Purpose
Femoral component malrotation in total knee arthroplasty (TKA) is clinically proven to cause dissatisfaction and impaired function. This study is an attempt to characterize the tibiofemoral ...kinematics following femoral malrotation in posterior stabilized (PS) TKA. It was hypothesized that internal malrotation would introduce the most pronounced changes.
Methods
Six fresh-frozen cadaver specimens were mounted in a kinematic rig. Three motion patterns were applied with the native knee and following PS TKA (passive motion, open chain extension, and squatting) while infrared cameras recorded the trajectories of markers attached to femur and tibia. Three different femoral implants were tested: a conventional posterior stabilized component, and adapted components of the same implant with 5° of intrinsic external and internal rotation, respectively.
Results
The implantation of the PS TKA resulted in less tibial internal rotation (squat 33–70°,
p
< 0.05) and the medial femoral condyle shifted posteriorly especially in deep flexion (squat 84–111°,
p
< 0.05). Internal component malrotation caused internal rotation and abduction of the tibia in flexion (squat 33–111°,
p
< 0.05), an elevated (squat 43–111°,
p
< 0.05) and more anterior (passive 61–126°,
p
< 0.05) located medial femoral condyle and a lateral femoral condyle located more posterior and inferior (squat 73–111°,
p
< 0.05) than in the neutrally aligned TKA. External component malrotation caused only little changes under passive motion. Under a squat there was less internal rotation and more adduction to the tibia (33–111°,
p
< 0.05). The medial femoral condyle was moved more posterior (squat 59–97°,
p
< 0.05), the lateral femoral condyle more superior (squat 54–105°,
p
< 0.05) than in the neutrally aligned TKA.
Conclusion
The greatest differences to the native tibiofemoral kinematics were introduced by internal rotation of the femoral component. Also neutrally and externally rotated femoral components introduce kinematic changes, but to a lesser extent. With respect to the alterations introduced to kinematics internal malrotation should be avoided when performing PS TKA.
Total knee arthroplasties (TKAs) for patients aged ≤35 years are rare but necessary for patients who have diseases such as juvenile idiopathic arthritis, osteonecrosis, osteoarthritis, and rheumatoid ...arthritis. Few studies have examined the 10-year and 20-year survivorship and clinical outcomes of TKAs for young patients.
A retrospective registry review identified 185 TKAs in 119 patients aged ≤ 35 years performed between 1985 and 2010 at a single institution. The primary outcome was implant survivorship free of revision. Patient-reported outcomes were assessed at 2 time points: 2011 to 2012 and 2018 to 2019. The average age was 26 years (range, 12 to 35). Mean follow-up was 17 years (range, 8 to 33).
Survivorship decreased from 84% (95% confidence interval CI: 79 to 90) at 5 years to 70% (95% CI: 64 to 77) at 10 years and to 37% (95% CI: 29 to 45) at 20 years. The most common reasons for revision were aseptic loosening (6%) and infection (4%). Risk factors for revision included increasing age at time of surgery (Hazards Ratio HR 1.3, P = .01) and use of constrained (HR 1.7, P = .05) or hinged prostheses (HR 4.3, P = .02). There were 86% of patients reporting that their surgery resulted in “a great improvement” or better.
Survivorship of TKAs in young patients is less favorable than expected. However, for the patients who responded to our surveys, TKA demonstrated substantial pain relief and improvement in function at 17-year follow-up. Revision risk increased with older age and higher levels of constraint.
sup.240Pu and sup.242Pu targets were prepared by the so-called "molecular plating" on an Al backing for cross-section measurements. The total activity of the actinide layer was determined by ...low-geometry alpha-particle counting. The atomic abundances of the two Pu materials were determined via thermal ionization mass spectrometry. A radiochemical separation was performed by anion exchange prior to the preparation of the layers to prevent interferences with the ingrown daughter nuclide during the characterization and cross-section measurements. The targets were prepared to be used in the project "Metrology for New Generation Nuclear Power Plants" (MetroFission), within the frame of the European Metrology Research Programme, directed to improve the knowledge on neutron cross sections through metrological approaches. For measurements of the sup.240Pu(n, f) and sup.242Pu(n, f) cross sections sup.240Pu and sup.242Pu targets were produced and mounted in fission chambers. In parallel, sup.240Pu and sup.242Pu targets were prepared for the ANDES project (Accurate Nuclear Data for Nuclear Energy Sustainability) for cross-section measurements at the CERN neutron time-of-flight facility n_TOF, and the Van de Graaff laboratories of IRMM and CNRS/CENBG. Keywords sup.240Pu * sup.242Pu * Target * Molecular plating * Radiochemical separation * Alpha-particle counting * Thermal ionization mass spectrometry * Metrology
Introduction
No evidence-based guidelines are available to determine the appropriate stem length, and whether or not to cement stems in revision total knee arthroplasty (TKA). Therefore, the ...objective of this study was to compare stresses and relative movement of cemented and uncemented stems of different lengths using a finite element analysis.
