Purpose:
Small field output correction factors have been studied by several research groups for the PTW 60019 microDiamond (MD) dosimeter, by comparing the response of such a device with both ...reference dosimeters and Monte Carlo simulations. A general good agreement is observed for field sizes down to about 1 cm. However, evident inconsistencies can be noticed when comparing some experimental results and Monte Carlo simulations obtained for smaller irradiation fields. This issue was tentatively attributed by some authors to unintentional large variations of the MD active surface area. The aim of the present study is a nondestructive experimental determination of the MD active surface area and active volume.
Methods:
Ten MD dosimeters, one MD prototype, and three synthetic diamond samples were investigated in the present work. 2D maps of the MD response were recorded under scanned soft x-ray microbeam irradiation, leading to an experimental determination of the device active surface area. Profiles of the device responses were measured as well. In order to evaluate the MD active volume, the thickness of the diamond sensing layer was independently evaluated by capacitance measurements and alpha particle detection experiments. The MD sensitivity, measured at the PTW calibration laboratory, was also used to calculate the device active volume thickness.
Results:
An average active surface area diameter of (2.19 ± 0.02) mm was evaluated by 2D maps and response profiles of all the MDs. Average active volume thicknesses of (1.01 ± 0.13) μm and (0.97 ± 0.14) μm were derived by capacitance and sensitivity measurements, respectively. The obtained results are well in agreement with the nominal values reported in the manufacturer dosimeter specifications. A homogeneous response was observed over the whole device active area. Besides the one from the device active volume, no contributions from other components of the housing nor from encapsulation materials were observed in the 2D response maps.
Conclusions:
The obtained results demonstrate the high reproducibility of the MD fabrication process. The observed discrepancies among the output correction factors reported by several authors for MD response in very small fields are very unlikely to be ascribed to unintentional variations of the device active surface area and volume. It is the opinion of the authors that the role of the volume averaging as well as of other perturbation effects should be separately investigated instead, both experimentally and by Monte Carlo simulations, in order to better clarify the behaviour of the MD response in very small fields.
Purpose:
To determine the potentialities of synthetic single crystal diamond Schottky diodes for accurate dose measurements in radiation therapy small photon beams.
Methods:
The dosimetric properties ...of a diamond-based detector were assessed by comparison with a reference microionization chamber. The diamond device was operated at zero bias voltage under irradiation with high-energy radiotherapic photon beams. The stability of the detector response and its dose and dose rate dependence were measured. Different square field sizes ranging from 1 × 1 cm2 to 10 × 10 cm2 were used during comparative dose distribution measurements by means of percentage depth dose curves (PDDs), lateral beam profiles, and output factors. The angular and temperature dependence of the diamond detector response were also studied.
Results:
The detector response shows a deviation from linearity of less than ±0.5% in the 0.01–7 Gy range and dose rate dependence below ±0.5% in the 1–6 Gy/min range. PDDs and output factors are in good agreement with those measured by the reference ionization chamber within 1%. No angular dependence is observed by rotating the detector along its axis, while ∼3.5% maximum difference is measured by varying the radiation incidence angle in the polar direction. The temperature dependence was investigated as well and a ±0.2% variation of the detector response is found in the 18–40 °C range.
Conclusions:
The obtained results indicate the investigated synthetic diamond-based detector as a candidate for small field clinical radiation dosimetry in advanced radiation therapy techniques.
Purpose:
To investigate the dosimetric properties of a synthetic single crystal diamond Schottky diode for accurate relative dose measurements in large and small field high-energy clinical proton ...beams.
Methods:
The dosimetric properties of a synthetic single crystal diamond detector were assessed by comparison with a reference Markus parallel plate ionization chamber, an Exradin A16 microionization chamber, and Exradin T1a ion chamber. The diamond detector was operated at zero bias voltage at all times. Comparative dose distribution measurements were performed by means of Fractional depth dose curves and lateral beam profiles in clinical proton beams of energies 155 and 250 MeV for a 14 cm square cerrobend aperture and 126 MeV for 3, 2, and 1 cm diameter circular brass collimators. ICRU Report No. 78 recommended beam parameters were used to compare fractional depth dose curves and beam profiles obtained using the diamond detector and the reference ionization chamber. Warm-up/stability of the detector response and linearity with dose were evaluated in a 250 MeV proton beam and dose rate dependence was evaluated in a 126 MeV proton beam. Stem effect and the azimuthal angle dependence of the diode response were also evaluated.
