Type 1 diabetes is one of the most common endocrine problems in childhood and adolescence, and remains a serious chronic disorder with increased morbidity and mortality, and reduced quality of life. ...Technological innovations positively affect the management of type 1 diabetes. Closed-loop insulin delivery (artificial pancreas) is a recent medical innovation, aiming to reduce the risk of hypoglycemia while achieving tight control of glucose. Characterized by real-time glucose-responsive insulin administration, closed-loop systems combine glucose-sensing and insulin-delivery components. In the most viable and researched configuration, a disposable sensor measures interstitial glucose levels, which are fed into a control algorithm controlling delivery of a rapid-acting insulin analog into the subcutaneous tissue by an insulin pump. Research progress builds on an increasing use of insulin pumps and availability of glucose monitors. We review the current status of insulin delivery, focusing on clinical evaluations of closed-loop systems. Future goals are outlined, and benefits and limitations of closed-loop therapy contrasted. The clinical utility of these systems is constrained by inaccuracies in glucose sensing, inter- and intra-patient variability, and delays due to absorption of insulin from the subcutaneous tissue, all of which are being gradually addressed.
Celotno besedilo
Dostopno za:
DOBA, IZUM, KILJ, NUK, PILJ, PNG, SAZU, SIK, UILJ, UKNU, UL, UM, UPUK
Accuracy and frequency of glucose measurement is essential to achieve safe and efficacious glucose control in the intensive care unit. Emerging continuous glucose monitors provide frequent ...measurements, trending information, and alarms. The objective of this study was to establish the level of accuracy of continuous glucose monitoring (CGM) associated with safe and efficacious glucose control in the intensive care unit.
We evaluated 3 established glucose control protocols Yale, University of Washington, and Normoglycemia in Intensive Care Evaluation and Surviving Using Glucose Algorithm Regulation (NICE-SUGAR) by use of computer simulations. Insulin delivery was informed by intermittent blood glucose (BG) measurements or CGM levels with an increasing level of measurement error. Measures of glucose control included mean glucose, glucose variability, proportion of time glucose was in target range, and hypoglycemia episodes.
Apart from the Washington protocol, CGM with mean absolute relative deviation (MARD) ≤ 15% resulted in similar mean glucose as with the use of intermittent BG measurements. Glucose variability was also similar between CGM and BG-informed protocols. Frequency and duration of hypoglycemia were not worse by use of CGM with MARD ≤ 10%. Measures of glucose control varied more between protocols than at different levels of the CGM error.
The efficacy of CGM-informed and BG-informed commonly used glucose protocols is similar, but the risk of hypoglycemia may be reduced by use of CGM with MARD ≤ 10%. Protocol choice has greater influence on glucose control measures than the glucose measurement method.
We aimed to assess the feasibility and safety of hybrid closed-loop insulin delivery in children with type 1 diabetes aged 1-7 years as well as evaluate the role of diluted insulin on glucose ...control.
In an open-label, multicenter, multinational, randomized crossover study, 24 children with type 1 diabetes on insulin pump therapy (median age 5 years interquartile range 3-6 and mean ± SD HbA
7.4 ± 0.7% 57 ± 8 mmol/mol and total insulin 13.2 ± 4.8 units/day) underwent two 21-day periods of unrestricted living and we compared hybrid closed-loop with diluted insulin (U20) and hybrid closed-loop with standard strength insulin (U100) in random order. During both interventions, the Cambridge model predictive control algorithm was used.
The proportion of time that sensor glucose was in the target range between 3.9 and 10 mmol/L (primary end point) was not different between interventions (mean ± SD 72 ± 8% vs. 70 ± 7% for closed-loop with diluted insulin vs. closed-loop with standard insulin, respectively;
= 0.16). There was no difference in mean glucose levels (8.0 ± 0.8 vs. 8.2 ± 0.6 mmol/L;
= 0.14), glucose variability (SD of sensor glucose 3.1 ± 0.5 vs. 3.2 ± 0.4 mmol/L;
= 0.16), or the proportion of time spent with sensor glucose <3.9 mmol/L (4.5 ± 1.7% vs. 4.7 ± 1.4%;
= 0.47) or <2.8 mmol/L (0.6 ± 0.5% vs. 0.6 ± 0.4%;
> 0.99). Total daily insulin delivery did not differ (17.3 ± 5.6 vs. 18.9 ± 6.9 units/day;
= 0.07). No closed-loop-related severe hypoglycemia or ketoacidosis occurred.
