    Mitsuda, C; Kajita, T; Miyano, K; Moriyama, S; Nakahata, M; Takeuchi, Y; Tasaka, S

    Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 02/2003, Letnik: 497, Številka: 2
    Journal Article

    A super-high sensitivity radon detector for water which works as a real-time monitor of the radon concentration in water for the Super-Kamiokande detector was developed. Super-high sensitivity was achieved through the utilization of the membrane degassifier module, in addition to factors such as large detector volume, high electrostatic collection voltage, and high-quality electronics. As a result of the calibration study, a sensitivity of 14.6±2.1(stat.+syst.) (counts/day)/(mBq/m 3) is achieved. The background level of the radon detector for water is 34.5±5.8(stat.) (counts/day). This corresponds to the detection limit of 0.7 mBq/m 3 in a one-day measurement.