The article reports results from the largest international dose survey in paediatric computed tomography (CT) in 32 countries and proposes international diagnostic reference levels (DRLs) in terms of ...computed tomography dose index (CTDI vol) and dose length product (DLP). It also assesses whether mean or median values of individual facilities should be used. A total of 6115 individual patient data were recorded among four age groups: <1 y, >1-5 y, >5-10 y and >10-15 y. CTDIw, CTDI vol and DLP from the CT console were recorded in dedicated forms together with patient data and technical parameters. Statistical analysis was performed, and international DRLs were established at rounded 75th percentile values of distribution of median values from all CT facilities. The study presents evidence in favour of using median rather than mean of patient dose indices as the representative of typical local dose in a facility, and for establishing DRLs as third quartile of median values. International DRLs were established for paediatric CT examinations for routine head, chest and abdomen in the four age groups. DRLs for CTDI vol are similar to the reference values from other published reports, with some differences for chest and abdomen CT. Higher variations were observed between DLP values, based on a survey of whole multi-phase exams. It may be noted that other studies in literature were based on single phase only. DRLs reported in this article can be used in countries without sufficient medical physics support to identify non-optimised practice. Recommendations to improve the accuracy and importance of future surveys are provided.
Purpose:
IGRT has become an indispensable tool in modern radiotherapy with kV imaging used in many departments due to superior image quality and lower dose when compared to MV imaging. Many ...departments use manufacturer supplied protocols for imaging which are not always optimised between image quality and radiation dose (ALARA).
Methods:
Whole body phantom PBU-50 (Kyoto Kagaku ltd., Japan) for imaging in radiology has been imaged on Varian iX accelerator (Varian Medical Systems, USA) with OBI 1.5 system. Manufacturer’s default protocols were adapted by modifying kV and mAs values when imaging different anatomical regions of the phantom (head, thorax, abdomen, pelvis, extremities). Images with different settings were independently reviewed by two persons and their suitability for IGRT set-up correction protocols were evaluated. The suitable images with the lowest mAs were then selected. The entrance surface dose (ESD) for manufacturer’s default protocols and modified protocols were measured with RTI Black Piranha (RTI Group, Sweden) and compared. Image quality was also measured with kVQC phantom (Standard Imaging, USA) for different protocols. The modified protocols have been applied for clinical work.
Results:
For most cases optimized protocols reduced the ESD on average by a factor of 3(range 0.9–8.5). Further reduction in ESD has been observed by applying bow-tie filter designed for CBCT. The largest reduction in dose (12.2 times) was observed for Thorax lateral protocol. The dose was slightly increased (by 10%) for large pelvis AP protocol.
Conclusion:
Manufacturer’s default IGRT protocols could be optimised to reduce the ESD to the patient without losing the necessary image quality for patient set-up correction. For patient set-up with planar kV imaging the bony anatomy is mostly used and optimization should focus on this aspect. Therefore, the current approach with anthropomorphic phantom is more advantageous in optimization over standard kV quality control phantoms and SNR metrics.
The aims of this study were to investigate the frequency of computed tomography (CT) examinations for paediatric patients below 15 y of age in 128 CT facilities in 28 developing countries of Africa, ...Asia and Eastern Europe and to assess the magnitude of CT doses. Radiation dose data were available from 101 CT facilities in 19 countries. The dose assessment was performed in terms of weighted CT dose index (CTDIw), volume CT index and dose length product (DLP) for chest, chest (high resolution), lumbar spine, abdomen and pelvis CT examinations using standard methods. The results show that on average the frequency of paediatric CT examinations was 20, 16 and 5 % of all CT examinations in participating centres in Africa, Asia and Eastern Europe, respectively. Eleven CT facilities in six countries were found to use adult CT exposure parameters for paediatric patients, thus indicating limited awareness and the need for optimisation. CT images were of adequate quality for diagnosis. The CTDIw variations ranged up to a factor of 55 (Africa), 16.3 (Asia) and 6.6 (Eastern Europe). The corresponding DLP variations ranged by a factor of 10, 20 and 8, respectively. Generally, the CTDIw and DLP values in Japan are lower than the corresponding values in the three regions in this study. The study has indicated a stronger need in many developing countries to justify CT examinations in children and their optimisation. Awareness, training and monitoring of radiation doses is needed as a way forwards.