Unrestricted home use of day-and-night closed-loop in very young children with type 1 diabetes is feasible and safe. The use of diluted insulin during closed-loop does not provide additional benefits compared with standard strength insulin.
Automated insulin delivery systems, also known as closed-loop or ‘artificial pancreas’ systems, are transforming the management of type 1 diabetes. These systems consist of an algorithm which ...responds to real-time glucose sensor levels by automatically modulating insulin delivery through an insulin pump. We review the rapidly changing landscape of automated insulin-delivery systems over recent decades, from initial prototypes to the different hybrid closed-loop systems commercially available today. We discuss the growing body of clinical trials and real-world evidence demonstrating their glycaemic and psychosocial benefits. We also address future directions in automated insulin delivery such as dual-hormone systems and adjunct therapy as well as the challenges around ensuring equitable access to closed-loop technology.
There is considerable physiological and clinical evidence of harm and increased risk of death associated with dysglycemia in critical care. However, glycemic control (GC) currently leads to increased ...hypoglycemia, independently associated with a greater risk of death. Indeed, recent evidence suggests GC is difficult to safely and effectively achieve for all patients. In this review, leading experts in the field discuss this evidence and relevant data in diabetology, including the artificial pancreas, and suggest how safe, effective GC can be achieved in critically ill patients in ways seeking to mimic normal islet cell function. The review is structured around the specific clinical hurdles of: understanding the patient's metabolic state; designing GC to fit clinical practice, safety, efficacy, and workload; and the need for standardized metrics. These aspects are addressed by reviewing relevant recent advances in science and technology. Finally, we provide a set of concise recommendations to advance the safety, quality, consistency, and clinical uptake of GC in critical care. This review thus presents a roadmap toward better, more personalized metabolic care and improved patient outcomes.
To quantify age-related variability of insulin needs during day and night closed-loop insulin delivery.
We retrospectively analyzed data from hybrid closed-loop studies involving young children (1-6 ...years old,
= 20), children (7-12 years,
= 21), adolescents (13-17 years,
= 15), and adults (>18 years,
= 58) with type 1 diabetes. The coefficient of variation quantified variability of insulin needs during 3 weeks of unrestricted-living hybrid closed-loop use.
Data from 2,365 nights and 2,367 days in 114 participants were analyzed. The coefficient of variation of insulin delivery was higher in young children compared with adults (mean difference at nighttime 10.7 percentage points 95% CI 2.9-18.4,
= 0.003; daytime 6.4 percentage points 95% CI 2.0-10.9,
= 0.002) and compared with adolescents (mean difference at nighttime 10.2 percentage points 95% CI 0.0-20.4,
= 0.049; daytime 7.0 percentage points 95% CI 1.1-12.8,
= 0.014).
Diabetes management in young children is complicated by higher variability in insulin requirements, supporting fast-track clinical practice adoption of closed-loop in this vulnerable population.
Tight control of blood glucose concentration in people with type 1 diabetes predisposes to hypoglycaemia. We aimed to investigate whether day-and-night hybrid closed-loop insulin delivery can improve ...glucose control while alleviating the risk of hypoglycaemia in adults with HbA1c below 7·5% (58 mmol/mol).
In this open-label, randomised, crossover study, we recruited adults (aged ≥18 years) with type 1 diabetes and HbA1c below 7·5% from Addenbrooke's Hospital (Cambridge, UK) and Medical University of Graz (Graz, Austria). After a 2–4 week run-in period, participants were randomly assigned (1:1), using web-based randomly permuted blocks of four, to receive insulin via the day-and-night hybrid closed-loop system or usual pump therapy for 4 weeks, followed by a 2–4 week washout period and then the other intervention for 4 weeks. Treatment interventions were unsupervised and done under free-living conditions. During the closed-loop period, a model-predictive control algorithm directed insulin delivery, and prandial insulin delivery was calculated with a standard bolus wizard. The primary outcome was the proportion of time when sensor glucose concentration was in target range (3·9–10·0 mmol/L) over the 4 week study period. Analyses were by intention to treat. This study is registered with ClinicalTrials.gov, number NCT02727231, and is completed.
