and valves. We evaluated four dilatation models in which diameters of the aortic root were dilated by 5%, 15%, 30%, and 50%. Regional stress and strain were evaluated and leaflet coaptation percent was calculated under diastolic pressure.
Results. Root dilatation significantly increased regional leaflet stress and strain beyond that found in the normal model. Stress increases ranged from 57% to 399% and strain increases ranged from 39% to 189% in the 50% dilatation model. Leaflet stress and strain were disproportionately high at the attachment edge and coaptation area. Leaflet coaptation was decreased by 18% in the 50% root dilatation model.
Conclusions. Idiopathic root dilatation significantly increases leaflet stress and strain and reduces coaptation in an otherwise normal aortic valve. These alterations may affect valve-sparing aortic root replacement procedures.