In this randomized trial involving patients with acute ischemic stroke with proximal anterior circulation occlusion, endovascular treatment with a Solitaire stent retriever was more effective than ...medical therapy alone in improving functional outcomes at 90 days.
Recently completed prospective, randomized trials involving patients with acute stroke have consistently shown a clinical benefit for mechanical thrombectomy.
1
–
4
Our study, called the Randomized Trial of Revascularization with Solitaire FR Device versus Best Medical Therapy in the Treatment of Acute Stroke Due to Anterior Circulation Large Vessel Occlusion Presenting within Eight Hours of Symptom Onset (REVASCAT), shares the following four features with the previously cited trials: enrollment limited to patients with imaging-based evidence of proximal occlusion of the M1 segment (main trunk) of the middle cerebral artery with or without concomitant occlusion of the internal carotid artery, imaging-based exclusion . . .
Purpose:
Monte Carlo methods were used to generate lightly filtered high resolution x-ray spectra spanning from 20 kV to 640 kV.
Methods:
X-ray spectra were simulated for a conventional tungsten ...anode. The Monte Carlo N-Particle eXtended radiation transport code (MCNPX 2.6.0) was used to produce 35 spectra over the tube potential range from 20 kV to 640 kV, and cubic spline interpolation procedures were used to create piecewise polynomials characterizing the photon fluence per energy bin as a function of x-ray tube potential. Using these basis spectra and the cubic spline interpolation, 621 spectra were generated at 1 kV intervals from 20 to 640 kV. The tungsten anode spectral model using interpolating cubic splines (TASMICS) produces minimally filtered (0.8 mm Be) x-ray spectra with 1 keV energy resolution. The TASMICS spectra were compared mathematically with other, previously reported spectra.
Results:
Using pairedt-test analyses, no statistically significant difference (i.e., p > 0.05) was observed between compared spectra over energy bins above 1% of peak bremsstrahlung fluence. For all energy bins, the correlation of determination (R
2) demonstrated good correlation for all spectral comparisons. The mean overall difference (MOD) and mean absolute difference (MAD) were computed over energy bins (above 1% of peak bremsstrahlung fluence) and over all the kV permutations compared. MOD and MAD comparisons with previously reported spectra were 2.7% and 9.7%, respectively (TASMIP), 0.1% and 12.0%, respectively R. Birch and M. Marshall, “Computation of bremsstrahlung x-ray spectra and comparison with spectra measured with a Ge(Li) detector,” Phys. Med. Biol.
24, 505–517 (1979), 0.4% and 8.1%, respectively (Poludniowski), and 0.4% and 8.1%, respectively (AAPM TG 195). The effective energy of TASMICS spectra with 2.5 mm of added Al filtration ranged from 17 keV (at 20 kV) to 138 keV (at 640 kV); with 0.2 mm of added Cu filtration the effective energy was 9 keV at 20 kV and 169 keV at 640 kV.
Conclusions:
Ranging from 20 kV to 640 kV, 621 x-ray spectra were produced and are available at 1 kV tube potential intervals. The spectra are tabulated at 1 keV intervals. TASMICS spectra were shown to be largely equivalent to published spectral models and are available in spreadsheet format for interested users by emailing the corresponding author (JMB).
Purpose:
Current dosimetry methods in mammography assume that the breast is comprised of a homogeneous mixture of glandular and adipose tissues. Three‐dimensional (3D) dedicated breast CT (bCT) data ...sets were used previously to assess the complex anatomical structure within the breast, characterizing the statistical distribution of glandular tissue in the breast. The purpose of this work was to investigate the effect of bCT‐derived heterogeneous glandular distributions on dosimetry in mammography.
