The aim of the SYRMA-CT collaboration is to set-up the first clinical trial of phase-contrast breast CT with synchrotron radiation (SR). In order to combine high image quality and low delivered dose ...a number of innovative elements are merged: a CdTe single photon counting detector, state-of-the-art CT reconstruction and phase retrieval algorithms. To facilitate an accurate exam optimization, a Monte Carlo model was developed for dose calculation using GEANT4. In this study, high isotropic spatial resolution (120 μm)(3) CT scans of objects with dimensions and attenuation similar to a human breast were acquired, delivering mean glandular doses in the range of those delivered in clinical breast CT (5-25 mGy). Due to the spatial coherence of the SR beam and the long distance between sample and detector, the images contain, not only absorption, but also phase information from the samples. The application of a phase-retrieval procedure increases the contrast-to-noise ratio of the tomographic images, while the contrast remains almost constant. After applying the simultaneous algebraic reconstruction technique to low-dose phase-retrieved data sets (about 5 mGy) with a reduced number of projections, the spatial resolution was found to be equal to filtered back projection utilizing a four fold higher dose, while the contrast-to-noise ratio was reduced by 30%. These first results indicate the feasibility of clinical breast CT with SR.
To introduce the optimization of a customized GPU-based simultaneous algebraic reconstruction technique (cSART) in the field of phase-contrast breast computed tomography (bCT). The presented ...algorithm features a 3D bilateral regularization filter that can be tuned to yield optimal performance for clinical image visualization and tissues segmentation.
Acquisitions of a dedicated test object and a breast specimen were performed at Elettra, the Italian synchrotron radiation (SR) facility (Trieste, Italy) using a large area CdTe single-photon counting detector. Tomographic images were obtained at 5 mGy of mean glandular dose, with a 32 keV monochromatic x-ray beam in the free-space propagation mode. Three independent algorithms parameters were optimized by using contrast-to-noise ratio (CNR), spatial resolution, and noise texture metrics. The results obtained with the cSART algorithm were compared with conventional SART and filtered back projection (FBP) reconstructions. Image segmentation was performed both with gray scale-based and supervised machine-learning approaches.
Compared to conventional FBP reconstructions, results indicate that the proposed algorithm can yield images with a higher CNR (by 35% or more), retaining a high spatial resolution while preserving their textural properties. Alternatively, at the cost of an increased image 'patchiness', the cSART can be tuned to achieve a high-quality tissue segmentation, suggesting the possibility of performing an accurate glandularity estimation potentially of use in the realization of realistic 3D breast models starting from low radiation dose images.
The study indicates that dedicated iterative reconstruction techniques could provide significant advantages in phase-contrast bCT imaging. The proposed algorithm offers great flexibility in terms of image reconstruction optimization, either toward diagnostic evaluation or image segmentation.
The aim of this study was to optimise the experimental protocol and data analysis for in-vivo breast cancer x-ray imaging. Results are presented of the experiment at the SYRMEP beamline of Elettra ...Synchrotron using the propagation-based phase-contrast mammographic tomography method, which incorporates not only absorption, but also x-ray phase information. In this study the images of breast tissue samples, of a size corresponding to a full human breast, with radiologically acceptable x-ray doses were obtained, and the degree of improvement of the image quality (from the diagnostic point of view) achievable using propagation-based phase-contrast image acquisition protocols with proper incorporation of x-ray phase retrieval into the reconstruction pipeline was investigated. Parameters such as the x-ray energy, sample-to-detector distance and data processing methods were tested, evaluated and optimized with respect to the estimated diagnostic value using a mastectomy sample with a malignant lesion. The results of quantitative evaluation of images were obtained by means of radiological assessment carried out by 13 experienced specialists. A comparative analysis was performed between the x-ray and the histological images of the specimen. The results of the analysis indicate that, within the investigated range of parameters, both the objective image quality characteristics and the subjective radiological scores of propagation-based phase-contrast images of breast tissues monotonically increase with the strength of phase contrast which in turn is directly proportional to the product of the radiation wavelength and the sample-to-detector distance. The outcomes of this study serve to define the practical imaging conditions and the CT reconstruction procedures appropriate for low-dose phase-contrast mammographic imaging of live patients at specially designed synchrotron beamlines.
•Phase-contrast CT addresses poor soft-tissue contrast in X-ray breast imaging.•Interactive 3D breast virtual visualizations available at different resolutions.•Spatial resolution bridging from ...radiological level to virtual histology level.•2D and 3D images may be combinedly used in a multidisciplinary clinical context.•Discernibility of breast structures and microcalcifications is enhanced.
Phase-contrast imaging techniques address the issue of poor soft-tissue contrast encountered in traditional X-ray imaging. This can be accomplished with the propagation-based phase-contrast technique by employing a coherent photon beam, which is available at synchrotron facilities, as well as long sample-to-detector distances. This study demonstrates the optimization of propagation-based phase-contrast computed tomography (CT) techniques for multiscale X-ray imaging of the breast at the Elettra synchrotron facility (Trieste, Italy).
Two whole breast mastectomy samples were acquired with propagation-based breast-CT using a monochromatic synchrotron beam at a pixel size of 60 µm. Paraffin-embedded blocks sampled from the same tissues were scanned with propagation-based micro-CT imaging using a polychromatic synchrotron beam at a pixel size of 4 µm. Images of both methodologies and of the same sample were spatially registered.
The resulting images showed the transition from whole breast imaging with propagation-based breast-CT methodology to virtual histology with propagation-based micro-CT imaging of the same sample. Additionally, conventional histological images were matched to virtual histology images.
Phase-contrast images offer a high resolution with low noise, which allows for a highly precise match between virtual and conventional histology. Furthermore, those techniques allow a clear discernment of breast structures, lesions, and microcalcifications, being a promising clinically-compatible tool for breast imaging in a multiscale approach, to either assist in the detection of cancer in full volume breast samples or to complement structure identification in paraffin-embedded breast tissue samples.
Conventional contrast agents utilized in diagnostic radiology are based on x-ray absorption properties; alternative physical principles capable of providing a contrast enhancement in radiographs have ...never been applied. This study exploits the possibility of using a novel type of contrast media based on x-ray scattering. The contrast agents consist of microbubble echo-enhancing agents, usually applied in ultrasound examinations, which are invisible with conventional x-ray absorption techniques. The experiment was carried out at the medical beamline of the synchrotron radiation laboratory ELETTRA in Trieste, Italy. A flat silicon analyzer crystal typically used for diffraction-enhanced imaging was utilized as a tool for detecting the scattering properties of the contrast agents. In analyzer-based imaging, it is possible to detect the scattering properties of the sample by shifting the analyzer crystal to selected positions of its reflectivity curve. In particular, when the sample consists of a large number of micro-particles an overall effect can be observed. Phantoms containing contrast agents based on microbubbles were imaged at different angular positions of the analyzer crystal. High visibility of the details was demonstrated, and a strong contrast enhancement was measured compared to normal x-ray absorption techniques.