  • Noise studies of n-strip on...
    Robinson, D.; Allport, P.P.; Bizzell, J.; Buttar, C.; Carter, A.A.; Carter, J.R.; Goodrick, M.; Greenall, A.; Hill, J.C.; Morgan, D.; Munday, D.J.; Ohsugi, T.; Phillips, P.W.; Riedler, P.; Smith, N.A.; Terada, S.; Turner, P.R.; Unno, Y.

    Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 1999, Letnik: 426, Številka: 1
    Journal Article

    N-strip on n-bulk silicon microstrip detectors were irradiated at the CERN PS to 3×10 14 p cm -2 and their post-irradiation performance evaluated using fast binary readout electronics. Strip noise measurements demonstrate that detectors using conventional p-stop strip isolation are vulnerable to microdischarge at high bias voltages after irradiation. However, a novel isolation technique is shown to suppress microdischarge and lead to excellent post-irradiation characteristics.