We propose to detect and to study neutrino neutral current coherent scattering off atomic nuclei with a two-phase emission detector using liquid xenon as a working medium. Expected signals and ...backgrounds are calculated for two possible experimental sites: the Kalinin Nuclear Power Plant in the Russian Federation and the Spallation Neutron Source at the Oak Ridge National Laboratory in the U.S.A. Both sites have advantages as well as limitations. The experiment looks feasible at either location.
A SensL MicroFC-SMT-60035 6×6 mm2 silicon photo-multiplier coated with a NOL-1 wavelength shifter have been tested in the liquid xenon to detect the 175-nm scintillation light. For comparison, a ...Hamamatsu vacuum ultraviolet sensitive MPPC VUV3 3×3 mm2 was tested under the same conditions. The photodetection efficiency of 13.1±2.5% and 6.0±1.0%, correspondingly, is obtained.
We present that the next-generation Enriched Xenon Observatory (nEXO) is a proposed experiment to search for neutrinoless double- β (0νββ) decay in 136Xe with a target half-life sensitivity of ...approximately 1028 yr using 5 × 103 kg of isotopically enriched liquid-xenon in a time projection chamber. This improvement of two orders of magnitude in sensitivity over current limits is obtained by a significant increase of the 136Xe mass, the monolithic and homogeneous configuration of the active medium, and the multiparameter measurements of the interactions enabled by the time projection chamber. Finally, the detector concept and anticipated performance are presented based upon demonstrated realizable background rates.
A xenon gas purity monitor for EXO Dobi, A.; Hall, C.; Herrin, S. ...
Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment,
12/2011, Letnik:
659, Številka:
1
Journal Article
Recenzirano
Odprti dostop
We discuss the design, operation, and calibration of two versions of a xenon gas purity monitor (GPM) developed for the EXO double beta decay program. The devices are sensitive to concentrations of ...oxygen well below 1ppb at an ambient gas pressure of one atmosphere or more. The theory of operation of the GPM is discussed along with the interactions of oxygen and other impurities with the GPM's tungsten filament. Lab tests and experiences in commissioning the EXO-200 double beta decay experiment are described. These devices can also be used on other noble gases.
Many extensions of the standard model of particle physics suggest that neutrinos should be Majorana-type fermions-that is, that neutrinos are their own anti-particles-but this assumption is difficult ...to confirm. Observation of neutrinoless double-β decay (0νββ), a spontaneous transition that may occur in several candidate nuclei, would verify the Majorana nature of the neutrino and constrain the absolute scale of the neutrino mass spectrum. Recent searches carried out with (76)Ge (the GERDA experiment) and (136)Xe (the KamLAND-Zen and EXO (Enriched Xenon Observatory)-200 experiments) have established the lifetime of this decay to be longer than 10(25) years, corresponding to a limit on the neutrino mass of 0.2-0.4 electronvolts. Here we report new results from EXO-200 based on a large (136)Xe exposure that represents an almost fourfold increase from our earlier published data sets. We have improved the detector resolution and revised the data analysis. The half-life sensitivity we obtain is 1.9 × 10(25) years, an improvement by a factor of 2.7 on previous EXO-200 results. We find no statistically significant evidence for 0νββ decay and set a half-life limit of 1.1 × 10(25) years at the 90 per cent confidence level. The high sensitivity holds promise for further running of the EXO-200 detector and future 0νββ decay searches with an improved Xe-based experiment, nEXO.
Celotno besedilo
Dostopno za:
DOBA, IJS, IZUM, KILJ, KISLJ, NUK, PILJ, PNG, SAZU, SIK, UILJ, UKNU, UL, UM, UPUK
The EXO-200 Collaboration is searching for neutrinoless double β decay using a liquid xenon (LXe) time projection chamber. This measurement relies on modeling the transport of charge deposits ...produced by interactions in the LXe to allow discrimination between signal and background events. Here we present measurements of the transverse diffusion constant and drift velocity of electrons at drift fields between 20 V/cm and 615 V/cm using EXO-200 data. At the operating field of 380 V/cm EXO-200 measures a drift velocity of 1.705+0.014 -0.010 mm/μs and a transverse diffusion coefficient of 55 ± 4 cm2/s.
A search for Lorentz- and CPT-violating signals in the double beta decay spectrum of super(136) Xe has been performed using an exposure of 100kgmiddotyr with the EXO-200 detector. No significant ...evidence of the spectral modification due to isotropic Lorentz-violation was found, and a two-sided limit of -2.65x10 super(-5)GeV(ProQuest: Formulae and/or non-USASCII text omitted) <7.60x10 super(-6)GeV (90% C.L.) is placed on the relevant coefficient within the Standard-Model Extension (SME). This is the first experimental study of the effect of the SME-defined oscillation-free and momentum-independent neutrino coupling operator on the double beta decay process.