Within the ORAMED project (Optimization of Radiation Protection of Medical Staff) a coordinated measurement program for occupationally exposed medical staff was performed in different hospitals in ...Europe (
www.oramed-fp7.eu). The main objective was to obtain a set of standardized data on extremity and eye lens doses for staff involved in interventional radiology and cardiology and to optimize radiation protection. Special attention was given to the measurement of the doses to the eye lenses. In this paper an overview will be given of the measured eye lens doses and the main influence factors for these doses. The measured eye lens doses are extrapolated to annual doses. The extrapolations showed that monitoring of the eye lens should be performed on routine basis.
► Eye lens dose limit of 150 mSv is generally not reached in IR/IC. ► Monitoring of eye lens dose is recommended (except for ERCP) in IR/IC. ► New eye lens dose limit of 20 mSv will be surpassed by many physicians in IR/IC.
Within the ORAMED project a coordinated measurement program for occupationally exposed medical staff was performed in different hospitals in Europe. The main objectives of ORAMED were to obtain a set ...of standardized data on doses for staff in interventional cardiology and radiology and to optimize staff protection. Doses were measured with thermoluminescent dosemeters on the ring finger and wrist of both hands, on legs and at the level of the eyes of the main operator performing interventional procedures. In this paper an overview of the doses per procedure measured during 646 interventional cardiology procedures is given for cardiac angiographies and angioplasties (CA/PTCA), radiofrequency ablations (RFA) and pacemaker and defibrillator implantations (PM/ICD). 31% of the monitored procedures were associated with no collective protective equipment, whereas 44% involved a ceiling screen and a table curtain. Although associated with the smallest air kerma – area product (KAP), PM/ICD procedures led to the highest doses. As expected, KAP and doses values exhibited a very large variability. The left side of the operator, most frequently the closest to the X-ray scattering region, was more exposed than his right side. An analysis of the effect of parameters influencing the doses, namely collective protective equipment, X-ray tube configuration and catheter access route, was performed on the doses normalized to KAP. Ceiling screen and table curtain were observed to reduce normalized doses by atmost a factor 4, much smaller than theoretical attenuation factors typical for such protections, i.e. from 10 to 100. This observation was understood as their inappropriate use by the operators and their non-optimized design. Configurations with tube above the patient led to higher normalized doses to the operator than tube below, but the effect of using a biplane X-ray suite was more complex to analyze. For CA/PTCA procedures, the upper part of the operator’s body received higher normalized doses for radial than for femoral catheter access, by atmost a factor 5. This could be seen for cases with no collective protection. The eyes were observed to receive the maximum fraction of the annual dose limit almost as frequently as legs and hands, and clearly the most frequently, if the former 150 mSv and new 20 mSv recommended limits for the lens of the eye are considered, respectively.
► 643 interventional cardiology procedures were monitored. ► In 31% of the cases no collective protective equipment was used. ► The left side of the operator is generally the most exposed. ► Effect of collective protective equipment, tube configuration and catheter access on doses is studied. ► The eyes receive most frequently the largest fraction of the annual dose limit if the new 20 mSv limit is considered.
The main aim of the Work Package 1 (WP1) of the ORAMED project, Collaborative Project (2008–2011), supported by the European Commission within its 7th Framework Programme, was to obtain a set of ...standardized data on extremity and eye lens doses for staff in interventional radiology and cardiology (IR/IC) workplaces and to recommend a series of guidelines on radiation protection in order to both guarantee and optimize staff protection. Within the project, coordinated measurements were performed in 34 hospitals in 6 European countries. Furthermore, simulations of the most representative workplaces in IR and IC were performed to determine the main parameters that influence the extremity and eye lens doses. The work presented in this paper shows the recommendations that were formulated by the results obtained from both measurements and simulations. The presented guidelines are directed to operators, assistant personnel, radiation protection officers and medical physics experts. They concern radiation protection issues, such as the use of room protective equipment, as well as the positioning of the extremity and eye lens dosemeters for routine monitoring.
