Purpose
The purpose of this study was to explore gains in predictive model performance for radiation pneumonitis (RP) using pretreatment CT radiomics features extracted from the normal lung volume.
...Methods
A total of 192 patients treated for nonsmall cell lung cancer with definitive radiotherapy were considered in the current study. In addition to clinical and dosimetric data, CT radiomics features were extracted from the total lung volume defined using the treatment planning scan. A total of 6851 features (15 clinical, 298 total lung and heart dosimetric, and 6538 image features) were gathered and considered candidate predictors for modeling of RP grade ≥3. Models were built with the least absolute shrinkage and selection operator (LASSO) logistic regression and applied to the set of candidate predictors with 50 iterations of tenfold nested cross‐validation.
Results
In the current cohort, 30 of 192 patients (15.6%) presented with RP grade ≥3. Average cross‐validated AUC (CV‐AUC) using only the clinical and dosimetric parameters was 0.51. CV‐AUC was 0.68 when total lung CT radiomics features were added. Analysis with the entire set of available predictors revealed seven different image features selected in at least 40% of the model fits.
Conclusions
We have successfully incorporated CT radiomics features into a framework for building predictive RP models via LASSO logistic regression. Addition of normal lung image features produced superior model performance relative to traditional dosimetric and clinical predictors of RP, suggesting that pretreatment CT radiomics features should be considered in the context of RP prediction.
Lung Size and the Risk of Radiation Pneumonitis Briere, Tina Marie; Krafft, Shane; Liao, Zhongxing ...
International journal of radiation oncology, biology, physics,
02/2016, Letnik:
94, Številka:
2
Journal Article
Recenzirano
The purpose of this study was to identify patient populations treated for non-small cell lung cancer (NSCLC) who may be more at risk of radiation pneumonitis.
A total of 579 patients receiving ...fractionated 3D conformal or intensity modulated radiation therapy (IMRT) for NSCLC were included in the study. Statistical analysis was performed to search for cohorts of patients with higher incidences of radiation pneumonitis. In addition to conventional risk factors, total and spared lung volumes were analyzed. The Lyman-Kutcher-Burman (LKB) and cure models were then used to fit the incidence of radiation pneumonitis as a function of lung dose and other factors.
Total lung volumes with a sparing of less than 1854 cc at 40 Gy were associated with a significantly higher incidence of radiation pneumonitis at 6 months (38% vs 12% for patients with larger volumes, P<.001). This patient cohort was overwhelmingly female and represented 22% of the total female population of patients and nearly 30% of the cases of radiation pneumonitis. An LKB fit to normal tissue complication probability (NTCP) including volume as a dose modifying factor resulted in a dose that results in a 50% probability of complication for the smaller spared volume cohort that was 9 Gy lower than the fit to all mean lung dose data and improved the ability to predict radiation pneumonitis (P<.001). Using an effective dose parameter of n=0.42 instead of mean lung dose further improved the LKB fit. Fits to the data using the cure model produced similar results.
Spared lung volume should be considered when treating NSCLC patients. Separate dose constraints based on smaller spared lung volume should be considered. Smaller spared lung volume patients should be followed closely for signs of radiation pneumonitis.
Charge coupled devices (CCDs) are being increasingly used in radiation therapy for dosimetric purposes. However, CCDs are sensitive to stray radiation. This effect induces transient noise. ...Radiation-induced noise strongly alters the image and therefore limits its quantitative analysis. The purpose of this work is to characterize the radiation-induced noise and to develop filtration algorithms to restore image quality. Two models of CCD were used for measurements close to a medical linac. The structure of the transient noise was first characterized. Then, four methods of noise filtration were compared: median filtering of a time series of identical images, uniform median filtering of single images, an adaptive filter with switching mechanism, and a modified version of the adaptive switch filter. The intensity distribution of noisy pixels was similar in both cameras. However, the spatial distribution of the noise was different: The average noise cluster size was
1.2
±
0.6
and
3.2
±
2.7
pixels
for the U2000 and the Luca, respectively. The median of a time series of images resulted in the best filtration and minimal image distortion. For applications where time series is impractical, the adaptive switch filter must be used to reduce image distortion. Our modified version of the switch filter can be used in order to handle nonisolated groups of noisy pixels.
Patients with glioblastoma multiforme (GBM) require radiotherapy as part of definitive management. Our institution has adopted the use of volumetric arc therapy (VMAT) due to superior sparing of the ...adjacent organs at risk (OARs) compared to intensity modulated radiation therapy (IMRT). Here we report our clinical experience by analyzing target coverage and sparing of OARs for 90 clinical treatment plans.
VMAT and IMRT patient cohorts comprising 45 patients each were included in this study. For all patients, the planning target volume (PTV) received 50 Gy in 30 fractions, and the simultaneous integrated boost PTV received 60 Gy. The characteristics of the two patient cohorts were examined for similarity. The doses to target volumes and OARs, including brain, brainstem, hippocampi, optic nerves, eyes, and cochleae were then compared using statistical analysis. Target coverage and normal tissue sparing for six patients with both clinical IMRT and VMAT plans were analyzed.
