A detection system of the conventional PET tomograph is set-up to record data from annihilation into two photons with energy of 511 keV, and it gives information on the density distribution of a ...radiopharmaceutical in the body of the object. In this paper we explore the possibility of performing the three gamma photons imaging based on ortho-positronium annihilation, as well as the possibility of positronium mean lifetime imaging with the J-PET tomograph constructed from plastic scintillators. For this purposes simulations of the ortho-positronium formation and its annihilation into three photons were performed taking into account distributions of photons' momenta as predicted by the theory of quantum electrodynamics and the response of the J-PET tomograph. In order to test the proposed ortho-positronium lifetime image reconstruction method, we concentrate on the decay of the ortho-positronium into three photons and applications of radiopharmaceuticals labeled with isotopes emitting a prompt gamma. The proposed method of imaging is based on the determination of hit-times and hit-positions of registered photons which enables the reconstruction of the time and position of the annihilation point as well as the lifetime of the ortho-positronium on an event-by-event basis. We have simulated the production of the positronium in point-like sources and in a cylindrical phantom composed of a set of different materials in which the ortho-positronium lifetime varied from 2.0 ns to 3.0 ns, as expected for ortho-positronium created in the human body. The presented reconstruction method for total-body J-PET like detector allows to achieve a mean lifetime resolution of 40 ps. Recent positron annihilation lifetime spectroscopy measurements of cancerous and healthy uterine tissues show that this sensitivity may allow to study the morphological changes in cell structures.
A novel whole-body positron emission tomography (PET) system based on plastic scintillators is developed by the J-PET Collaboration. It consists of plastic scintillator strips arranged axially in the ...form of a cylinder, allowing the cost-effective construction of the total-body PET system. In order to determine the properties of the scanner prototype and optimize its geometry, advanced computer simulations were performed using the GATE (Geant4 application for tomographic emission) software. The spatial resolution, sensitivity, scatter fraction and noise equivalent count rate were estimated according to the National Electrical Manufacturers Association norm, as a function of the length of the tomograph, the number of detection layers, the diameter of the tomographic chamber and for various types of applied readout. For the single-layer geometry with a diameter of 85 cm, a strip length of 100 cm, a cross-section of 4 mm × 20 mm and silicon photomultipliers with an additional layer of wavelength shifter as the readout, the spatial resolution (full width at half maximum) in the centre of the scanner is equal to 3 mm (radial, tangential) and 6 mm (axial). For the analogous double-layer geometry with the same readout, diameter and scintillator length, with a strip cross-section of 7 mm × 20 mm, a noise equivalent count rate peak of 300 kcps was reached at 40 kBq cc−1 activity concentration, the scatter fraction is estimated to be about 35% and the sensitivity at the centre amounts to 14.9 cps kBq−1. Sensitivity profiles were also determined.
We present a study of the application of the Jagiellonian positron emission tomograph (J-PET) for the registration of gamma quanta from decays of ortho-positronium (o-Ps). The J-PET is the first ...positron emission tomography scanner based on organic scintillators in contrast to all current PET scanners based on inorganic crystals. Monte Carlo simulations show that the J-PET as an axially symmetric and high acceptance scanner can be used as a multi-purpose detector well suited to pursue research including e.g. tests of discrete symmetries in decays of ortho-positronium in addition to the medical imaging. The gamma quanta originating from o-Ps decay interact in the plastic scintillators predominantly via the Compton effect, making the direct measurement of their energy impossible. Nevertheless, it is shown in this paper that the J-PET scanner will enable studies of the
o-Ps
→
3
γ
decays with angular and energy resolution equal to
σ
(
θ
)
≈
0
.
4
∘
and
σ
(
E
)
≈
4.1
keV
, respectively. An order of magnitude shorter decay time of signals from plastic scintillators with respect to the inorganic crystals results not only in better timing properties crucial for the reduction of physical and instrumental background, but also suppresses significantly the pile-ups, thus enabling compensation of the lower efficiency of the plastic scintillators by performing measurements with higher positron source activities.
Celotno besedilo
Dostopno za:
DOBA, IZUM, KILJ, NUK, PILJ, PNG, SAZU, SIK, UILJ, UKNU, UL, UM, UPUK
Reactive octahedral silsesquioxanes of rod-like octakis(3-chloropropyl)octasilsesquioxane - T
(CH
CH
CH
Cl)
and spherical octavinyloctasilsesquioxane - T
(CHdouble bond, length as m-dashCH
)
...structure can undergo reversible thermally induced phase transitions in the solid state. The phase behaviour has been studied with differential scanning calorimetry (DSC, including temperature modulated DSC), X-ray diffraction, dielectric relaxation spectroscopy (DRS), and nuclear magnetic resonance spectroscopy in the solid state (SS NMR), as well as positron annihilation lifetime spectroscopy (PALS) and polarized optical microscopy (POM). The mechanisms involving fitting the molecules into most symmetrical crystal lattices vary for species of different structure. Thermal energy can be used to expand the crystal lattice leading to thermochromism in the case of T
(CHdouble bond, length as m-dashCH
)
or conversely to an unusual negative thermal expansion of crystals of T
(CH
CH
CH
Cl)
that results in their self-actuation. The complex behaviour is reflected in unusual changes in the capacitance and fractional free volume of the material. These phenomena can be used for molecular design of advanced well-defined hybrid materials capable of reversible thermally induced structural transformations. The findings present a new perspective for POSS-based flexible metal-organic frameworks (MOF) of cooperative structural transformability via entropy-based translational sub-net sliding.
