Akademska digitalna zbirka SLovenije - logo
E-viri
Recenzirano Odprti dostop
  • Intra-examiner and inter-ex...
    McCoy, Matthew; Campbell, Ismay; Stone, Pamela; Fedorchuk, Curtis; Wijayawardana, Sameera; Easley, Kirk

    PloS one, 02/2011, Letnik: 6, Številka: 2
    Journal Article

    The objective of this study was to evaluate the intra-examiner and inter-examiner reproducibility of paraspinal thermography using an infrared scanner. The thermal functions of a commercially available infrared scanner (Insight Subluxation Station®) were evaluated for clinical reliability. Two practicing clinicians conducted the measures on 100 subjects. Intra class correlation coefficients (ICCs) and concordance correlation coefficients (CCCs) were calculated from the collected data. Mean bilateral paraspinal skin temperature was 89.78° F and ranged from 88.77° F to 91.43° F. Intra class correlation coefficients (ICCs) for agreement and consistency ranged from 0.959 to 0.976. Concordance correlation coefficients (CCCs) ranged from 0.783 to 0.859 with tight confidence intervals indicating robust estimates of these quantities. This study revealed excellent intra-examiner and inter-examiner reproducibility of paraspinal thermography using a commercially available unit.