Akademska digitalna zbirka SLovenije - logo
E-viri
Recenzirano Odprti dostop
  • Physical evaluation of an u...
    Oostveen, Luuk J.; Boedeker, Kirsten L.; Brink, Monique; Prokop, Mathias; de Lange, Frank; Sechopoulos, Ioannis

    European radiology, 05/2020, Letnik: 30, Številka: 5
    Journal Article

    Objectives To evaluate the technical performance of an ultra-high-resolution CT (UHRCT) system. Methods The physico-technical capabilities of a novel commercial UHRCT system were assessed and compared with those of a current-generation multi-detector (MDCT) system. The super-high-resolution (SHR) mode of the system uses 0.25 mm (at isocentre) detector elements (dels) in the in-plane and longitudinal directions, while the high-resolution (HR) mode bins two dels in the longitudinal direction. The normal-resolution (NR) mode bins dels 2 × 2, resulting in a del-size equivalent to that of the MDCT system. In general, standard procedures and phantoms were used to perform these assessments. Results The UHRCT MTF (10% MTF 4.1 lp/mm) is twice as high as that of the MDCT (10% MTF 1.9 lp/mm), which is comparable to the MTF in the NR mode (10% MTF 1.7 lp/mm). The width of the slice sensitivity profile in the SHR mode (FWHM 0.45 mm) is about 60% of that of the MDCT (FWHM 0.77 mm). Uniformity and CT numbers are within the expected range. Noise in the high-resolution modes has a higher magnitude and higher frequency components compared with MDCT. Low-contrast visibility is lower for the NR, HR and SHR modes compared with MDCT, but about a 14%, for NR, and 23%, for HR and SHR, dose increase gives the same results. Conclusions HR and SHR mode scanning results in double the spatial resolution, with about a 23% increase in dose required to achieve the same low-contrast detectability. Key Points • Resolution on UHRCT is up to twice as high as for the tested MDCT. • With abdominal settings, UHRCT needs higher dose for the same low-contrast detectability as MDCT, but dose is still below achievable levels as defined by current diagnostic reference levels. • The UHRCT system used in normal-resolution mode yields comparable resolution and noise characteristics as the MDCT system.