Between March 21 and June 24, 2016, we recruited 31 participants, of whom 29 were randomised. One participant withdrew during the first closed-loop period because of dissatisfaction with study devices and glucose control. The proportion of time when sensor glucose concentration was in target range was 10·5 percentage points higher (95% CI 7·6–13·4; p<0·0001) during closed-loop delivery compared with usual pump therapy (65·6% SD 8·1 when participants used usual pump therapy vs 76·2% 6·4 when they used closed-loop). Compared with usual pump therapy, closed-loop delivery also reduced the proportion of time spent in hypoglycaemia: the proportion of time with glucose concentration below 3·5 mmol/L was reduced by 65% (53–74, p<0·0001) and below 2·8 mmol/L by 76% (59–86, p<0·0001). No episodes of serious hypoglycaemia or other serious adverse events occurred.
Use of day-and-night hybrid closed-loop insulin delivery under unsupervised, free-living conditions for 4 weeks in adults with type 1 diabetes and HbA1c below 7·5% is safe and well tolerated, improves glucose control, and reduces hypoglycaemia burden. Larger and longer studies are warranted.
Swiss National Science Foundation (P1BEP3_165297), JDRF, UK National Institute for Health Research Cambridge Biomedical Research Centre, and Wellcome Strategic Award (100574/Z/12/Z).
The Artificial Pancreas and Type 1 Diabetes Nwokolo, Munachiso; Hovorka, Roman
The journal of clinical endocrinology and metabolism,
07/2023, Letnik:
108, Številka:
7
Journal Article
Recenzirano
Odprti dostop
Abstract
Diabetes technologies represent a paradigm shift in type 1 diabetes care. Continuous subcutaneous insulin infusion (CSII) pumps and continuous glucose monitors (CGM) improve glycated ...hemoglobin (HbA1c) levels, enhance time in optimal glycemic range, limit severe hypoglycemia, and reduce diabetes distress. The artificial pancreas or closed-loop system connects these devices via a control algorithm programmed to maintain target glucose, partially relieving the person living with diabetes of this constant responsibility. Automating insulin delivery reduces the input required from those wearing the device, leading to better physiological and psychosocial outcomes. Hybrid closed-loop therapy systems, requiring user-initiated prandial insulin doses, are the most advanced closed-loop systems commercially available. Fully closed-loop systems, requiring no user-initiated insulin boluses, and dual hormone systems have been shown to be safe and efficacious in the research setting. Clinical adoption of closed-loop therapy remains in early stages despite recent technological advances. People living with diabetes, health care professionals, and regulatory agencies continue to navigate the complex path to equitable access. We review the available devices, evidence, clinical implications, and barriers regarding these innovatory technologies.
The growing incidence of type 2 diabetes (T2D) is a significant health concern, representing 90% of diabetes cases worldwide. As the disease progresses, resultant insulin deficiency and ...hyperglycaemia necessitates insulin therapy in many cases. It has been recognized that a significant number of people who have a clinical requirement for insulin therapy, as well as their healthcare professionals, are reluctant to intensify treatment with insulin due to fear of hypoglycaemia, poor understanding of treatment regimens or lack of engagement, and are therefore at higher risk of developing complications from poor glycaemic control. Over the past decade, the rise of diabetes technologies, including dosing advisors, continuous glucose monitoring systems, insulin pumps and automated insulin delivery systems, has led to great improvements in the therapies available, particularly to those requiring insulin. Although the focus has largely been on delivering these therapies to the type 1 diabetes population, it is becoming increasingly recognized that people with T2D face similar challenges to achieve recommended glycaemic standards and also have the potential to benefit from these advances. In this review, we discuss diabetes technologies that are currently available for people with T2D and the evidence supporting their use, as well as future prospects. We conclude that there is a clinical need to extend the use of these technologies to the T2D population to curb the consequences of suboptimal disease management in this group.