  • Comparison of the accuracy ...
    Ando, Wataru; Takao, Masaki; Hamada, Hidetoshi; Uemura, Keisuke; Sugano, Nobuhiko

    International orthopaedics, 07/2021, Letnik: 45, Številka: 7
    Journal Article

    Purpose We investigated the accuracy of the acetabular cup position and orientation in robotic-assisted total hip arthroplasty (rTHA) compared to navigated THA (nTHA) using computed tomography (CT) for patients with osteoarthritis secondary to developmental dysplasia of the hip (DDH). Methods We studied 31 hips of 28 patients who underwent rTHA and 119 hips of 112 patients who underwent nTHA with the same target cup orientation. After propensity score matching, each group comprised 29 hips. Post-operative cup position and orientation were measured from the postoperative CT data. Errors from the target cup position and orientation were compared between the two groups. Results Post-operatively, the absolute error of the anteroposterior and superoinferior cup positions from the target position was significantly lower in the rTHA group than in the nTHA group. The change in the post-operative radiographic inclination from the target orientation was lower in the rTHA group than in the nTHA group. Screw fixation for cup implantation was required for three hips in the nTHA group but not in the rTHA group. Conclusion rTHA achieved more precise cup implantation with reduced variation from the target orientation compared to nTHA in patients with osteoarthritis secondary to DDH.