E-viri
Recenzirano
-
Fuss, M; Salter, B J; Caron, J-L; Vollmer, D G; Herman, T S
Acta neurochirurgica 147, Številka: 11Journal Article
Presentation of intensity-modulated radiosurgery (IMRS) for the treatment of inoperable, complex shaped pediatric arterio-venous malformations AVM. Between 03/99 and 11/04, IMRS was delivered to seven children aged six to 18 years. Prescribed minimum doses ranged from 17.5 to 20 Gy (median 18 Gy). Radiosurgery planning and delivery used a serial tomotherapeutic IMRT technique (Peacock IMRT, North American Scientific/Nomos, Cranberry Township, PA) over two to four couch angles. A linear accelerator attached binary multi-leaf collimator was used to generate pencil beams of 10 mm by either 8.5 or 4.0 mm. Treatment planning employed an inverse treatment planning optimization algorithm. Parameters submitted to the treatment planning system were: prescription dose (PD), volume of target allowed to receive less dose (standard 3%), minimum dose (0.5 Gy less than PD), and maximum dose (200% of PD). Planning system specific IMRS target and tissue types were selected to prioritize dose conformality over dose homogeneity. The prescription isodose encompassed at least 95% of the target volume. We calculated conformality (CI) and homogeneity indices (HI) to characterize the quality of IMRS plans, and summarized preliminary clinical outcomes. Target volumes ranged from 0.71 to 63.02 cm(3) (median 13.8 cm(3), 6/7 AVM larger than 10 cm(3)). Median CI was 1.07 (range 1.05 to 1.7) according to RTOG criteria. Median HI was 1.12 (range 1.09 to 1.23). During limited follow-up (median 32 months, range 5 to 53 months), two AVM completely obliterated at 19 and 22 months, and partial obliteration (>75%) was observed in three cases. No treatment-related side effects, other than acute nausea and temporary headaches interpreted as being associated with changes in cerebral blood distribution, were observed. IMRS can allow for highly conformal planning and delivery of radiosurgery radiation doses even if pediatric AVM target volumes are large and/or highly complex in shape. This technique has been seen to result in favorable preliminary outcomes, thus supporting future exploration of this technique in pediatric and adult patients.
![loading ... loading ...](themes/default/img/ajax-loading.gif)
Vnos na polico
Trajna povezava
- URL:
Faktor vpliva
Dostop do baze podatkov JCR je dovoljen samo uporabnikom iz Slovenije. Vaš trenutni IP-naslov ni na seznamu dovoljenih za dostop, zato je potrebna avtentikacija z ustreznim računom AAI.
Leto | Faktor vpliva | Izdaja | Kategorija | Razvrstitev | ||||
---|---|---|---|---|---|---|---|---|
JCR | SNIP | JCR | SNIP | JCR | SNIP | JCR | SNIP |
Baze podatkov, v katerih je revija indeksirana
Ime baze podatkov | Področje | Leto |
---|
Povezave do osebnih bibliografij avtorjev | Povezave do podatkov o raziskovalcih v sistemu SICRIS |
---|
Vir: Osebne bibliografije
in: SICRIS
To gradivo vam je dostopno v celotnem besedilu. Če kljub temu želite naročiti gradivo, kliknite gumb Nadaljuj.