Results:
A maximum deviation in diamond detector signal from the average reading of less than 0.5% was found during the warm-up irradiation procedure. The detector response showed a good linear behavior as a function of dose with observed deviations below 0.5% over a dose range from 50 to 500 cGy. The detector response was dose rate independent, with deviations below 0.5% in the investigated dose rates ranging from 85 to 300 cGy/min. Stem effect and azimuthal angle dependence of the diode signal were within 0.5%. Fractional depth dose curves and lateral beam profiles obtained with the diamond detector were in good agreement with those measured using reference dosimeters.
Conclusions:
The observed dosimetric properties of the synthetic single crystal diamond detector indicate that its behavior is proton energy independent and dose rate independent in the investigated energy and dose rate range and it is suitable for accurate relative dosimetric measurements in large as well as in small field high energy clinical proton beams.
Purpose
The purpose of this study was to investigate for the first time the performance of a synthetic single crystal diamond detector for the microdosimetric characterization of clinical 62 MeV ...ocular therapy proton beams.
Methods
A novel diamond microdosimeter with a well‐defined sensitive volume was fabricated and tested with a monoenergetic and spread‐out Bragg peak (SOBP) of the CATANA therapeutic proton beam in Catania, Italy. The whole sensitive volume of the detector has an active planar‐sectional area of 100 µm × 100 µm and a thickness of approximately 6.3 um. Microdosimetric measurements were performed at several water equivalent depths, corresponding to positions of clinical relevance. From the measured spectra, microdosimetric quantities such as the frequency mean lineal energy (y¯F), dose mean lineal energy (y¯D) as well as microdosimetric relative biological effectiveness (RBEµ) values were derived for each depth along both a pristine Bragg curve and SOBP. Finally, Geant4 Monte Carlo simulations were performed modeling the detector geometry and CATANA beamline in order to calculate the average linear energy transfer (LET) values in the diamond active layer and water.
Results
The microdosimetric spectra acquired by the diamond microdosimeter show different shapes as a function of the water equivalent depths. No spectral distortion, due to pile‐up events and polarization effects, was observed. The experimental spectra have a very low detection threshold due to the electronic noise during the irradiation of about 1 keV/μm. The y¯F and y¯D values were in agreement with expected trends, showing a sharp increase in mean lineal energy at the distal edge of the Bragg peak. In addition, a good agreement between the mean lineal energy values and the calculated average LET ones was also observed. Finally, the RBE values evaluated with the diamond microdosimeter were in excellent agreement with those obtained with a mini tissue equivalent proportional counter as well as with radiobiological measurements in the same proton beam field.
Conclusions
The microdosimetric performance of the tested synthetic single crystal diamond microdosimeter clearly indicates its suitability for quality assurance in clinical proton therapy beam.
Ion stopping in warm dense matter is a process of fundamental importance for the understanding of the properties of dense plasmas, the realization and the interpretation of experiments involving ...ion-beam-heated warm dense matter samples, and for inertial confinement fusion research. The theoretical description of the ion stopping power in warm dense matter is difficult notably due to electron coupling and degeneracy, and measurements are still largely missing. In particular, the low-velocity stopping range, that features the largest modelling uncertainties, remains virtually unexplored. Here, we report proton energy-loss measurements in warm dense plasma at unprecedented low projectile velocities. Our energy-loss data, combined with a precise target characterization based on plasma-emission measurements using two independent spectroscopy diagnostics, demonstrate a significant deviation of the stopping power from classical models in this regime. In particular, we show that our results are in closest agreement with recent first-principles simulations based on time-dependent density functional theory.
Purpose:
To investigate the dosimetric properties of synthetic single crystal diamond based Schottky diodes under irradiation with therapeutic electron beams from linear accelerators.
Methods:
A ...single crystal diamond detector was fabricated and tested under 6, 8, 10, 12, and 15 MeV electron beams. The detector performances were evaluated using three types of commercial detectors as reference dosimeters: an Advanced Markus plane parallel ionization chamber, a Semiflex cylindrical ionization chamber, and ap-type silicon detector. Preirradiation, linearity with dose, dose rate dependence, output factors, lateral field profiles, and percentage depth dose profiles were investigated and discussed.