The article reports results from the largest international dose survey in paediatric computed tomography (CT) in 32 countries and proposes international diagnostic reference levels (DRLs) in terms of ...computed tomography dose index (CTDI vol) and dose length product (DLP). It also assesses whether mean or median values of individual facilities should be used. A total of 6115 individual patient data were recorded among four age groups: <1 y, >1-5 y, >5-10 y and >10-15 y. CTDIw, CTDI vol and DLP from the CT console were recorded in dedicated forms together with patient data and technical parameters. Statistical analysis was performed, and international DRLs were established at rounded 75th percentile values of distribution of median values from all CT facilities. The study presents evidence in favour of using median rather than mean of patient dose indices as the representative of typical local dose in a facility, and for establishing DRLs as third quartile of median values. International DRLs were established for paediatric CT examinations for routine head, chest and abdomen in the four age groups. DRLs for CTDI vol are similar to the reference values from other published reports, with some differences for chest and abdomen CT. Higher variations were observed between DLP values, based on a survey of whole multi-phase exams. It may be noted that other studies in literature were based on single phase only. DRLs reported in this article can be used in countries without sufficient medical physics support to identify non-optimised practice. Recommendations to improve the accuracy and importance of future surveys are provided.
An intensive measurement campaign was launched in different hospitals in Europe within work package 1 of the ORAMED project (Optimization of RAdiation protection for MEDical staff). Its main ...objective was to obtain a set of standardized data on extremity and eye lens doses for staff in interventional radiology (IR) and cardiology (IC) and to optimize staff protection. The monitored procedures were divided in three main categories: cardiac, general angiography and endoscopic retrograde cholangio-pancreatography(ERCP) procedures. Using a common measurement protocol, information such as the protective equipment used (lead table curtain, transparent lead glass ceiling screen, patient shielding, whole body shielding or special cabin etc.) as well as Kerma Area Product (KAP) values and access of the catheter were recorded.
This study was performed with a final database of more than 1300 procedures performed in 34 European hospitals. Its objectives were firstly to determine if the measured extremity doses could be correlated to the KAP values; secondly to check if the doses to the eyes could be linked to the doses to the hands (finger or wrist positions) and finally if the doses to the fingers could be estimated based on the doses to the wrists. General correlations were very difficult to find and their strength was mostly influenced by three main parameters: the X-ray tube configuration, the room collective radioprotective equipment and the access of the catheter. The KAP value can provide a simple mean to estimate the extremity doses of the operator given that it is assessed correctly for the operator when he is actually using the X-ray tube. Moreover, this study showed that the doses to the left finger are strongly correlated to the doses to the left wrist when no ceiling shield is used. It is also possible to estimate the doses to the eyes given the doses to the left finger or left wrist but the X-ray tube configuration and the access have to be considered.
The annual finger doses of three workers were estimated and were found to exceed the 3/10thof the annual limit. Therefore, for the hands routine monitoring is necessary especially for interventional cardiology procedures, except for ERCP procedures. If the table shield is properly used, leg doses can be reduced significantly and no routine monitoring is necessary.
► Data on extremity and eye lens doses was collected in order to optimize staff protection. Correlations were investigated and extremity annual doses were estimated. ► If assessed correctly, the Kerma area product value can provide a simple mean to estimate the extremity doses of the operator. ► For certain X-ray tube configuration and access, doses to the left finger or left wrist can help estimating eye lens doses. ► Hands should be routinely monitored whereas it is not necessary for the legs if they are properly shielded.
The introduction of interventional radiology (IR) procedures in the 20th century has demonstrated significant advantages over surgery procedures. As a result, their number is continuously rising in ...diagnostic, as well as, in therapy field and is connected with progress in highly sophisticated equipment used for these purposes. Nowadays, in the European countries more than 400 fluoroscopically guided IR procedures were identified with a 10–12% increase in the number of IR examinations every year (
UNSCEAR, 2010). Depending on the complexity of the different types of the interventions large differences in the radiation doses of the staff are observed.