Methods:
bCT‐derived breast diameters, volumes, and 3D fibroglandular distributions were used to design realistic compressed breast models comprised of heterogeneous distributions of glandular tissue. The bCT‐derived glandular distributions were fit to biGaussian functions and used as probability density maps to assign the density distributions within compressed breast models. The MCNPX 2.6.0 Monte Carlo code was used to estimate monoenergetic normalized mean glandular dose “DgN(E)” values in mammography geometry. The DgN(E) values were then weighted by typical mammography x‐ray spectra to determine polyenergetic DgN (pDgN) coefficients for heterogeneous (pDgNhetero) and homogeneous (pDgNhomo) cases. The dependence of estimated pDgN values on phantom size, volumetric glandular fraction (VGF), x‐ray technique factors, and location of the heterogeneous glandular distributions was investigated.
Results:
The pDgNhetero coefficients were on average 35.3% (SD, 4.1) and 24.2% (SD, 3.0) lower than the pDgNhomo coefficients for the Mo–Mo and W–Rh x‐ray spectra, respectively, across all phantom sizes and VGFs when the glandular distributions were centered within the breast phantom in the coronal plane. At constant breast size, increasing VGF from 7.3% to 19.1% lead to a reduction in pDgNhetero relative to pDgNhomo of 23.6%–27.4% for a W–Rh spectrum. Displacement of the glandular distribution, at a distance equal to 10% of the compressed breast width in the superior and inferior directions, resulted in a 37.3% and a −26.6% change in the pDgNhetero coefficient, respectively, relative to the centered distribution for the Mo–Mo spectrum. Lateral displacement of the glandular distribution, at a distance equal to 10% of the compressed breast width, resulted in a 1.5% change in the pDgNhetero coefficient relative to the centered distribution for the W–Rh spectrum.
Conclusions:
Introducing bCT‐derived heterogeneous glandular distributions into mammography phantom design resulted in decreased glandular dose relative to the widely used homogeneous assumption. A homogeneous distribution overestimates the amount of glandular tissue near the entrant surface of the breast, where dose deposition is exponentially higher. While these findings are based on clinically measured distributions of glandular tissue using a large cohort of women, future work is required to improve the classification of glandular distributions based on breast size and overall glandular fraction.
Abstract Background We aimed to evaluate health-related quality of life (QOL), dyspnea and functional exercise capacity during the year following the diagnosis of a first episode of pulmonary ...embolism. Methods Prospective multicenter cohort study of 100 patients with acute pulmonary embolism recruited at 5 Canadian hospitals from 2010-2013. We measured the outcomes QOL (by SF-36 and PEmb-QOL measures), dyspnea (by the University of California San Diego Shortness of Breath Questionnaire (SOBQ)) and six-minute walk distance at Baseline, 1, 3, 6, and 12 months after acute pulmonary embolism. CT pulmonary angiography was performed at baseline, echocardiogram was performed within 10 days, and cardiopulmonary exercise testing was performed at 1 and 12 months. Predictors of change in QOL, dyspnea, and six-minute walk distance were assessed by repeated measures mixed effects models analysis. Results Mean age was 50.0 years, 57% were male, and 80% were treated as out-patients. Mean scores for all outcomes improved during 1 year follow-up: from baseline to 12 months, mean SF-36 physical component score improved by 8.8 points, SF-36 mental component score by 5.3 points, PEmb-QoL by -32.1 points, and SOBQ by -16.3 points, and six-minute walk distance improved by 40 m. Independent predictors of reduced improvement over time were female sex, higher BMI and percent-predicted VO2 peak <80% on 1 month cardiopulmonary exercise test for all outcomes; prior lung disease and higher pulmonary artery systolic pressure on 10-day echocardiogram for the outcomes SF-36 physical component score and dyspnea score; and higher main pulmonary artery diameter on baseline CT pulmonary angiography for the outcome PEmb-QoL score. Conclusions On average, QOL, dyspnea, and walking distance improve during the year after pulmonary embolism. However, a number of clinical and physiological predictors of reduced improvement over time were identified, most notably female sex, higher BMI and exercise limitation on 1- month cardiopulmonary exercise test. Our results provide new information on patient-relevant prognosis after pulmonary embolism.
Purpose
The purpose of this work was to develop and make available x‐ray spectra for some of the most widely used digital mammography (DM), breast tomosynthesis (BT), and breast CT (bCT) systems in ...North America.
