Abstract
Aims
The Valve Academic Research Consortium (VARC), founded in 2010, was intended to (i) identify appropriate clinical endpoints and (ii) standardize definitions of these endpoints for ...transcatheter and surgical aortic valve clinical trials. Rapid evolution of the field, including the emergence of new complications, expanding clinical indications, and novel therapy strategies have mandated further refinement and expansion of these definitions to ensure clinical relevance. This document provides an update of the most appropriate clinical endpoint definitions to be used in the conduct of transcatheter and surgical aortic valve clinical research.
Methods and results
Several years after the publication of the VARC-2 manuscript, an in-person meeting was held involving over 50 independent clinical experts representing several professional societies, academic research organizations, the US Food and Drug Administration (FDA), and industry representatives to (i) evaluate utilization of VARC endpoint definitions in clinical research, (ii) discuss the scope of this focused update, and (iii) review and revise specific clinical endpoint definitions. A writing committee of independent experts was convened and subsequently met to further address outstanding issues. There were ongoing discussions with FDA and many experts to develop a new classification schema for bioprosthetic valve dysfunction and failure. Overall, this multi-disciplinary process has resulted in important recommendations for data reporting, clinical research methods, and updated endpoint definitions. New definitions or modifications of existing definitions are being proposed for repeat hospitalizations, access site-related complications, bleeding events, conduction disturbances, cardiac structural complications, and bioprosthetic valve dysfunction and failure (including valve leaflet thickening and thrombosis). A more granular 5-class grading scheme for paravalvular regurgitation (PVR) is being proposed to help refine the assessment of PVR. Finally, more specific recommendations on quality-of-life assessments have been included, which have been targeted to specific clinical study designs.
Conclusions
Acknowledging the dynamic and evolving nature of less-invasive aortic valve therapies, further refinements of clinical research processes are required. The adoption of these updated and newly proposed VARC-3 endpoints and definitions will ensure homogenous event reporting, accurate adjudication, and appropriate comparisons of clinical research studies involving devices and new therapeutic strategies.
Graphical Abstract
Owing to recent advances in computing power, iterative reconstruction (IR) algorithms have become a clinically viable option in computed tomographic (CT) imaging. Substantial evidence is accumulating ...about the advantages of IR algorithms over established analytical methods, such as filtered back projection. IR improves image quality through cyclic image processing. Although all available solutions share the common mechanism of artifact reduction and/or potential for radiation dose savings, chiefly due to image noise suppression, the magnitude of these effects depends on the specific IR algorithm. In the first section of this contribution, the technical bases of IR are briefly reviewed and the currently available algorithms released by the major CT manufacturers are described. In the second part, the current status of their clinical implementation is surveyed. Regardless of the applied IR algorithm, the available evidence attests to the substantial potential of IR algorithms for overcoming traditional limitations in CT imaging.
Fractional flow reserve (FFR) computation from coronary computed tomography angiography (CTA) datasets (FFRCT) has emerged as a promising noninvasive test to assess hemodynamic severity of coronary ...artery disease (CAD), but has not yet been compared with traditional functional imaging.
The purpose of this study was to evaluate the diagnostic performance of FFRCT and compare it with coronary CTA, single-photon emission computed tomography (SPECT), and positron emission tomography (PET) for ischemia diagnosis.
This subanalysis involved 208 prospectively included patients with suspected stable CAD, who underwent 256-slice coronary CTA, 99mTc-tetrofosmin SPECT, 15OH2O PET, and routine 3-vessel invasive FFR measurements. FFRCT values were retrospectively derived from the coronary CTA images. Images from each modality were interpreted by core laboratories, and their diagnostic performances were compared using invasively measured FFR ≤0.80 as the reference standard.
In total, 505 of 612 (83%) vessels could be evaluated with FFRCT. FFRCT showed a diagnostic accuracy, sensitivity, and specificity of 87%, 90%, and 86% on a per-vessel basis and 78%, 96%, and 63% on a per-patient basis, respectively. Area under the receiver-operating characteristic curve (AUC) for identification of ischemia-causing lesions was significantly greater for FFRCT (0.94 and 0.92) in comparison with coronary CTA (0.83 and 0.81; p < 0.01 for both) and SPECT (0.70 and 0.75; p < 0.01 for both), on a per-vessel and -patient level, respectively. FFRCT also outperformed PET on a per-vessel basis (AUC 0.87; p < 0.01), but not on a per-patient basis (AUC 0.91; p = 0.56). In the intention-to-diagnose analysis, PET showed the highest per-patient and -vessel AUC followed by FFRCT (0.86 vs. 0.83; p = 0.157; and 0.90 vs. 0.79; p = 0.005, respectively).
In this study, FFRCT showed higher diagnostic performance than standard coronary CTA, SPECT, and PET for vessel-specific ischemia, provided coronary CTA images were evaluable by FFRCT, whereas PET had a favorable performance in per-patient and intention-to-diagnose analysis. Still, in patients in whom 3-vessel FFRCT could be analyzed, FFRCT holds clinical potential to provide anatomic and hemodynamic significance of coronary lesions.
Display omitted
In a trial, patients who had undergone successful TAVR were assigned to rivaroxaban or antiplatelet therapy. In this substudy in patients who underwent CT, leaflet thickening and reduced leaflet ...motion at 90 days were less common with rivaroxaban. However, in the main trial, rivaroxaban was associated with a higher risk of death or thromboembolic complications and a higher risk of bleeding.
Abstract Computed tomography (CT) plays an important role in the workup of patients who are candidates for implantation of a catheter-based aortic valve, a procedure referred to as transcatheter ...aortic valve implantation (TAVI) or transcatheter aortic valve replacement (TAVR). Contrast-enhanced CT imaging provides information on the suitability of the peripheral access vessels to accommodate the relatively large sheaths necessary to introduce the prosthesis. CT imaging also provides accurate dimensions of the ascending aorta, aortic root, and aortic annulus which are of importance for prosthesis sizing, and initial data indicate that compared with echocardiographic sizing, CT-based sizing of the prosthesis may lead to better results for postprocedural aortic valve regurgitation. Finally, CT permits one to predict appropriate fluoroscopic projections which are oriented orthogonal to the aortic valve plane. This consensus document provides recommendations about the use of CT imaging in patients scheduled for TAVR/TAVI, including data acquisition, interpretation, and reporting.