The staff that carries out IR procedures is likely to receive relatively high radiation doses, because IR procedures require the operator to remain close to the patient and close to the primary radiation beam. In spite of the fact that the operator is shielded by protective apron, the hands, eyes and legs remain practically unshielded. For this reason, one of the aims of the ORAMED project was to provide a set of standardized data on extremity doses for the personnel that are involved in IR procedures and to optimize their protection by evaluating the various factors that affect the doses. In the framework of work package 1 of the ORAMED project the impact of protective equipment, tube configuration and access routes were analyzed for the selected IR procedures. The position of maximum dose measured is also investigated. The results of the extremity doses in IR workplaces are presented in this study together with the influence of the above mentioned parameters on the doses.
► We present a set of data on extremity doses for staff in selected interventional radiology procedures. ► We studied the influence of different parameters. ► The measured doses are analyzed according to the operators skill,his position during work, tube configuration, etc. ► Maximum doses recorded for all types of embolisation, in all locations of dosemeters on the staff body. ► The results were compared with established dose limits and extremity doses published by another authors.
The main objective of WP1 of the ORAMED (Optimization of RAdiation protection for MEDical staff) project is to obtain a set of standardised data on extremity and eye lens doses for staff in ...interventional radiology (IR) and cardiology (IC) and to optimise staff protection. A coordinated measurement program in different hospitals in Europe will help towards this direction. This study aims at analysing the first results of the measurement campaign performed in IR and IC procedures in 34 European hospitals. The highest doses were found for pacemakers, renal angioplasties and embolisations. Left finger and wrist seem to receive the highest extremity doses, while the highest eye lens doses are measured during embolisations. Finally, it was concluded that it is difficult to find a general correlation between kerma area product and extremity or eye lens doses.
Patient doses for a few common fluoroscopy-guided procedures in interventional radiology (IR) (excluding cardiology) were collected from a few radiological departments in 13 European countries. The ...major aim was to evaluate patient doses for the basis of the reference levels. In total, data for 20 procedures for about 1300 patients were collected. There were many-fold variations in the number of IR equipment and procedures per population, in the entrance dose rates, and in the patient dose data (total dose area product or DAP, fluoroscopy time and number of frames). There was no clear correlation between the total DAP and entrance dose rate, or between the total DAP and fluoroscopy time, indicating that a number of parameters affect the differences. Because of the limited number of patients, preliminary reference levels were proposed only for a few procedures. There is a need to improve the optimisation of IR procedures and their definitions and grouping, in order to account for their different complexities.
In interventional radiology, for an accurate determination of effective dose to the staff, measurements with two dosemeters have been recommended, one located above and one under the protective ...apron. Such ‘double dosimetry’ practices and the algorithms used for the determination of effective dose were reviewed in this study by circulating a questionnaire and by an extensive literature search. The results indicated that regulations for double dosimetry almost do not exist and there is no firm consensus on the most suitable calculation algorithms. The calculation of effective dose is mainly based on the single dosemeter measurements, in which either personal dose equivalent, directly, (dosemeter below the apron) or a fraction of personal dose equivalent (dosemeter above the apron) is taken as an assessment of effective dose. The most recent studies suggest that there might not be just one double dosimetry algorithm that would be optimum for all interventional radiology procedures. Further investigations in several critical configurations of interventional radiology procedures are needed to assess the suitability of the proposed algorithms.
The combination of fluoroscopically guided interventional procedures with computed tomography (CTF) has become widespread around the world. The benefits of CTF include the ability to obtain a ...real-time visualization of the entire body, increased target accuracy and improved visualization of biopsy needles. Modern CTF units work with variable frame rates for image selection, and therefore the dose distributions for patients and staff can considerably vary, creating growing concern in terms of the occupational exposure of interventionists and the drawback of a higher exposure of the patient. A literature review of the latest CTF publications is summarized in this article. A wide range of CTF studies reveal different treatment methods used in clinical practice, and therefore the differences in the exposures between them; as well as in the radiation protection tools and dose monitoring. Further optimization of radiation protection methods, harmonization of exposure patterns as well as training and education of CTF staff on the basis of the information in the survey, are strongly recommended.