PTV coverage of at least 95% was achieved for all plans, and the median mean dose to the boost PTV differed by only 0.1 Gy between the IMRT and VMAT plans. Superior sparing of the brainstem was found with VMAT, with a median difference in mean dose being 9.4 Gy. The ipsilateral cochlear mean dose was lower by 19.7 Gy, and the contralateral cochlea was lower by 9.5 Gy. The total treatment time was reduced by 5 min. The difference in the ipsilateral hippocampal D
was 12 Gy, though this is not statistically significant (P = 0.03).
VMAT for GBM patients can provide similar target coverage, superior sparing of the brainstem and cochleae, and be delivered in a shorter period of time compared with IMRT. The shorter treatment time may improve clinical efficiency and the quality of the treatment experience. Based on institutional clinical experience, use of VMAT for the treatment of GBMs appears to offer no inferiority in comparison to IMRT and may offer distinct advantages, especially for patients who may require re-irradiation.
We investigated the correlation between the motions of an external marker and internal fiducials implanted in the liver for 8 patients undergoing respiratory-based computed tomography ...(four-dimensional CT 4D-CT) procedures.
The internal fiducials were gold seeds, 3 mm in length and 1.2 mm in diameter. Four patients each had one implanted fiducial, and the other four had three implanted fiducials. The external marker was a plastic box, which is part of the Real-Time Position Management System (RPM) used to track the patient's respiration. Each patient received a standard helical CT scan followed by a time-correlated CT-image acquisition (4D-CT). The 4D-CT images were reconstructed in 10 separate phases covering the entire respiratory cycle.
The internal fiducial motion is predominant in the superior-inferior direction, with a range of 7.5-17.5 mm. The correlation between external respiration and internal fiducial motion is best during expiration. For 2 patients with their three fiducials separated by a maximum of 3.2 cm, the motions of the fiducials were well correlated, whereas for 2 patients with more widely spaced fiducials, there was less correlation.
In general, there is a good correlation between internal fiducial motion imaged by 4D-CT and external marker motion. We have demonstrated that gating may be best performed at the end of the respiratory cycle. Special attention should be paid to gating for patients whose fiducials do not move in synchrony, because targeting on the correct respiratory amplitude alone would not guarantee that the entire tumor volume is within the treatment field.
Complex radiation therapy techniques require dosimetric verification of treatment planning and delivery. The authors investigated a liquid scintillator (LS) system for application for real-time ...high-energy photon beam dosimetry. The system was comprised of a transparent acrylic tank filled with liquid scintillating material, an opaque outer tank, and a CCD camera. A series of images was acquired when the tank with liquid scintillator was irradiated with a 6 MV photon beam, and the light data measured with the CCD camera were filtered to correct for scattering of the optical light inside the liquid scintillator. Depth-dose and lateral profiles as well as two-dimensional (2D) dose distributions were found to agree with results from the treatment planning system. Further, the corrected light output was found to be linear with dose, dose rate independent, and is robust for single or multiple acquisitions. The short time needed for image acquisition and processing could make this system ideal for fast verification of the beam characteristics of the treatment machine. This new detector system shows a potential usefulness of the LS for 2D QA.
The goal of this work was to demonstrate the feasibility of using a plastic scintillation detector (PSD) incorporated into a prostate immobilization device to verify doses in vivo delivered during ...intensity-modulated radiation therapy (IMRT) and volumetric modulated-arc therapy (VMAT) for prostate cancer. The treatment plans for both modalities had been developed for a patient undergoing prostate radiation therapy. First, a study was performed to test the dependence, if any, of PSD accuracy on the number and type of calibration conditions. This study included PSD measurements of each treatment plan being delivered under quality assurance (QA) conditions using a rigid QA phantom. PSD results obtained under these conditions were compared to ionization chamber measurements. After an optimal set of calibration factors had been found, the PSD was combined with a commercial endorectal balloon used for rectal distension and prostate immobilization during external beam radiotherapy. This PSD-enhanced endorectal balloon was placed inside of a deformable anthropomorphic phantom designed to simulate male pelvic anatomy. PSD results obtained under these so-called “simulated treatment conditions” were compared to doses calculated by the treatment planning system (TPS). With the PSD still inserted in the pelvic phantom, each plan was delivered once again after applying a shift of 1 cm anterior to the original isocenter to simulate a treatment setup error.
The mean total accumulated dose measured using the PSD differed the TPS-calculated doses by less than 1% for both treatment modalities simulated treatment conditions using the pelvic phantom. When the isocenter was shifted, the PSD results differed from the TPS calculations of mean dose by 1.2% (for IMRT) and 10.1% (for VMAT); in both cases, the doses were within the dose range calculated over the detector volume for these regions of steep dose gradient. Our results suggest that the system could benefit prostate cancer patient treatment by providing accurate in vivo dose reports during treatment and verify in real-time whether treatments are being delivered according to the prescribed plan.
