The integration of 1.5 T MRI functionality with a radiotherapy linear accelerator (linac) has been pursued since 1999 by the UMC Utrecht in close collaboration with Elekta and Philips. The idea ...behind this integrated device is to offer unrivalled, online and real-time, soft-tissue visualization of the tumour and the surroundings for more precise radiation delivery. The proof of concept of this device was given in 2009 by demonstrating simultaneous irradiation and MR imaging on phantoms, since then the device has been further developed and commercialized by Elekta. The aim of this work is to demonstrate the clinical feasibility of online, high-precision, high-field MRI guidance of radiotherapy using the first clinical prototype MRI-Linac. Four patients with lumbar spine bone metastases were treated with a 3 or 5 beam step-and-shoot IMRT plan. The IMRT plan was created while the patient was on the treatment table and based on the online 1.5 T MR images; pre-treatment CT was deformably registered to the online MRI to obtain Hounsfield values. Bone metastases were chosen as the first site as these tumors can be clearly visualized on MRI and the surrounding spine bone can be detected on the integrated portal imager. This way the portal images served as an independent verification of the MRI based guidance to quantify the geometric precision of radiation delivery. Dosimetric accuracy was assessed post-treatment from phantom measurements with an ionization chamber and film. Absolute doses were found to be highly accurate, with deviations ranging from 0.0% to 1.7% in the isocenter. The geometrical, MRI based targeting as confirmed using portal images was better than 0.5 mm, ranging from 0.2 mm to 0.4 mm. In conclusion, high precision, high-field, 1.5 T MRI guided radiotherapy is clinically feasible.
To evaluate prostate intrafraction motion using MRI during the full course of online adaptive MR-Linac radiotherapy (RT) fractions, in preparation of MR-guided extremely hypofractionated RT.
Five low ...and intermediate risk prostate cancer patients were treated with 20 × 3.1 Gy fractions on a 1.5T MR-Linac. Each fraction, initial MRI (Pre) scans were obtained at the start of every treatment session. Pre-treatment planning MRI contours were propagated and adapted to this Pre scan after which plan re-optimization was started in the treatment planning system followed by dose delivery. 3D Cine-MR imaging was started simultaneously with beam-on and acquired over the full beam-on period. Prostate intrafraction motion in this cine-MR was determined with a previously validated soft-tissue contrast based tracking algorithm. In addition, absolute accuracy of the method was determined using a 4D phantom.
Prostate motion was completely automatically determined over the full on-couch period (approx. 45 min) with no identified mis-registrations. The translation 95% confidence intervals are within clinically applied margins of 5 mm, and plan adaption for intrafraction motion was required in only 4 out of 100 fractions.
This is the first study to investigate prostate intrafraction motions during entire MR-guided RT sessions on an MR-Linac. We have shown that high quality 3D cine-MR imaging and prostate tracking during RT is feasible with beam-on. The clinically applied margins of 5 mm have proven to be sufficient for these treatments and may potentially be further reduced using intrafraction plan adaptation guided by cine-MR imaging.
The hybrid MRI-radiotherapy machines, like the MR-linac (Elekta AB, Stockholm, Sweden) installed at the UMC Utrecht (Utrecht, The Netherlands), will be able to provide real-time patient imaging ...during treatment. In order to take advantage of the system's capabilities and enable online adaptive treatments, a new generation of software should be developed, ranging from motion estimation to treatment plan adaptation. In this work we present a proof of principle adaptive pipeline designed for high precision stereotactic body radiation therapy (SBRT) suitable for sites affected by respiratory motion, like renal cell carcinoma (RCC). We utilized our research MRL treatment planning system (MRLTP) to simulate a single fraction 25 Gy free-breathing SBRT treatment for RCC by performing inter-beam replanning for two patients and one volunteer. The simulated pipeline included a combination of (pre-beam) 4D-MRI and (online) 2D cine-MR acquisitions. The 4DMRI was used to generate the mid-position reference volume, while the cine-MRI, via an in-house motion model, provided three-dimensional (3D) deformable vector fields (DVFs) describing the anatomical changes during treatment. During the treatment fraction, at an inter-beam interval, the mid-position volume of the patient was updated and the delivered dose was accurately reconstructed on the underlying motion calculated by the model. Fast online replanning, targeting the latest anatomy and incorporating the previously delivered dose was then simulated with MRLTP. The adaptive treatment was compared to a conventional mid-position SBRT plan with a 3 mm planning target volume margin reconstructed on the same motion trace. We demonstrate that our system produced tighter dose distributions and thus spared the healthy tissue, while delivering more dose to the target. The pipeline was able to account for baseline variations/drifts that occurred during treatment ensuring target coverage at the end of the treatment fraction.
To compare toxicity rates in patients with localized prostate cancer treated with standard fractionated external beam radiotherapy (EBRT) with or without an additional integrated boost to the ...macroscopically visible tumour.
