Monitoring of regional lung function in interventional COPD trials requires alternative endpoints beyond global parameters such as FEV1. T1 relaxation times of the lung might allow to draw ...conclusions on tissue composition, blood volume and oxygen fraction. The aim of this study was to evaluate the potential value of lung Magnetic resonance imaging (MRI) with native and oxygen-enhanced T1 mapping for the assessment of COPD patients in comparison with contrast enhanced perfusion MRI.
20 COPD patients (GOLD I-IV) underwent a coronal 2-dimensional inversion recovery snapshot flash sequence (8 slices/lung) at room air and during inhalation of pure oxygen, as well as dynamic contrast-enhanced first-pass perfusion imaging. Regional distribution of T1 at room air (T1), oxygen-induced T1 shortening (ΔT1) and peak enhancement were rated by 2 chest radiologists in consensus using a semi-quantitative 3-point scale in a zone-based approach.
Abnormal T1 and ΔT1 were highly prevalent in the patient cohort. T1 and ΔT1 correlated positively with perfusion abnormalities (r = 0.81 and r = 0.80; p&0.001), and with each other (r = 0.80; p<0.001). In GOLD stages I and II ΔT1 was normal in 16/29 lung zones with mildly abnormal perfusion (15/16 with abnormal T1). The extent of T1 (r = 0.45; p<0.05), ΔT1 (r = 0.52; p<0.05) and perfusion abnormalities (r = 0.52; p<0.05) showed a moderate correlation with GOLD stage.
Native and oxygen-enhanced T1 mapping correlated with lung perfusion deficits and severity of COPD. Under the assumption that T1 at room air correlates with the regional pulmonary blood pool and that oxygen-enhanced T1 reflects lung ventilation, both techniques in combination are principally suitable to characterize ventilation-perfusion imbalance. This appears valuable for the assessment of regional lung characteristics in COPD trials without administration of i.v. contrast.
Celotno besedilo
Dostopno za:
DOBA, IZUM, KILJ, NUK, PILJ, PNG, SAZU, SIK, UILJ, UKNU, UL, UM, UPUK
To apply the MB (multiband) excitation and blipped-CAIPI (blipped-controlled aliasing in parallel imaging) techniques in a spin and gradient-echo (SAGE) EPI sequence to improve the slice coverage for ...vessel architecture imaging (VAI).
Both MB excitation and blipped-CAIPI with in-plane parallel imaging were incorporated into a gradient-echo (GE)/spin-echo (SE) EPI sequence for simultaneous tracking of the dynamic MR signal changes in both GE and SE contrasts after the injection of contrast agent. MB and singleband (SB) excitation were compared using a 20-channel head coil at 3 Tesla, and high-resolution MB VAI could be performed in 32 glioma patients.
Whole-brain covered high resolution VAI can be achieved after applying multiband excitation with a factor of 2 and in-plane parallel imaging with a factor of 3. The quality of the images resulting from MB acceleration was comparable to those from the SB method: images were reconstructed without any loss of spatial resolution or severe distortions. In addition, MB and SB signal-to-noise ratios (SNR) were similar. A relative low g-factor induced from the MB acceleration method was achieved after using a blipped-CAIPI technique (1.35 for GE and 1.33 for SE imaging). Performing quantitative VAI, we found that, among all VAI parametric maps, microvessel type indicator (MTI), distance map (I) and vascular-induced bolus peak-time shift (VIPS) were highly correlated. Likewise, VAI parametric maps of slope, slope length and short axis were highly correlated.
Multiband accelerated SAGE successfully doubles the number of readout slices in the same measurement time when compared to conventional readout sequences. The corresponding VAI parametric maps provide insights into the complexity and heterogeneity of vascular changes in glioma.
Celotno besedilo
Dostopno za:
DOBA, IZUM, KILJ, NUK, PILJ, PNG, SAZU, SIK, UILJ, UKNU, UL, UM, UPUK
Magnetic Resonance Imaging of Lung Perfusion Triphan, Simon M.F.; Bauman, Grzegorz; Konietzke, Philip ...