J-PET is a detector optimized for registration of photons from the electron–positron annihilation via plastic scintillators where photons interact predominantly via Compton scattering. Registration ...of both primary and scattered photons enables to determinate the linear polarization of the primary photon on the event by event basis with a certain probability. Here we present quantitative results on the feasibility of such polarization measurements of photons from the decay of positronium with the J-PET and explore the physical limitations for the resolution of the polarization determination of 511 keV photons via Compton scattering. For scattering angles of about 82
∘
(where the best contrast for polarization measurement is theoretically predicted) we find that the single event resolution for the determination of the polarization is about 40
∘
(predominantly due to properties of the Compton effect). However, for samples larger than ten thousand events the J-PET is capable of determining relative average polarization of these photons with the precision of about few degrees. The obtained results open new perspectives for studies of various physics phenomena such as quantum entanglement and tests of discrete symmetries in decays of positronium and extend the energy range of polarization measurements by five orders of magnitude beyond the optical wavelength regime.
Celotno besedilo
Dostopno za:
DOBA, IZUM, KILJ, NUK, PILJ, PNG, SAZU, SIK, UILJ, UKNU, UL, UM, UPUK
Samples of healthy and neoplastic myometrial tissues were investigated using Positron Annihilation Lifetime Spectroscopy (PALS). Meaningful differences between normal and diseased tissues were ...observed for each patient. The differences were also clearly visible for various kinds of lesions in each patient. The set of lifetimes and intensities obtained from PALS was correlated with the histopathological examinations of the same fragments of tissues. Strong coincidence between PALS parameters and histopathological findings was observed only in the case of a very precise correlation of the investigated area in both techniques. Measurements and discussion presented here were carried out to develop a method for measuring the sub-nanometric structure of human tissues. This kind of investigation, using positron probe, creates an opportunity of a new application in Positron Emission Tomography (PET).
In this paper we estimate the time resolution of the J-PET scanner built from plastic scintillators. We incorporate the method of signal processing using the Tikhonov regularization framework and the ...kernel density estimation method. We obtain simple, closed-form analytical formulae for time resolution. The proposed method is validated using signals registered by means of the single detection unit of the J-PET tomograph built from a 30 cm long plastic scintillator strip. It is shown that the experimental and theoretical results obtained for the J-PET scanner equipped with vacuum tube photomultipliers are consistent.
Purpose
In living organisms, the positron-electron annihilation (occurring during the PET imaging) proceeds in about 30% via creation of a metastable ortho-positronium atom. In the tissue, due to the ...pick-off and conversion processes, over 98% of ortho-positronia annihilate into two 511 keV photons. In this article, we assess the feasibility for reconstruction of the mean ortho-positronium lifetime image based on annihilations into two photons. The main objectives of this work include the (i) estimation of the sensitivity of the total-body PET scanners for the ortho-positronium mean lifetime imaging using 2
γ
annihilations and (ii) estimation of the spatial and time resolution of the ortho-positronium image as a function of the coincidence resolving time (CRT) of the scanner.
Methods
Simulations are conducted assuming that radiopharmaceutical is labeled with
44
S
c
isotope emitting one positron and one prompt gamma. The image is reconstructed on the basis of triple coincidence events. The ortho-positronium lifetime spectrum is determined for each voxel of the image. Calculations were performed for cases of total-body detectors build of (i) LYSO scintillators as used in the EXPLORER PET and (ii) plastic scintillators as anticipated for the cost-effective total-body J-PET scanner. To assess the spatial and time resolution, the four cases were considered assuming that CRT is equal to 500 ps, 140 ps, 50 ps, and 10 ps.
Results
The estimated total-body PET sensitivity for the registration and selection of image forming triple coincidences (2
γ
+
γ
prompt
) is larger by a factor of 13.5 (for LYSO PET) and by factor of 5.2 (for plastic PET) with respect to the sensitivity for the standard 2
γ
imaging by LYSO PET scanners with AFOV = 20 cm. The spatial resolution of the ortho-positronium image is comparable with the resolution achievable when using TOF-FBP algorithms already for CRT = 50 ps. For the 20-min scan, the resolution better than 20 ps is expected for the mean ortho-positronium lifetime image determination.
Conclusions
Ortho-positronium mean lifetime imaging based on the annihilations into two photons and prompt gamma is shown to be feasible with the advent of the high sensitivity total-body PET systems and time resolution of the order of tens of picoseconds.