Advances in diabetes technology have transformed the treatment paradigm for type 1 diabetes, yet the burden of disease is significant. We report on a pivotal safety study of the first tubeless, ...on-body automated insulin delivery system with customizable glycemic targets.
This single-arm, multicenter, prospective study enrolled 112 children (age 6-13.9 years) and 129 adults (age 14-70 years). A 2-week standard therapy phase (usual insulin regimen) was followed by 3 months of automated insulin delivery. Primary safety outcomes were incidence of severe hypoglycemia and diabetic ketoacidosis. Primary effectiveness outcomes were change in HbA
and percent time in sensor glucose range 70-180 mg/dL ("time in range").
A total of 235 participants (98% of enrolled, including 111 children and 124 adults) completed the study. HbA
was significantly reduced in children by 0.71% (7.8 mmol/mol) (mean ± SD: 7.67 ± 0.95% to 6.99 ± 0.63% 60 ± 10.4 mmol/mol to 53 ± 6.9 mmol/mol,
< 0.0001) and in adults by 0.38% (4.2 mmol/mol) (7.16 ± 0.86% to 6.78 ± 0.68% 55 ± 9.4 mmol/mol to 51 ± 7.4 mmol/mol,
< 0.0001). Time in range was improved from standard therapy by 15.6 ± 11.5% or 3.7 h/day in children and 9.3 ± 11.8% or 2.2 h/day in adults (both
< 0.0001). This was accomplished with a reduction in time in hypoglycemia <70 mg/dL among adults (median interquartile range: 2.00% 0.63, 4.06 to 1.09% 0.46, 1.75,
< 0.0001), while this parameter remained the same in children. There were three severe hypoglycemia events not attributable to automated insulin delivery malfunction and one diabetic ketoacidosis event from an infusion site failure.
This tubeless automated insulin delivery system was safe and allowed participants to significantly improve HbA
levels and time in target glucose range with a very low occurrence of hypoglycemia.
The objective of this study was to assess the safety and effectiveness of the first commercial configuration of a tubeless automated insulin delivery system, Omnipod
5, in children (6-13.9 years) and ...adults (14-70 years) with type 1 diabetes (T1D) in an outpatient setting.
This was a single-arm, multicenter, prospective clinical study. Data were collected over a 14-day standard therapy (ST) phase followed by a 14-day hybrid closed-loop (HCL) phase, where participants (
= 36) spent 72 h at each of three prespecified glucose targets (130, 140, and 150 mg/dL, 9 days total) then 5 days with free choice of glucose targets (110-150 mg/dL) using the Omnipod 5. Remote safety monitoring alerts were enabled during the HCL phase. Primary endpoints were difference in time in range (TIR) (70-180 mg/dL) between ST and HCL phases and proportion of participants reporting serious device-related adverse events.
Mean TIR was significantly higher among children in the free-choice period overall (64.9% ± 12.2%,
< 0.01) and when using a 110 mg/dL target (71.2% ± 10.2%,
< 0.01), a 130 mg/dL target (61.5% ± 7.7%,
< 0.01), and a 140 mg/dL target (64.8% ± 11.6%,
< 0.01), and among adults using a 130 mg/dL target (75.1% ± 11.6%,
< 0.05), compared to the ST phase (children: 51.0% ± 13.3% and adults: 65.6% ± 15.7%). There were no serious device-related adverse events reported during the HCL phase, nor were there episodes of severe hypoglycemia or diabetic ketoacidosis.
The Omnipod 5 System was safe and effective when used at glucose targets from 110 to 150 mg/dL for 14 days at home in children and adults with T1D.
Very young children with type 1 diabetes often struggle to achieve glycemic targets, putting them at risk for long-term complications and creating an immense management burden for caregivers. We ...conducted the first evaluation of the Omnipod 5 Automated Insulin Delivery System in this population.
A total of 80 children aged 2.0-5.9 years used the investigational system in a single-arm study for 13 weeks following 14 days of baseline data collection with their usual therapy.
There were no episodes of severe hypoglycemia or diabetic ketoacidosis. By study end, HbA1c decreased by 0.55% (6.0 mmol/mol) (P < 0.0001). Time with sensor glucose levels in target range 70-180 mg/dL increased by 10.9%, or 2.6 h/day (P < 0.0001), while time with levels <70 mg/dL declined by median 0.27% (P = 0.0204).
Use of the automated insulin delivery system was safe, and participants experienced improved glycemic measures and reduced hypoglycemia during the study phase compared with baseline.