FLAME is a phase 3 multicentre RCT (NCT01168479) of patients with pathologically confirmed localized intermediate or high-risk prostate cancer. The standard treatment arm (n = 287) received a dose to the entire prostate of 77 Gy in 35 fractions. The dose-escalated treatment arm (n = 284) received 77 Gy in 35 fractions to the entire prostate, with an integrated boost up to 95 Gy to the multi-parametric MRI-defined (macroscopic) tumour within the prostate. Treatment related toxicity was measured using the CTCAE version 3.0. Grade 2 or worse GU or GI events up to two years were compared between groups by presenting proportions and by Generalized Estimating Equations (GEE) analyses for repeated measures.
Ninety percent of the 571 men randomly assigned between September 2009 and January 2015 had high-risk disease (Ash 2000), of whom nearly 66% were prescribed hormonal therapy up to three years. Median follow-up was 55 months at the time of this analysis. Toxicity prevalence rates for both GI and GU increased until the end of treatment and regressed thereafter, with no obvious differences across treatment groups. Late cumulative GI toxicity rates were 11.1% and 10.2% for the standard and dose-escalated group, respectively. These rates were 22.6% and 27.1% for GU toxicity. GEE analyses showed that both GU toxicity and GI toxicity (≥grade 2) up to two years after treatment were similar between arms (OR 1.02 95%CI 0.78–1.33p = 0.81 and (OR 1.19 95%CI 0.82–1.73p = 0.38), respectively.
In intermediate- and high-risk prostate cancer patients, focal dose escalation integrated with standard EBRT did not result in an increase in GU and GI toxicity when compared to the standard treatment up to two years after treatment. This suggests that the described focal dose escalation technique is safe and feasible.
Magnetic resonance imaging (MRI) is often combined with computed tomography (CT) in prostate radiotherapy to optimise delineation of the target and organs-at-risk (OAR) while maintaining accurate ...dose calculation. Such a dual-modality workflow requires two separate imaging sessions, and it has some fundamental and logistical drawbacks. Due to the availability of new MRI hardware and software solutions, CT examinations can be omitted for prostate radiotherapy simulations. All information for treatment planning, including electron density maps and bony anatomy, can nowadays be obtained with MRI. Such an MRI-only simulation workflow reduces delineation ambiguities, eases planning logistics, and improves patient comfort; however, careful validation of the complete MRI-only workflow is warranted. The first institutes are now adopting this MRI-only workflow for prostate radiotherapy. In this article, we will review technology and workflow requirements for an MRI-only prostate simulation workflow.
•MRI-only workflows are being implemented in several centres for prostate cancer;•Before the introduction of an MRI-only workflow in clinical practice, the building blocks of this workflow should be implemented and validated.•MRI-only workflows are useful both in the context of X-ray or CBCT-guided radiotherapy as well as MRI-guided radiotherapy.
External beam radiotherapy (EBRT) is an important cornerstone in the treatment of localized prostate cancer. Current image-guided radiotherapy (IGRT) techniques allow for more accurate and precise ...delivery of radiation treatment by the use of imaging before each fraction. Magnetic resonance guided radiotherapy (MRgRT) is the next step in IGRT with hybrid systems combining linear accelerators with MRI-scanners. With MRgRT, it is possible to visualize pelvic anatomy in great detail and subsequently perform replanning of the radiation dose distribution before each radiotherapy fraction. This technique has the potential to increase the therapeutic window of EBRT, by improved normal tissue sparing due to margin reduction and more accurate target dose delivery. This is particularly promising for prostate cancer, with its biology lending itself to ultra-hypofractionation, reducing radiotherapy treatment to as little as five fractions. Also, recent studies have shown that focal dose escalation to the intraprostatic tumor to high ablative doses can substantially increase disease-free survival. In this article, we discuss these unique opportunities as well as the potential future benefits of MRgRT in prostate cancer treatment.
To investigate the dosimetric impact of intrafraction translation and rotation motion of the prostate, as extracted from daily acquired post-treatment 3D cine-MR based on soft-tissue contrast, in ...extremely hypofractionated (SBRT) prostate patients. Accurate dose reconstruction is performed by using a prostate intrafraction motion trace which is obtained with a soft-tissue based rigid registration method on 3D cine-MR dynamics with a temporal resolution of 11 s. The recorded motion of each time-point was applied to the planning CT, resulting in the respective dynamic volume used for dose calculation. For each treatment fraction, the treatment delivery record was generated by proportionally splitting the plan into 11 s intervals based on the delivered monitor units. For each fraction the doses of all partial plan/dynamic volume combinations were calculated and were summed to lead to the motion-affected fraction dose. Finally, for each patient the five fraction doses were summed, yielding the total treatment dose. Both daily and total doses were compared to the original reference dose of the respective patient to assess the impact of the intrafraction motion. Depending on the underlying motion of the prostate, different types of motion-affected dose distributions were observed. The planning target volumes (PTVs) ensured CTV_30 (seminal vesicles) D99% coverage for all patients, CTV_35 (prostate corpus) coverage for 97% of the patients and GTV_50 (local boost) for 83% of the patients when compared against the strict planning target D99% value. The dosimetric impact due to prostate intrafraction motion in extremely hypofractionated treatments was determined. The presented study is an essential step towards establishing the actual delivered dose to the patient during radiotherapy fractions.
