    Lockhart, Catherine M; MacDonald, Lawrence R; Alessio, Adam M; McDougald, Wendy A; Doot, Robert K; Kinahan, Paul E

    Journal of Nuclear Medicine, 02/2011, Letnik: 52, Številka: 2
    Journal Article

    The purpose of this study was to measure the errors introduced by regular calibration of PET/CT scanners and to minimize the effect of calibration error on standardized uptake value measurements. Global calibration factors from 2 PET/CT scanners were recorded for 3.5 and 1.8 y, comparing manufacturer-recommended protocols with modified protocols to evaluate error contributions due to operator-influenced procedures. Dose calibrator measurements were evaluated using National Institute of Standards and Technology-traceable sources. Dose calibrator variability was less than 1%, although there was a consistent bias. Global scaling variability was reduced from 6% to 4% for scanner 1 and from 11% to 4% for scanner 2 when quality assurance and quality control procedures were applied to the calibration protocol. When calibrations were done using a (68)Ge/(68)Ga phantom, the variability for both scanners was reduced to approximately 3%. Applying quality assurance and quality control procedures to scanner calibration reduces variability, but there is a still a residual longitudinal scanner variability of 3%-4%. The procedures proposed here reduce the impact of operator error on scanner calibration and thereby minimize longitudinal variability in standardized uptake value measurements.