    Mandić, I.; Cindro, V.; Gorišek, A.; Hiti, B.; Kramberger, G.; Zavrtanik, M.; Mikuž, M.; Hemperek, T.

    Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 09/2018, Volume: 903
    Journal Article

    The edge transient-current technique (Edge-TCT) and charge-collection measurements with passive test structures made with the LFoundry 150-nm CMOS process on a p-type substrate with an initial resistivity of over 3 kΩcm are presented. The measurements were made before and after irradiation with reactor neutrons up to 2⋅1015 neq/cm2. Two sets of devices were investigated: unthinned (700 μm) with the substrate biased through the implant on top and thinned (200 μm) with a processed and metallised backplane. The depletion depth was estimated with the Edge-TCT and the collected charge was measured with a 90Sr source using an external amplifier having a 25-ns shaping time. The depletion depth for a given bias voltage decreased with the increasing neutron fluence, but it was still larger than 70 μm at 250 V after the highest fluence. After irradiation a much higher collected charge was measured for the thinned detectors with a processed backplane compared to the unthinned devices, although the same or an even larger depletion depth was measured in the unthinned devices. The most probable value of the collected charge of over 5000 electrons was measured with a thinned device also after irradiation to 2⋅1015 neq/cm2. This is sufficient to ensure the successful operation of these detectors at the outer layer of the pixel detector in the ATLAS experiment at the upgraded HL-LHC.