Akademska digitalna zbirka SLovenije - logo
E-viri
Celotno besedilo
Recenzirano
  • Dose Measurements in a Phan...
    Sáez-Vergara, J.C.; Romero, A.M.; Fernández, C.; Gómez, S.; Vázquez, J.; Olivares, M.P.

    Radiation protection dosimetry, 09/1999, Letnik: 85, Številka: 1-4
    Journal Article, Conference Proceeding

    A study reproducing usual exposure conditions in a special care baby unit has been performed to measure doses using TL materials in a versatile phantom specially designed for neonates having X ray examinations. The phantom offers the possibilities of reproducing different patient thicknesses and representing either a solid or hollow lung region. The results of the dose measurements using TL materials at the entrance, exit and both laterals of the phantom during different chest radiograph conditions are presented. Test conditions were reproduced in both hollow and solid chest cages simulating patient thicknesses of 5, 6 and 7 cm. The study was completed using two types of TL materials, LiF:Mg,Cu,P and LiF:Mg,Ti, in order to analyse and correct the differences on energy response between the two phosphors.