Journal of magnetic resonance imaging,
March 2024, 2024-Mar, 2024-03-00, 20240301, Letnik:
59, Številka:
3
Journal Article
Recenzirano
Odprti dostop
“Lung perfusion” in the context of imaging conventionally refers to the delivery of blood to the pulmonary capillary bed through the pulmonary arteries originating from the right ventricle required ...for oxygenation. The most important physiological mechanism in the context of imaging is the so‐called hypoxic pulmonary vasoconstriction (HPV, also known as “Euler‐Liljestrand‐Reflex”), which couples lung perfusion to lung ventilation. In obstructive airway diseases such as asthma, chronic‐obstructive pulmonary disease (COPD), cystic fibrosis (CF), and asthma, HPV downregulates pulmonary perfusion in order to redistribute blood flow to functional lung areas in order to conserve optimal oxygenation. Imaging of lung perfusion can be seen as a reflection of lung ventilation in obstructive airway diseases. Other conditions that primarily affect lung perfusion are pulmonary vascular diseases, pulmonary hypertension, or (chronic) pulmonary embolism, which also lead to inhomogeneity in pulmonary capillary blood distribution. Several magnetic resonance imaging (MRI) techniques either dependent on exogenous contrast materials, exploiting periodical lung signal variations with cardiac action, or relying on intrinsic lung voxel attributes have been demonstrated to visualize lung perfusion. Additional post‐processing may add temporal information and provide quantitative information related to blood flow. The most widely used and robust technique, dynamic‐contrast enhanced MRI, is available in clinical routine assessment of COPD, CF, and pulmonary vascular disease. Non‐contrast techniques are important research tools currently requiring clinical validation and cross‐correlation in the absence of a viable standard of reference. First data on many of these techniques in the context of observational studies assessing therapy effects have just become available.
Level of Evidence
5
Technical Efficacy
Stage 5
4D perfusion magnetic resonance imaging (MRI) with intravenous injection of contrast agent allows for a radiation-free assessment of regional lung function. It is therefore a valuable method to ...monitor response to treatment in patients with chronic obstructive pulmonary disease (COPD). This study was designed to evaluate its potential for monitoring short-term response to hyperoxia in COPD patients.
19 prospectively enrolled COPD patients (median age 66y) underwent paired dynamic contrast-enhanced 4D perfusion MRI within 35min, first breathing 100% oxygen (injection 1, O2) and then room air (injection 2, RA), which was repeated on two consecutive days (day 1 and 2). Post-processing software was employed to calculate mean transit time (MTT), pulmonary blood volume (PBV) and pulmonary blood flow (PBF), based on the indicator dilution theory, for the automatically segmented whole lung and 12 regions of equal volume.
Comparing O2 with RA conditions, PBF and PBV were found to be significantly lower at O2, consistently on both days (p<10-8). Comparing day 2 to day 1, MTT was shorter by 0.59±0.63 s (p<10-8), PBF was higher by 22±80 ml/min/100ml (p<3·10-4), and PBV tended to be lower by 0.2±7.2 ml/100ml (p = 0.159) at both, RA and O2, conditions.
The second injection (RA) yielded higher PBF and PBV, which apparently contradicts the established hypothesis that hyperoxia increases lung perfusion. Quantification of 4D perfusion MRI by current software approaches may thus be limited by residual circulating contrast agent in the short-term and even the next day.
Celotno besedilo
Dostopno za:
DOBA, IZUM, KILJ, NUK, PILJ, PNG, SAZU, SIK, UILJ, UKNU, UL, UM, UPUK
As there is an increasing number of multicentre lung imaging studies with MRI in patients, dedicated reference phantoms are required to allow for the assessment and comparison of image quality in ...multi-vendor and multi-centre environments. However, appropriate phantoms for this purpose are so far not available commercially. It was therefore the purpose of this project to design and apply a cost-effective and simple to use reference phantom which addresses the specific requirements for imaging the lungs with MRI.
The phantom was designed to simulate 4 compartments (lung, blood, muscle and fat) which reflect the specific conditions in proton-MRI of the chest. Multiple phantom instances were produced and measured at 15 sites using a contemporary proton-MRI protocol designed for an in vivo COPD study at intervals over the course of the study. Measures of signal- and contrast-to-noise ratio, as well as structure and edge depiction were extracted from conventionally acquired images using software written for this purpose.
For the signal to noise ratio, low intra-scanner variability was found with 4.5% in the lung compartment, 4.0% for blood, 3.3% for muscle and 3.7% for fat. The inter-scanner variability was substantially higher, with 41%, 32%, 27% and 32% for the same order of compartments. In addition, measures of structure and edge depiction were found to both vary significantly among several scanner types and among scanners of the same model which were equipped with different gradient systems.
