    Shivdasani, Mohit N; Sinclair, Nicholas C; Dimitrov, Peter N; Varsamidis, Mary; Ayton, Lauren N; Luu, Chi D; Perera, Thushara; McDermott, Hugh J; Blamey, Peter J

    Investigative ophthalmology & visual science, 10/2014, Letnik: 55, Številka: 10
    Journal Article

    The suprachoroidal location for a retinal prosthesis provides advantages over other locations in terms of a simplified surgical procedure and a potentially more stable electrode-neural interface. The aim of this study was to assess the factors affecting perceptual thresholds, and to optimize stimulus parameters to achieve the lowest thresholds in patients implanted with a suprachoroidal retinal prosthesis. Three patients with profound vision loss from retinitis pigmentosa were implanted with a suprachoroidal array. Perceptual thresholds measured on individual electrodes were analyzed as a function of stimulus (return configuration, pulse polarity, pulse width, interphase gap, and rate), electrode (area and number of ganged electrodes), and clinical (retinal thickness and electrode-retina distance) parameters. A total of 92.8% of 904 measurements made up to 680 days post implantation yielded thresholds (range, 44-436 nanocoulombs nC) below the safe charge limit. Thresholds were found to vary between individuals and to depend significantly on electrode-retina distance, negligibly on retinal thickness, and not on electrode area or the number of ganged electrodes. Lowest thresholds were achieved when using a monopolar return, anodic-first polarity, short pulse widths (100 μs) combined with long interphase gaps (500 μs), and high stimulation rates (≥400 pulses per second pps). With suprachoroidal stimulation, anodic-first pulses with a monopolar return are most efficacious. To enable high rates, an appropriate combination of pulse width and interphase gap must be chosen to ensure low thresholds and electrode voltages. Electrode-retina distance needs to be monitored carefully owing to its influence on thresholds. These results inform implantable stimulator specifications for a suprachoroidal retinal prosthesis. (ClinicalTrials.gov number, NCT01603576.).