The use of radiochromic film (RCF) dosimetry in radiation therapy is extensive due to its high level of achievable accuracy for a wide range of dose values and its suitability under a variety of ...measurement conditions. However, since the publication of the 1998 AAPM Task Group 55, Report No. 63 on RCF dosimetry, the chemistry, composition, and readout systems for RCFs have evolved steadily. There are several challenges in using the new RCFs, readout systems and validation of the results depending on their applications. Accurate RCF dosimetry requires understanding of RCF selection, handling and calibration methods, calibration curves, dose conversion methods, correction methodologies as well as selection, operation and quality assurance (QA) programs of the readout systems. Acquiring this level of knowledge is not straight forward, even for some experienced users. This Task Group report addresses these issues and provides a basic understanding of available RCF models, dosimetric characteristics and properties, advantages and limitations, configurations, and overall elemental compositions of the RCFs that have changed over the past 20 yr. In addition, this report provides specific guidelines for data processing and analysis schemes and correction methodologies for clinical applications in radiation therapy.
Highlights • Review of technical aspects for radiochromic film dosimetry system. • Recipe on how to design a reference radiochromic film dosimetry system. • Summary of parameters that define ...radiochromic film dosimetry protocol.
Purpose
To introduce a model that reproducibly linearizes the response from radiochromic film (RCF) dosimetry systems at extended dose range. To introduce a correction method, generated from the same ...scanned images, which corrects for scanner temporal response variation and scanner bed inhomogeneity.
Methods
Six calibration curves were established for different lot numbers of EBT3 GAFCHROMIC™ film model based on four EPSON scanners 10000XL (2 units), 11000XL, 12000XL at three different centers. These films were calibrated in terms of absorbed dose to water based on TG51 protocol or TRS398 with dose ranges up to 40 Gy. The film response was defined in terms of a proposed normalized pixel value (nPVRGB) as a summation of first‐order equations based on information from red, green, and blue channels. The fitting parameters of these equations are chosen in a way that makes the film response equal to dose at the time of calibration. An integrated set of correction factors (one per color channel) was also introduced. These factors account for the spatial and temporal changes in scanning states during calibration and measurements. The combination of nPVRGB and this “fingerprint” correction formed the basis of this new protocol and it was tested against net optical density (netODX=R,G,B) single‐channel dosimetry in terms of accuracy, precision, scanner response variability, scanner bed inhomogeneity, noise, and long‐term stability.
Results
Incorporating multichannel features (RGB) into the normalized pixel value produced linear response to absorbed dose (slope of 1) in all six RCF dosimetry systems considered in this study. The “fingerprint” correction factors of each of these six systems displayed unique patterns at the time of calibration. The application of nPVRGB to all of these six systems could achieve a level of accuracy of ± 2.0% in the dose range of interest within modeled uncertainty level of 2.0%–3.0% depending on the dose level. Consistent positioning of control and measurement film pieces and integrating the multichannel correction into the response function formalism mitigated possible scanner response variations of as much as ± 10% at lower doses and scanner bed inhomogeneity of ± 8% to the established level of uncertainty at the time of calibration. The system was also able to maintain the same level of accuracy after 3 and 6 months post calibration.
Conclusions
Combining response linearity with the integrated correction for scanner response variation lead to a sustainable and practical RCF dosimetry system that mitigated systematic response shifts and it has the potential to reduce errors in reporting relative information from the film response.
Purpose:
Radiochromic film provides dose measurement at high spatial resolution, but often is not preferred for routine evaluation of patient-specific intensity modulated radiation therapy (IMRT) ...plans owing to ease-of-use factors. The authors have established an efficient protocol that combines calibration and measurement in a single scan and enables measurement results to be obtained in less than 30 min. This avoids complications due to postexposure changes in radiochromic film that delay the completion of a measurement, often for up to 24 h, in commonly used methods. In addition, the protocol addresses the accuracy and integrity of the measurement by eliminating environmental and interscan variability issues.
Methods:
The authors collected dose–response data from six production lots of Gafchromic EBT3 film and three production lots of EBT2 film at doses up to 480 cGy. In this work, the authors used seven different scanners of two different models—Epson 10000XL and V700; postexposure times before scanning from 30 min to 9 days; ambient temperatures for scanning spanning 11 °C; and two film orientations. Scanning was in 48-bit RGB format at 72 dpi resolution. Dose evaluation was conducted using a triple-channel dosimetry method. To evaluate the measurement protocol, patient specific IMRT and volumetric modulated arc therapy (VMAT) plans were exposed onto EBT3 films on a Varian Trilogy Linac. Film scanning was done following the protocol under a number of different conditions and the dose maps were analyzed to demonstrate the equivalence of results.