The described reference phantom reproducibly quantified image quality aspects and detected substantial inter-scanner variability in a typical pulmonary multicentre proton MRI study, while variability was greater in lung tissue compared to other tissue types. Accordingly, appropriate reference phantoms can help to detect bias in multicentre in vivo study results and could also be used to harmonize equipment or data.
Celotno besedilo
Dostopno za:
DOBA, IZUM, KILJ, NUK, PILJ, PNG, SAZU, SIK, UILJ, UKNU, UL, UM, UPUK
Objectives
Pulmonary perfusion abnormalities are prevalent in patients with chronic obstructive pulmonary disease (COPD), are potentially reversible, and may be associated with emphysema development. ...Therefore, we aimed to evaluate the clinical meaningfulness of perfusion defects in percent (QDP) using DCE-MRI.
Methods
We investigated a subset of baseline DCE-MRIs, paired inspiratory/expiratory CTs, and pulmonary function testing (PFT) of 83 subjects (age = 65.7 ± 9.0 years, patients-at-risk, and all GOLD groups) from one center of the “COSYCONET” COPD cohort. QDP was computed from DCE-MRI using an in-house developed quantification pipeline, including four different approaches: Otsu’s method, k-means clustering, texture analysis, and 80
th
percentile threshold. QDP was compared with visual MRI perfusion scoring, CT parametric response mapping (PRM) indices of emphysema (PRM
Emph
) and functional small airway disease (PRM
fSAD
), and FEV1/FVC from PFT.
Results
All QDP approaches showed high correlations with the MRI perfusion score (
r
= 0.67 to 0.72,
p
< 0.001), with the highest association based on Otsu’s method (
r
= 0.72,
p
< 0.001). QDP correlated significantly with all PRM indices (
p
< 0.001), with the strongest correlations with PRM
Emph
(
r
= 0.70 to 0.75,
p
< 0.001). QDP was distinctly higher than PRM
Emph
(mean difference = 35.85 to 40.40) and PRM
fSAD
(mean difference = 15.12 to 19.68), but in close agreement when combining both PRM indices (mean difference = 1.47 to 6.03) for all QDP approaches. QDP correlated moderately with FEV1/FVC (
r
= − 0.54 to − 0.41,
p
< 0.001).
Conclusion
QDP is associated with established markers of disease severity and the extent corresponds to the CT-derived combined extent of PRM
Emph
and PRM
fSAD
. We propose to use QDP based on Otsu’s method for future clinical studies in COPD.
Key Points
• QDP quantified from DCE-MRI is associated with visual MRI perfusion score, CT PRM indices, and PFT.
• The extent of QDP from DCE-MRI corresponds to the combined extent of PRM
Emph
and PRM
fSAD
from CT.
• Assessing pulmonary perfusion abnormalities using DCE-MRI with QDP improved the correlations with CT PRM indices and PFT compared to the quantification of pulmonary blood flow and volume.
Purpose
Magnetic resonance imaging (MRI) of the lung can be used for diagnosis and monitoring of interstitial lung disease. Biophysical models of alveolar lung tissue are needed to understand the ...complex interplay of susceptibility, diffusion, and flow effects, and their influence on magnetic resonance (MR) spin dephasing.
Methods
In this work, we present a method for modeling the signal decay of lung tissue by utilizing a two‐compartment model, which considers the different spin dephasing mechanisms in the alveolar vasculature and interstitial tissue. This allows calculating the magnetization dynamics and the MR lineshape, which can be measured noninvasively using clinical MR scanners.
Results
The accuracy of the method was evaluated using finite element simulations and the experimentally measured lineshapes of a healthy volunteer. In this comparison, the model performs well, indicating that the relevant spin dephasing mechanisms are correctly taken into account.
Conclusions
The proposed method can be used to estimate the influence of blood flow and alveolar geometry on the MR lineshape of lung tissue.
Non-invasive end-points for interventional trials and tailored treatment regimes in chronic obstructive pulmonary disease (COPD) for monitoring regionally different manifestations of lung disease ...instead of global assessment of lung function with spirometry would be valuable. Proton nuclear magnetic resonance imaging (1H-MRI) allows for a radiation-free assessment of regional structure and function. The aim of this study was to evaluate the short-term reproducibility of a comprehensive morpho-functional lung MRI protocol in COPD.