Methods
The Monte Carlo code MCNP6 was used to simulate minimally filtered (only beryllium) x‐ray spectra at 8 tube potentials from 20 to 49 kV for DM/BT, and 9 tube potentials from 35 to 70 kV for bCT. Vendor‐specific anode compositions, effective anode angles, focal spot sizes, source‐to‐detector distances, and beryllium filtration were simulated. For each 0.5 keV energy bin in all simulated spectra, the fluence was interpolated using cubic splines across the range of simulated tube potentials to produce spectra in 1 kV increments from 20 to 49 kV for DM/BT and from 35 to 70 kV for bCT. The HVL of simulated spectra with conventional filtration (at 35 kV for DM/BT and 49 kV for bCT) was used to assess spectral differences resulting from variations in: (a) focal spot size (0.1 and 0.3 mm IEC), (b) solid angle at the detector (i.e., small and large FOV size), and (c) geometrical specifications for vendors that employ the same anode composition.
Results
Averaged across all DM/BT vendors, variations in focal spot and FOV size resulted in HVL differences of 2.2% and 0.9%, respectively. Comparing anode compositions separately, the HVL differences for Mo (GE, Siemens) and W (Hologic, Philips, and Siemens) spectra were 0.3% and 0.6%, respectively. Both the commercial Koning and prototype “Doheny” (UC Davis) bCT systems utilize W anodes with a 0.3 mm focal spot. Averaged across both bCT systems, variations in FOV size resulted in a 2.2% difference in HVL. In addition, the Koning spectrum was slightly harder than Doheny with a 4.2% difference in HVL. Therefore to reduce redundancy, a generic DM/BT system and a generic bCT system were used to generate the new spectra reported herein. The spectral models for application to DM/BT were dubbed the Molybdenum, Rhodium, and Tungsten Anode Spectral Models using Interpolating Cubic Splines (MASMICSM‐T, RASMICSM‐T, and TASMICSM‐T; subscript “M‐T” indicating mammography and tomosynthesis). When compared against reference models (MASMIPM, RASMIPM, and TASMIPM; subscript “M” indicating mammography), the new spectral models were in close agreement with mean differences of 1.3%, −1.3%, and −3.3%, respectively, across tube potential comparisons of 20, 30, and 40 kV with conventional filtration. TASMICSbCT‐generated bCT spectra were also in close agreement with the reference TASMIP model with a mean difference of −0.8%, across tube potential comparisons of 35, 49, and 70 kV with 1.5 mm Al filtration.
Conclusions
The Mo, Rh, and W anode spectra for application in DM and BT (MASMICSM‐T, RASMICSM‐T, and TASMICSM‐T) and the W anode spectra for bCT (TASMICSbCT) as described in this study should be useful for individuals interested in modeling the performance of modern breast x‐ray imaging systems including dual‐energy mammography which extends to 49 kV. These new spectra are tabulated in spreadsheet form and are made available to any interested party.
Purpose
Clinical use of dedicated breast computed tomography (bCT) requires relatively short scan times necessitating systems with high frame rates. This in turn impacts the x‐ray tube operating ...range. We characterize the effects of tube voltage, beam filtration, dose, and object size on contrast and noise properties related to soft tissue and iodine contrast agents as a way to optimize imaging protocols for soft tissue and iodine contrast at high frame rates.
Methods
This study design uses the signal‐difference‐to‐noise ratio (SDNR), noise‐equivalent quanta (NEQ), and detectability (d´) as measures of imaging performance for a prototype breast CT scanner that utilizes a pulsed x‐ray tube (with a 4 ms pulse width) at 43.5 fps acquisition rate. We assess a range of kV, filtration, breast phantom size, and mean glandular dose (MGD). Performance measures are estimated from images of adipose‐equivalent breast phantoms machined to have a representative size and shape of small, medium, and large breasts. Water (glandular tissue equivalent) and iodine contrast (5 mg/ml) were used to fill two cylindrical wells in the phantoms.