For the successful completion of medical interventional procedures, several concepts, such as daily positioning compensation, dose accumulation or delineation propagation, rely on establishing a ...spatial coherence between planning images and images acquired at different time instants over the course of the therapy. To meet this need, image-based motion estimation and compensation relies on fast, automatic, accurate and precise registration algorithms. However, image registration quickly becomes a challenging and computationally intensive task, especially when multiple imaging modalities are involved. In the current study, a novel framework is introduced to reduce the computational overhead of variational registration methods. The proposed framework selects representative voxels of the registration process, based on a supervoxel algorithm. Costly calculations are hereby restrained to a subset of voxels, leading to a less expensive spatial regularized interpolation process. The novel framework is tested in conjunction with the recently proposed EVolution multi-modal registration method. This results in an algorithm requiring a low number of input parameters, is easily parallelizable and provides an elastic voxel-wise deformation with a subvoxel accuracy. The performance of the proposed accelerated registration method is evaluated on cross-contrast abdominal T1/T2 MR-scans undergoing a known deformation and annotated CT-images of the lung. We also analyze the ability of the method to capture slow physiological drifts during MR-guided high intensity focused ultrasound therapies and to perform multi-modal CT/MR registration in the abdomen. Results have shown that computation time can be reduced by 75% on the same hardware with no negative impact on the accuracy.
•Different image post-processing introduces ADC variation between centres.•Aligning ADC calculation methods between centres improves reproducibility.•Calculation-related ADC variation is larger than ...delineation-related ADC variation.
The apparent diffusion coefficient (ADC), a potential imaging biomarker for radiotherapy response, needs to be reproducible before translation into clinical use. The aim of this study was to evaluate the multi-centre delineation- and calculation-related ADC variation and give recommendations to minimize it.
Nine centres received identical diffusion-weighted and anatomical magnetic resonance images of different cancerous tumours (adrenal gland, pelvic oligo metastasis, pancreas, and prostate). All centres delineated the gross tumour volume (GTV), clinical target volume (CTV), and viable tumour volume (VTV), and calculated ADCs using both their local calculation methods and each of the following calculation conditions: b-values 0–500 vs. 150–500 s/mm2, region-of-interest (ROI)-based vs. voxel-based calculation, and mean vs. median. ADC variation was assessed using the mean coefficient of variation across delineations (CVD) and calculation methods (CVC). Absolute ADC differences between calculation conditions were evaluated using Friedman’s test. Recommendations for ADC calculation were formulated based on observations and discussions within the Elekta MRI-linac consortium image analysis working group.
The median (range) CVD and CVC were 0.06 (0.02–0.32) and 0.17 (0.08–0.26), respectively. The ADC estimates differed 18% between b-value sets and 4% between ROI/voxel-based calculation (p-values < 0.01). No significant difference was observed between mean and median (p = 0.64). Aligning calculation conditions between centres reduced CVC to 0.04 (0.01–0.16). CVD was comparable between ROI types.
Overall, calculation methods had a larger impact on ADC reproducibility compared to delineation. Based on the results, significant sources of variation were identified, which should be considered when initiating new studies, in particular multi-centre investigations.
To develop an automated radiotherapy treatment planning and optimization workflow to efficiently create patient specifically optimized clinical grade treatment plans for prostate cancer and to ...implement it in clinical practice. A two-phased planning and optimization workflow was developed to automatically generate 77Gy 5-field simultaneously integrated boost intensity modulated radiation therapy (SIB-IMRT) plans for prostate cancer treatment. A retrospective planning study (n = 100) was performed in which automatically and manually generated treatment plans were compared. A clinical pilot (n = 21) was performed to investigate the usability of our method. Operator time for the planning process was reduced to <5 min. The retrospective planning study showed that 98 plans met all clinical constraints. Significant improvements were made in the volume receiving 72Gy (V72Gy) for the bladder and rectum and the mean dose of the bladder and the body. A reduced plan variance was observed. During the clinical pilot 20 automatically generated plans met all constraints and 17 plans were selected for treatment. The automated radiotherapy treatment planning and optimization workflow is capable of efficiently generating patient specifically optimized and improved clinical grade plans. It has now been adopted as the current standard workflow in our clinic to generate treatment plans for prostate cancer.