Materials and methods
A finite element model was created for a synthetic tibia. Two stem lengths (95 and 160 mm) and two types of fixation (cemented or press fit) of a hinged TKA were examined. The average compressive stress distribution in different regions of interest, as well as implant micromotions, was determined and compared during lunge and squat motor tasks.
Results
Both long and short stems in revision TKA lead to high stresses, primarily in the region around the stem tip. The presence of cement reduces the stresses in the bone in every region along the stem. Short stem configurations are less affected by the presence of cement than the long stem configuration. Press-fit stems showed higher micromotions compared to cemented stems.
Conclusions
Lowest stresses and micromotion were found for long cemented stems. Cementless stems showed more micromotion and increased stress levels especially at the level of the stem tip, which may explain the clinical phenomenon of stem-end pain following revision knee arthroplasty. These findings will help the surgeon with optimal individual implant choice.
Purpose
Balancing unicondylar knee arthroplasty (UKA) is challenging. If not performed properly, it may lead to implant loosening or progression of osteoarthritis in the preserved compartment. This ...study was aimed to document the biomechanical effects of improper balancing. We hypothesised that overstuffing would lead to more valgus, higher strain in the medial collateral ligament (sMCL), and higher lateral contact force.
Methods
Six fresh-frozen cadaver specimens were mounted in a kinematic rig. Three motion patterns were applied with the native knee and following medial UKA (passive motion, open-chain extension, and squatting), while infrared cameras recorded the trajectories of markers attached to femur and tibia. Three inlay thicknesses were tested (8, 9, 10 mm).
Results
Overstuffed knees were in more valgus and showed less tibial rotation and higher strains in the sMCL (
p
< 0.05). Lateral contact forces were higher in some specimens and lower in others. Stiffening of the medial compartment by UKA, even well balanced, already leads to a knee more in valgus with a more stressed sMCL. Overstuffing increases these effects. Knees with a tight sMCL may even see lower lateral contact force. Biomechanics were closest to the native knee with understuffing.
Conclusion
The first two hypotheses were confirmed, but not the latter. This underlines the importance of optimal balancing. Overstuffing should certainly be avoided. Although kinematics is only slightly affected, contact forces and ligament strains are considerably changed and this might be of more clinical importance. It is advisable to use thinner inlays, if stability is not compromised.
Neutron capture cross section measurements have been performed at the time-of-flight facility GELINA of the EC-JRC-Geel. Prompt gamma rays, originating from a natural silver sample, were detected by ...a pair of C6D6 liquid scintillation detectors. The total energy detection principle in combination with the pulse height weighting technique has been used. In this contribution the experimental details together with the data reduction process are described. In addition, first results of calculations with REFIT are presented to verify the quality of recommended cross section data in the resolved resonance region.
Introduction
Patient-specific instrumentation (PSI) was introduced in an attempt to reduce positional outliers of components in total knee arthroplasty (TKA). It was hypothesized that PSI could help ...with the positioning of tibial components in optimal rotational alignment.
Methods
A magnetic resonance imaging (MRI) analysis of 58 patients following TKA was conducted. Of these, 30 operations were performed using PSI and 28 using conventional instrumentation. The rotation of the tibial components was determined in MRI using three different reference lines: a tangent to the dorsal tibial condyles, the tibial epicondylar line, and the tibial tubercle. Deviations >9° were considered outliers. Also internal rotation >1° was considered an outlier. Data were analyzed statistically for positional outliers using the Chi-squared test.
Results
There was excellent inter- and intraobserver reliability with low standard deviations for the determination of tibial component rotation using the tangent to the dorsal condyles and the tibial epicondylar line as reference. Using the dorsal tangent as reference, there were eight components in excessive external rotation (28.6 %) and one component being in relative internal rotation (5.4°) in the conventional group, while there were two components in excessive external rotation in the PSI group (6.7 %). Using the tibial epicondyles as reference, there were seven components in excessive external rotation (21.4 %) and one component being in relative internal rotation (4.4°) in the conventional group; while there were two components in excessive external rotation in the PSI group (6.7 %). These differences were statistically significant (
p
< 0.05). Measurements based on the tibial tubercle showed poor reproducibility in terms of intra- and interobserver reliability and was of little use in the context of the research question.