Results
Air kerma levels required for obtaining an MGD of 6 mGy ranged from 7.1 to 9.1 mGy and are reported across all kV, filtration, and breast phantom sizes. However, at 50 kV, the thick filters (0.3 mm of Cu or Gd) exceeded the maximum available mA of the x‐ray generator, and hence, these conditions were excluded from subsequent analysis. There was a strong positive association between measurements of SDNR and d’ (R2 > 0.97) within the range of parameters investigated in this work. A significant decrease in soft tissue SDNR was observed for increasing phantom size and increasing kV with a maximum SDNR at 50 kV with 0.2 mm Cu or 0.2 mm Gd filtration. For iodine contrast SDNR, a significant decrease was observed with increasing phantom size, but a decrease in SDNR for increasing kV was only observed for 70 kV (50 and 60 kV were not significantly different). Thicker Gd filtration (0.3 mm Gd) resulted in a significant increase in iodine SDNR and decrease in soft tissue SDNR but requires significantly more tube current to deliver the same MGD.
Conclusions
The choice of 60 kV with 0.2 mm Gd filtration provides a good trade‐off for maximizing both soft tissue and iodine contrast. This scanning technique takes advantage of the ~50 keV Gd k‐edge to produce contrast and can be achieved within operating range of the x‐ray generator used in this work. Imaging at 60 kV allows for a greater range in dose delivered to the large breast sizes when uniform image quality is desired across all breast sizes. While imaging performance metrics (i.e., detectability index and SDNR) were shown to be strongly correlated, the methodologies presented in this work for the estimation of NEQ (and subsequently d') provides a meaningful description of the spatial resolution and noise characteristics of this prototype bCT system across a range of beam quality, dose, and object sizes.
The studies by Sahbaee et al were well performed, and the results are provocative, but the authors of this editorial suggest that there are limitations to all modeling studies and that the results ...should be considered as only the first chapter in a much longer story about the role of contrast agents on radiation dose at CT.
Purpose
Small airways with inner diameters less than 2 mm are sites of major airflow limitations in patients with chronic obstructive pulmonary disease (COPD) and asthma. The purpose of this study is ...to investigate the limitations for accurate assessment of small airway dimensions using both high‐resolution CT (HRCT) and conventional normal‐resolution CT at low dose levels.
Methods
To model the normal human airways from the 3rd to 20th generations, a cylindrical polyurethane phantom with 14 airway tubes of inner diameters (ID) ranging from 0.3 to 3.4 mm and wall thicknesses (WT) ranging from 0.15 to 1.6 mm was placed within an Anthropomorphic QRM‐Thorax phantom. The Aquilion Precision (Canon Medical Systems Corporation) HRCT scanner was used to acquire images at 80, 100, and 120 kV using high resolution mode (HR, 0.25 mm × 160 detector configuration) and normal‐resolution (NR) mode (0.5 mm × 80 detector configuration). The HR data were reconstructed using a 1024 × 1024 matrix (0.22 × 0.22 × 0.25 mm voxel size) and the NR data were reconstructed using a 512 × 512 matrix (0.43 × 0.43 × 0.50 mm). Two reconstruction algorithms (filtered back projection; FBP and an adaptive iterative dose reduction 3D algorithm; AIDR 3D) and three reconstruction kernels (FC30, FC52, and FC56) were investigated. The CTDIvol dose values ranged from 0.2 to 6.2 mGy. A refined automated full‐width half‐maximum (FWHM) method was used for the measurement of airway dimensions, where the density profiles were computed by radial oversampling using a polar coordinate system. Both ID and WT were compared to the known dimensions using a regression model, and the root‐mean‐square error (RMSE) and average error were computed across all 14 airway tubes.
Results
The results indicate that the ID can be measured within a 15% error down to approximately 0.8 and 2.0 mm using the HR and NR modes, respectively. The overall RMSE (and average error) of ID measurements for HR and NR were 0.10 mm (−0.70%) and 0.31 mm (−2.63%), respectively. The RMSE (and average error) of WT measurements using HR and NR were 0.10 mm (23.27%) and 0.27 mm (53.56%), respectively. The WT measurement using HR yielded a factor of two improvement in accuracy as compared to NR.