Results:
The results indicated that the dose–response data could be fit by a set of related rational functions leading to the description of a generic calibration curve. A simplified dosimetry protocol was established where dose–response data for a specific film lot, scanner, and scanning conditions could be derived from two films exposed to known doses. In most cases only one calibrated exposure was required since the dose for one of the films could be zero. Using the Gamma test criterion of 2%/2 mm to evaluate the measurements, similar passing rates ranging between about 95% and 99% for the fields studied were obtained from application films digitized under a variety of conditions all of them different than the conditions under which the calibration films were scanned.
Conclusions:
The authors have developed a simplified and efficient protocol to measure doses delivered by an IMRT or VMAT plan using only the patient film, one calibration film, one unexposed film, and applying a single scan to acquire a digital image for calculation and analysis. The simplification and timesaving offer a potential practical solution for using radiochromic film for routine treatment plan quality assurance without sacrificing spatial resolution for convenience.
In this study, a measurement protocol is presented that improves the precision of dose measurements using a flat-bed document scanner in conjunction with two new GafChromic® film models, HS and ...Prototype A EBT exposed to
6
MV
photon beams. We established two sources of uncertainties in dose measurements, governed by measurement and calibration curve fit parameters contributions. We have quantitatively assessed the influence of different steps in the protocol on the overall dose measurement uncertainty. Applying the protocol described in this paper on the Agfa Arcus II flat-bed document scanner, the overall one-sigma dose measurement uncertainty for an uniform field amounts to 2% or less for doses above around
0.4
Gy
in the case of the EBT (Prototype A), and for doses above
5
Gy
in the case of the HS model GafChromic® film using a region of interest
2
×
2
mm
2
in size.
We investigated the dose-response of the external beam therapy 3 (EBT3) films for proton and carbon ion clinical beams, in comparison with conventional radiotherapy beams; we also measured the film ...response along the energy deposition-curve in water. We performed measurements at three hadrontherapy centres by delivering monoenergetic pencil beams (protons: 63-230 MeV; carbon ions: 115-400 MeV/u), at 0.4-20 Gy dose to water, in the plateau of the depth-dose curve. We also irradiated the films to clinical MV-photon and electron beams. We placed the EBT3 films in water along the whole depth-dose curve for 148.8 MeV protons and 398.9 MeV/u carbon ions, in comparison with measurements provided by a plane-parallel ionization chamber. For protons, the response of EBT3 in the plateau of the depth-dose curve is not different from that of photons, within experimental uncertainties. For carbon ions, we observed an energy dependent under-response of EBT3 film, from 16% to 29% with respect to photon beams. Moreover, we observed an under-response in the Bragg peak region of about 10% for 148.8 MeV protons and of about 42% for 398.9 MeV/u carbon ions. For proton and carbon ion clinical beams, an under-response occurs at the Bragg peak. For carbon ions, we also observed an under-response of the EBT3 in the plateau of the depth-dose curve. This effect is the highest at the lowest initial energy of the clinical beams, a phenomenon related to the corresponding higher LET in the film sensitive layer. This behavior should be properly modeled when using EBT3 films for accurate 3D dosimetry.
Film is an excellent dosimeter for verification of dose distributions due to its high spatial resolution. Irradiated film can be digitized with low-cost, transmission, flatbed scanners. However, a ...disadvantage is their lateral scan effect (LSE): a scanner readout change over its lateral scan axis. Although anisotropic light scattering was presented as the origin of the LSE, this paper presents an alternative cause. Hereto, LSE for two flatbed scanners (Epson 1680 Expression Pro and Epson 10000XL), and Gafchromic film (EBT, EBT2, EBT3) was investigated, focused on three effects: cross talk, optical path length and polarization. Cross talk was examined using triangular sheets of various optical densities. The optical path length effect was studied using absorptive and reflective neutral density filters with well-defined optical characteristics (OD range 0.2-2.0). Linear polarizer sheets were used to investigate light polarization on the CCD signal in absence and presence of (un)irradiated Gafchromic film. Film dose values ranged between 0.2 to 9 Gy, i.e. an optical density range between 0.25 to 1.1. Measurements were performed in the scanner's transmission mode, with red-green-blue channels. LSE was found to depend on scanner construction and film type. Its magnitude depends on dose: for 9 Gy increasing up to 14% at maximum lateral position. Cross talk was only significant in high contrast regions, up to 2% for very small fields. The optical path length effect introduced by film on the scanner causes 3% for pixels in the extreme lateral position. Light polarization due to film and the scanner's optical mirror system is the main contributor, different in magnitude for the red, green and blue channel. We concluded that any Gafchromic EBT type film scanned with a flatbed scanner will face these optical effects. Accurate dosimetry requires correction of LSE, therefore, determination of the LSE per color channel and dose delivered to the film.