20 prospectively enrolled COPD patients (GOLD I-IV) underwent 1H-MRI of the lung at 1.5T on two consecutive days, including sequences for morphology, 4D contrast-enhanced perfusion, and respiratory mechanics. Image quality and COPD-related morphological and functional changes were evaluated in consensus by three chest radiologists using a dedicated MRI-based visual scoring system. Test-retest reliability was calculated per each individual lung lobe for the extent of large airway (bronchiectasis, wall thickening, mucus plugging) and small airway abnormalities (tree in bud, peripheral bronchiectasis, mucus plugging), consolidations, nodules, parenchymal defects and perfusion defects. The presence of tracheal narrowing, dystelectasis, pleural effusion, pulmonary trunk ectasia, right ventricular enlargement and, finally, motion patterns of diaphragma and chest wall were addressed.
Median global scores 10(Q1:8.00;Q3:16.00) vs.11(Q1:6.00;Q3:15.00) as well as category subscores were similar between both timepoints, and kappa statistics indicated "almost perfect" global agreement (ĸ = 0.86, 95%CI = 0.81-0.91). Most subscores showed at least "substantial" agreement of MRI1 and MRI2 (ĸ = 0.64-1.00), whereas the agreement for the diagnosis of dystelectasis/effusion (ĸ = 0.42, 95%CI = 0.00-0.93) was "moderate" and of tracheal abnormalities (ĸ = 0.21, 95%CI = 0.00-0.75) "fair". Most MRI acquisitions showed at least diagnostic quality at MRI1 (276 of 278) and MRI2 (259 of 264).
Morpho-functional 1H-MRI can be obtained with reproducible image quality and high short-term test-retest reliability for COPD-related morphological and functional changes of the lung. This underlines its potential value for the monitoring of regional lung characteristics in COPD trials.
Celotno besedilo
Dostopno za:
DOBA, IZUM, KILJ, NUK, PILJ, PNG, SAZU, SIK, UILJ, UKNU, UL, UM, UPUK
Outracing Lung Signal Decay – Potential of Ultrashort Echo Time MRI Wielpütz, Mark Oliver; Triphan, Simon M. F.; Ohno, Yoshiharu ...
RöFo : Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebende Verfahren,
05/2019, Letnik:
191, Številka:
5
Journal Article
Recenzirano
Odprti dostop
Abstract
Background
Magnetic resonance imaging (MRI) of the pulmonary parenchyma is generally hampered by multiple challenges related to patient respiratory- and circulation-related motion, low ...proton density and extremely fast signal decay due to the structure of the lungs evolved for gas exchange.
Methods
Systematic literature database research as well as annual participation in conferences dedicated to pulmonary MRI for more than the past 20 years by at least one member of the author team.
Results and Conclusion
The problem of motion has been addressed in the past by developments such as triggering, gating and parallel imaging. The second problem has, in part, turned out to be an advantage in those diseases that lead to an increase in lung substance and thus an increase in signal relative to the background. To reduce signal decay, ultrashort echo time (UTE) methods were developed to minimize the time between excitation and readout. Having been postulated a while ago, improved hardware and software now open up the possibility of achieving echo times shorter than 200 µs, increasing lung signal significantly by forestalling signal decay and more effectively using the few protons available. Such UTE techniques may not only improve structural imaging of the lung but also enhance functional imaging, including ventilation and perfusion imaging as well as quantitative parameter mapping. Because of accelerating progress in this field of lung MRI, the review at hand seeks to introduce some technical properties as well as to summarize the growing data from applications in humans and disease, which promise that UTE MRI will play an important role in the morphological and functional assessment of the lung in the near future.
Key Points:
Ultrashort echo time MRI is technically feasible with state-of-the-art scanner hardware.
UTE MRI allows for CT-like image quality for structural lung imaging.
Preliminary studies show improvements over conventional morphological imaging in lung cancer and airways diseases.
UTE may improve sensitivity for functional processes like perfusion and tissue characterization.
Citation Format
Wielpütz MO, Triphan SM, Ohno Y et al. Outracing Lung Signal Decay – Potential of Ultrashort Echo Time MRI. Fortschr Röntgenstr 2018; DOI: 10.1055/a-0715-2246