Conclusions
High‐resolution CT can provide more accurate measurements of airway dimensions as compared with NR CT, potentially improving quantitative assessment of pathologies such as COPD and asthma. The HR mode acquired and reconstructed with AIDR3D and the FC52 kernel provides most accurate measurement of airway dimensions. Low‐dose HR measurements at dose level above 0.9 mGy can provide improved accuracy on both inner diameters and wall thicknesses compared to full dose NR airway phantom measurements.
Purpose
To present a dataset of computational digital breast phantoms derived from high‐resolution three‐dimensional (3D) clinical breast images for the use in virtual clinical trials in ...two‐dimensional (2D) and 3D x‐ray breast imaging.
Acquisition and validation methods
Uncompressed computational breast phantoms for investigations in dedicated breast CT (BCT) were derived from 150 clinical 3D breast images acquired via a BCT scanner at UC Davis (California, USA). Each image voxel was classified in one out of the four main materials presented in the field of view: fibroglandular tissue, adipose tissue, skin tissue, and air. For the image classification, a semi‐automatic software was developed. The semi‐automatic classification was compared via manual glandular classification performed by two researchers. A total of 60 compressed computational phantoms for virtual clinical trials in digital mammography (DM) and digital breast tomosynthesis (DBT) were obtained from the corresponding uncompressed phantoms via a software algorithm simulating the compression and the elastic deformation of the breast, using the tissue’s elastic coefficient. This process was evaluated in terms of glandular fraction modification introduced by the compression procedure. The generated cohort of 150 uncompressed computational breast phantoms presented a mean value of the glandular fraction by mass of 12.3%; the average diameter of the breast evaluated at the center of mass was 105 mm. Despite the slight differences between the two manual segmentations, the resulting glandular tissue segmentation did not consistently differ from that obtained via the semi‐automatic classification. The difference between the glandular fraction by mass before and after the compression was 2.1% on average. The 60 compressed phantoms presented an average glandular fraction by mass of 12.1% and an average compressed thickness of 61 mm.
Data format and access
The generated digital breast phantoms are stored in DICOM files. Image voxels can present one out of four values representing the different classified materials: 0 for the air, 1 for the adipose tissue, 2 for the glandular tissue, and 3 for the skin tissue. The generated computational phantoms datasets were stored in the Zenodo public repository for research purposes (http://doi.org/10.5281/zenodo.4529852, http://doi.org/10.5281/zenodo.4515360).
Potential applications
The dataset developed within the INFN AGATA project will be used for developing a platform for virtual clinical trials in x‐ray breast imaging and dosimetry. In addition, they will represent a valid support for introducing new breast models for dose estimates in 2D and 3D x‐ray breast imaging and as models for manufacturing anthropomorphic physical phantoms.
Antimicrobial resistance (AMR) in bacterial pathogens threatens global health, though the spread of AMR bacteria and AMR genes between humans, animals, and the environment is still largely unknown. ...Here, we investigated the role of wild birds in the epidemiology of AMR Escherichia coli. Using next-generation sequencing, we characterized cephalosporin-resistant E. coli cultured from sympatric gulls and bald eagles inhabiting a landfill habitat in Alaska to identify genetic determinants conferring AMR, explore potential transmission pathways of AMR bacteria and genes at this site, and investigate how their genetic diversity compares to isolates reported in other taxa. We found genetically diverse E. coli isolates with sequence types previously associated with human infections and resistance genes of clinical importance, including bla
and bla
. Identical resistance profiles were observed in genetically unrelated E. coli isolates from both gulls and bald eagles. Conversely, isolates with indistinguishable core-genomes were found to have different resistance profiles. Our findings support complex epidemiological interactions including bacterial strain sharing between gulls and bald eagles and horizontal gene transfer among E. coli harboured by birds. Results suggest that landfills may serve as a source for AMR acquisition and/or maintenance, including bacterial sequence types and AMR genes relevant to human health.