Radiochromic film has become an important tool to verify dose distributions in highly conformal radiation therapy such as IMRT. Recently, a new generation of these films, EBT3, has become available. ...EBT3 has the same composition and thickness of the sensitive layer of the previous EBT2 films, but its symmetric layer configuration allows the user to eliminate side orientation dependence, which is reported for EBT2 films. The most important EBT3 characteristics have been investigated, such as response at high‐dose levels, sensitivity to scanner orientation and postirradiation coloration, energy and dose rate dependence, and orientation dependence with respect to film side. Additionally, different IMRT fields were measured with both EBT3 and EBT2 films and evaluated using gamma index analysis. The results obtained show that most of the characteristics of EBT3 film are similar to the EBT2 film, but the orientation dependence with respect to film side is completely eliminated in EBT3 films. The study confirms that EBT3 film can be used for clinical practice in the same way as the previous EBT2 film.
PACS number: 87.56.Fc
Purpose:
In radiochromic film dosimetry systems, measurements are usually obtained from film images acquired on a CCD‐based flatbed scanner. The authors investigated factors affecting scan‐to‐scan ...response variability leading to increased dose measurement uncertainty.
Methods:
The authors used flatbed document scanners to repetitively scan EBT3 radiochromic films exposed to doses 0–1000 cGy, together with three neutral density filters and three blue optical filters. Scanning was performed under two conditions: scanner lid closed and scanner lid opened/closed between scans. The authors also placed a scanner in a cold room at 9 °C and later in a room at 22 °C and scanned EBT3 films to explore temperature effects. Finally, the authors investigated the effect of altering the distance between the film and the scanner's light source.
Results:
Using a measurement protocol to isolate the contribution of the CCD and electronic circuitry of the scanners, the authors found that the standard deviation of response measurements for the EBT3 film model was about 0.17% for one scanner and 0.09% for the second. When the lid of the first scanner was opened and closed between scans, the average scan‐to‐scan difference of responses increased from 0.12% to 0.27%. Increasing the sample temperature during scanning changed the RGB response values by about −0.17, −0.14, and −0.05%/°C, respectively. Reducing the film‐to‐light source distance increased the RBG response values about 1.1, 1.3, and 1.4%/mm, respectively. The authors observed that films and film samples were often not flat with some areas up to 8 mm away from the scanner's glass window.
Conclusions:
In the absence of measures to deal with the response irregularities, each factor the authors investigated could lead to dose uncertainty >2%. Those factors related to the film‐to‐light source distance could be particularly impactful since the authors observed many instances where the curl of film samples had the potential to cause dose uncertainty in excess of 5%. Two expedients will eliminate the uncertainties: a transparent sheet (preferably glass) placed over the scanned film keeps the film‐to‐light source distance constant, and an EBT3 reference film included in all scans provides correction factors for measured response values.
Purpose
The purpose of this study was to investigate the spectral response of the EBT3 radiochromic films to different beam qualities for radiation therapy dosimetry. Dose, dose rate, and interbatch ...dependencies on the spectral response of the films were investigated as well.
Methods
Pieces of EBT3 films placed between layers of solid water phantoms were irradiated with 6 and 15 MV photon beams, 6 and 10 MV‐flattening filter free (FFF) photon beams, and 6 and 20 MeV electron beams at dose levels between 0.4 and 50 Gy. Net absorbance was measured as a function of wavelength from the spectra acquired in the wavelength range of 400–800 nm using a fiber‐coupled spectrometer and broadband light source.
Results
No significant change was observed in the absorption spectra of the EBT3 film from the same batch irradiated with the same amount of dose using different beam qualities. Also, no spectral change with dose rate was observed. The measured net absorbance per Gy was independent of beam quality in the 1–50 Gy dose range. Slight differences in the spectral shape and absorption band positions were observed in film samples from different batches. The net absorbance spectra showed two absorption bands centered around 634–636 nm (primary) and 583–585 nm (secondary). However, depending on the film batch, for doses above a certain level the primary absorption band appears to “split” into two bands centered around ~624–628 and ~641–645 nm.
Conclusions
The spectral shape of the EBT3 radiochromic films irradiated with photons (including FFF) and electron beams is beam quality and dose rate independent; however, it varies with dose level, batch, and spectroscopy system used.