Discussion and conclusion
In this setup, PSI was effective in significantly reducing outliers of optimal rotational tibial component alignment during TKA. Anatomy of the proximal tibia does not deliver clear landmarks that are prominent and consistent. This makes both, MRI analysis as well as cutting jig production and intraoperative placement a challenge.
Severe acetabular deficiencies in cases of developmental dysplasia of the hip (DDH) often require complex reconstructive procedures in total hip arthroplasty (THA). The use of autologous femoral head ...grafts for acetabular reconstruction has been described, but few data is available about clinical results, the rates of non-union or aseptic loosening of acetabular components.
In a retrospective approach, 101 patients with 118 THA requiring autologous femoral head grafts to the acetabulum because of DDH were included. Six patients had died, another 6 were lost to follow-up, and 104 hips were available for clinical and radiological evaluation at a mean of 68 ± 15 (13 to 159) months.
The average Merle d'Aubigné hip score improved from 9 to 16 points. Seven implants had to be revised due to aseptic loosening (6.7%). The revisions were performed 90 ± 34 (56 to 159) months after implantation. The other hips showed a stable position of the sockets without any signs of bony non-union, severe radiolucencies at the implant-graft interface or significant resorption of the graft.
The use of autologous femoral head grafts with cementless cups in primary THA can achieve promising short- to midterm results in patients with dysplastic hips.
Assessment of shoulder mobility is essential for clinical follow-up of shoulder treatment. Only a few high sophisticated instruments for objective measurements of shoulder mobility are available. The ...interobserver dependency of conventional goniometer measurements is high. In the 1990s an isokinetic measuring system of BIODEX Inc. was introduced, which is a very complex but valid instrument. Since 2008 a new user-friendly system called DynaPort MiniMod TriGyro ShoulderTest-System (DP) is available. Aim of this study is the validation of this measuring instrument using the BIODEX-System.
The BIODEX is a computerized robotic dynamometer used for isokinetic testing and training of athletes. Because of its size the system needs to be installed in a separated room. The DP is a small, light-weighted three-dimensional gyroscope that is fixed on the distal upper patient arm, recording abduction, flexion and rotation. For direct comparison we fixed the DP on the lever arm of the BIODEX. The accuracy of measurement was determined at different positions, angles and distances from the centre of rotation (COR) as well as different velocities in a radius between 0° - 180° in steps of 20°. All measurements were repeated 10 times. As satisfactory accuracy a difference between both systems below 5° was defined. The statistical analysis was performed with a linear regression model.
The evaluation shows very high accuracy of measurements. The maximum average deviation is below 2.1°. For a small range of motion the DP is slightly underestimating comparing the BIODEX, whereas for higher angles increasing positive differences are observed. The distance to the COR as well as the position of the DP on the lever arm have no significant influence. Concerning different motion speeds significant but not relevant influence is detected. Unfortunately device related effects are observed, leading to differences between repeated measurements with any two different devices up to 8° at maximal range of motion (180°).
In summary the results shows high correlation and good reproducibility of measurements. All deviations are inside the tolerance interval of 5°, if one device is used. An unlikely systematic device effect is detected. These laboratory trials are promising for the validation of this system in humans. The challenge for both systems will be the changing of the COR in the shoulder joint at elevations higher than 90°.
Celotno besedilo
Dostopno za:
DOBA, IZUM, KILJ, NUK, PILJ, PNG, SAZU, SIK, UILJ, UKNU, UL, UM, UPUK
Primary stability of cartilage repair constructs is of the utmost importance in the clinical setting but few continuous passive motion (CPM) models are available. Our study aimed to establish a novel ...ex vivo CPM animal model and to evaluate the required motion cycles for testing the mechanical properties of a new cell-free collagen type I gel plug (CaReS®-1S).
A novel ex vivo CPM device was developed. Full-thickness cartilage defects (11 mm diameter by 6 mm deep) were created on the medial femoral condyle of porcine knee specimens. CaReS®-1S was implanted in 16 animals and each knee underwent continuous passive motion. After 0, 2000, 4000, 6000, and 8000 motions, standardized digital pictures of the grafts were taken, focusing on the worn surfaces. The percentage of worn surface on the total CaReS®-1S surface was evaluated with image processing software.
Significant differences in the worn surface were recorded between 0 and 2000 motion cycles (p < 0.0001). After 2000 motion cycles, there was no significant difference. No total delamination of CaReS®-1S with an empty defect site was recorded.
The ex vivo CPM animal model is appropriate in investigating CaReS®-1S durability under continuous passive motion. 2000 motion cycles appear adequate to assess the primary stability of type I collagen gels used to repair focal chondral defects.
Celotno besedilo
Dostopno za:
DOBA, IZUM, KILJ, NUK, PILJ, PNG, SAZU, SIK, UILJ, UKNU, UL, UM, UPUK