Results:
During preirradiation the diamond detector signal shows a weak decrease within 0.7% with respect to the plateau value and a final signal stability of 0.1% (1σ) is observed after about 5 Gy. A good linear behavior of the detector response as a function of the delivered dose is observed with deviations below ±0.3% in the dose range from 0.02 to 10 Gy. In addition, the detector response is dose rate independent, with deviations below 0.3% in the investigated dose rate range from 0.17 to 5.45 Gy/min. Percentage depth dose curves obtained from the diamond detector are in good agreement with the ones from the reference dosimeters. Lateral beam profile measurements show an overall good agreement among detectors, taking into account their respective geometrical features. The spatial resolution of solid state detectors is confirmed to be better than that of ionization chambers, being the one from the diamond detector comparable to that of the silicon diode. A good agreement within experimental uncertainties was also found in terms of output factor measurements between the diamond detector and reference dosimeters.
Conclusions:
The observed dosimetric properties indicate that the tested diamond detector is a suitable candidate for clinical electron beam dosimetry.
Purpose:
To investigate for the first time the dosimetric properties of a new commercial synthetic diamond detector (PTW microDiamond) in high‐energy scanned clinical carbon ion beams generated by a ...synchrotron at the CNAO facility.
Methods:
The detector response was evaluated in a water phantom with actively scanned carbon ion beams ranging from 115 to 380 MeV/u (30–250 mm Bragg peak depth in water). Homogeneous square fields of 3 × 3 and 6 × 6 cm2 were used. Short‐ and medium‐term (2 months) detector response stability, dependence on beam energy as well as ion type (carbon ions and protons), linearity with dose, and directional and dose‐rate dependence were investigated. The depth dose curve of a 280 MeV/u carbon ion beam, scanned over a 3 × 3 cm2 area, was measured with the microDiamond detector and compared to that measured using a PTW Advanced Markus ionization chamber, and also simulated using fluka Monte Carlo code. The detector response in two spread‐out‐Bragg‐peaks (SOBPs), respectively, centered at 9 and 21 cm depths in water and calculated using the treatment planning system (TPS) used at CNAO, was measured.
Results:
A negligible drift of detector sensitivity within the experimental session was seen, indicating that no detector preirradiation was needed. Short‐term response reproducibility around 1% (1 standard deviation) was found. Only 2% maximum variation of microDiamond sensitivity was observed among all the evaluated proton and carbon ion beam energies. The detector response showed a good linear behavior. Detector sensitivity was found to be dose‐rate independent, with a variation below 1.3% in the evaluated dose‐rate range. A very good agreement between measured and simulated Bragg curves with both microDiamond and Advanced Markus chamber was found, showing a negligible LET dependence of the tested detector. A depth dose curve was also measured by positioning the microDiamond with its main axis oriented orthogonally to the beam direction. A strong distortion in Bragg peak measurement was observed, confirming manufacturer recommendation on avoiding such configuration. Very good results were obtained for SOBP measurements, with a difference below 1% between measured and TPS‐calculated doses. The stability of detector sensitivity in the observation period was within the experimental uncertainty.
Conclusions:
Dosimetric characterization of a PTW microDiamond detector in high‐energy scanned carbon ion beams was performed. The results of the present study showed that this detector is suitable for dosimetry of clinical carbon ion beams, with a negligible LET and dose‐rate dependence.
Abstract
In proton therapy, most treatment planning systems (TPS) use a fixed relative biological effectiveness (RBE) of 1.1 all along the depth-dose profile. Innovative TPS are now investigated ...considering the variability of RBE with radiation quality. New TPS need an experimental verification in the quality assurance (QA) routine in clinics, but RBE data are usually obtained with radiobiological measurements that are time consuming and not suitable for daily QA. Microdosimetry is a useful tool based on physical measurements which can monitor the radiation quality. Several microdosimeters are available in different research institutions, which could potentially be used for the QA in TPS. In this study, the response functions of five detectors in the same 62-MeV proton Spread Out Bragg Peak is compared in terms of spectral distributions and their average values and microdosimetric RBE. Their different response function has been commented and must be considered in the clinical practice.