    Molinelli, S; Mairani, A; Mirandola, A; Vilches Freixas, G; Tessonnier, T; Giordanengo, S; Parodi, K; Ciocca, M; Orecchia, R

    Physics in medicine & biology, 06/2013, Letnik: 58, Številka: 11
    Journal Article

    During one year of clinical activity at the Italian National Center for Oncological Hadron Therapy 31 patients were treated with actively scanned proton beams. Results of patient-specific quality assurance procedures are presented here which assess the accuracy of a three-dimensional dose verification technique with the simultaneous use of multiple small-volume ionization chambers. To investigate critical cases of major deviations between treatment planning system (TPS) calculated and measured data points, a Monte Carlo (MC) simulation tool was implemented for plan verification in water. Starting from MC results, the impact of dose calculation, dose delivery and measurement set-up uncertainties on plan verification results was analyzed. All resulting patient-specific quality checks were within the acceptance threshold, which was set at 5% for both mean deviation between measured and calculated doses and standard deviation. The mean deviation between TPS dose calculation and measurement was less than ±3% in 86% of the cases. When all three sources of uncertainty were accounted for, simulated data sets showed a high level of agreement, with mean and maximum absolute deviation lower than 2.5% and 5%, respectively.