Despite continuous technological developments, transvenous pacemakers (PM) are still associated with significant immediate and long-term complications, mostly lead or pocket-related. Recent ...technological advances brought to the introduction in clinical practice of leadless PM for selected cohort of patients. These miniaturize devices are implanted through the femoral vein and advanced to the right ventricle, without leaving leads in place. Lack of upper extremity vascular access and/or high infective risk in patients requiring VVI pacing are the most common indications to leadless PM. The recently introduced MICRA AV leadless PM also allows ventricular synchronization through mechanical sensing of atrial contraction waves, thus solving the problem of AV synchronization. This review will discuss and summarize available clinical evidence on leadless PM, their performance compared to transvenous devices, current applications and future perspectives.
Objectives The goal of this study was to compare transesophageal echocardiography (TEE) and intracardiac echocardiography (ICE) for the diagnosis of cardiac device–related endocarditis (CDI). ...Background The diagnosis of infective endocarditis (IE) was established by using the modified Duke criteria based mainly on echocardiography and blood culture results. No previous studies have compared ICE with TEE for the diagnosis of IE. Methods We prospectively enrolled 162 patients (age 72 ± 11 years; 125 male) who underwent transvenous lead extraction: 152 with CDI and 10 with lead malfunction (control group). Using the modified Duke criteria, we divided the patients with infection into 3 groups: 44 with a “definite” diagnosis of IE (group 1), 52 with a “possible” diagnosis of IE (group 2), and 56 with a “rejected” diagnosis of IE (group 3). TEE and ICE were performed before the procedure. Results In group 1, ICE identified intracardiac masses (ICM) in all 44 patients; TEE identified ICM in 32 patients (73%). In group 2, 6 patients (11%) had ICE and TEE both positive for ICM, 8 patients (15%) had a negative TEE but a positive ICE, and 38 patients (73%) had ICE and TEE both negative. In group 3, 2 patients (3%) had ICM both at ICE and TEE, 1 patient (2%) had an ICM at ICE and a negative TEE, and 53 patients (95%) had no ICM at ICE and TEE. ICE and TEE were both negative in the control group. Conclusions ICE represents a useful technique for the diagnosis of ICM, thus providing improved imaging of right-sided leads and increasing the diagnostic yield compared with TEE.
Little is known about the prognostic significance of non-sustained ventricular tachycardia (NS-VT) in outpatients scheduled for routine pacemaker controls. We therefore sought to investigate the ...prognostic significance of non-sustained ventricular tachycardia on stored electrograms in pacemaker recipients.
We enrolled patients implanted with dual chamber pacemaker for atrioventricular block or sinus node dysfunction from 2010 to 2016, with LVEF> 45%, older than 18 years, with at least 3 device interrogations at follow-up. Data were collected about medical history, pharmacological therapy at implantation, pacemaker programming, NS-VT occurrence, long-term survival.
A total of 308 patients were included in the final analysis, with median follow-up time of 56 months. No ventricular arrhythmic episodes were documented in 221 patients (Group 1), whereas 87 had at least 1 episode of NS-VT during follow-up (Group 2). As a whole, 282 episodes of NS-VT were documented. There was a higher prevalence of previous myocardial infarction and slightly lower left ventricular ejection fraction (LVEF) in Group 2. The primary endpoint (all-cause mortality) occurred in 50 patients (22%) of Group 1 and 12 (14%) patients of Group 2 (p = 0.07). Clinical predictors of all-cause mortality at univariate analysis included age, LVEF and coronary artery disease (CAD). Only age and CAD, however, remained as predictors of mortality at multivariable analysis. A sizeable, but not statistically significant, portion of patients who died had a de novo occurrence of NS-VT at the last pacemaker check.
Our data do not support a prognostic role for the detection of NS-VT during pacemaker controls.