► Plastic scintillation detector (PSD) tested using high dose gradient radiotherapy beams. ► Measurements were made in real-time during beam delivery using rigid and deformable phantoms. ► PSD-measured doses agreed well with ionization chamber results and treatment planning system. ► Findings suggest PSDs are viable candidates for in vivo real-time dosimetry for radiotherapy.
The Pediatric Normal Tissue Effects in the Clinic (PENTEC) pulmonary task force reviewed dosimetric and clinical factors associated with radiation therapy (RT)-associated pulmonary toxicity in ...children.
Comprehensive search of PubMed (1965-2020) was conducted to assess available evidence and predictive models of RT-induced lung injury in pediatric cancer patients (<21 years old). Lung dose for radiation pneumonitis (RP) was obtained from dose-volume histogram (DVH) data. RP grade was obtained from standard criteria. Clinical pulmonary outcomes were evaluated using pulmonary function tests (PFTs), clinical assessment, and questionnaires.
More than 2,400 abstracts were identified; 460 articles had detailed treatment and toxicity data; and 11 articles with both detailed DVH and toxicity data were formally reviewed. Pooled cohorts treated during 1999 to 2016 included 277 and 507 patients age 0.04 to 22.7 years who were evaluable for acute and late RP analysis, respectively. After partial lung RT, there were 0.4% acute and 2.8% late grade 2, 0.4% acute and 0.8% late grade 3, and no grade 4 to 5 RP. RP risk after partial thoracic RT with mean lung dose (MLD) <14 Gy and total lung V
<30% is low. Clinical and self-reported pulmonary outcomes data included 8,628 patients treated during 1970 to 2013, age 0 to 21.9 years. At a median 2.9- to 21.9-year follow-up, patients were often asymptomatic; abnormal PFTs were common and severity correlated with lung dose. At ≥10-year follow-up, multi-institutional studies suggested associations between total or ipsilateral lung doses >10 Gy and pulmonary complications and deaths. After whole lung irradiation (WLI), pulmonary toxicity is higher; no dose response relationship was identified. Bleomycin and other chemotherapeutics at current dose regimens do not contribute substantially to adverse pulmonary outcomes after partial lung irradiation but increase risk with WLI.
After partial lung RT, acute pulmonary toxicity is uncommon; grade 2 to 3 RP incidences are <1%. Late toxicities, including subclinical/asymptomatic impaired pulmonary function, are more common (<4%). Incidence and severity appear to increase over time. Upon review of available literature, there appears to be low risk of pulmonary complications in children with MLD < 14 Gy and V
<30% using standard fractionated RT to partial lung volumes. A lack of robust data limit guidance on lung dose/volume constraints, highlighting the need for additional work to define factors associated with RT-induced lung injury.
Purpose
We developed and tested an automatic field‐in‐field (FIF) solution for whole‐brain radiotherapy (WBRT) planning that creates a homogeneous dose distribution by minimizing hotspots, resulting ...in clinically acceptable plans.
Methods
A configurable auto‐planning algorithm was developed to automatically generate FIF WBRT plans independent of the treatment planning system. Configurable parameters include the definition of hotspots, target volume, maximum number of subfields, and minimum number of monitor units per field. This algorithm iteratively identifies a hotspot, creates two opposing subfields, calculates the dose, and optimizes the beam weight based on user‐configured constraints of dose‐volume histogram coverage and least‐squared cost functions. The algorithm was retrospectively tested on 17 whole‐brain patients. First, an in‐house landmark‐based automated beam aperture technique was used to generate the treatment fields and initial plans. Second, the FIF algorithm was employed to optimize the plans using physician‐defined goals of 99.9% of the brain volume receiving 100% of the prescription dose (30 Gy in 10 fractions) and a target hotspot definition of 107% of the prescription dose. The final auto‐optimized plans were assessed for clinical acceptability by an experienced radiation oncologist using a five‐point scale.
Results
The FIF algorithm reduced the mean (± SD) plan hotspot percentage dose from 35.0 Gy (116.6%) ± 0.6 Gy (2.0%) to 32.6 Gy (108.8%) ± 0.4 Gy (1.2%). Also, it decreased the mean (± SD) hotspot V107% cm3 from 959 ± 498 cm3 to 145 ± 224 cm3. On average, plans were produced in 16 min without any user intervention. Furthermore, 76.5% of the auto‐plans were clinically acceptable (needing no or minor stylistic edits), and all of them were clinically acceptable after minor clinically necessary edits.
Conclusions
This algorithm successfully produced high‐quality WBRT plans and can improve treatment planning efficiency when incorporated into an automatic planning workflow.
Abstract Background and purpose Desmoplastic small round cell tumor (DSRCT) is a rare, aggressive malignancy. We report survival rates and toxicity associated with sequential multimodality treatment ...including whole abdominopelvic radiation therapy (WART). Material and methods Medical records of 32 patients with DSRCT treated at our institution were reviewed. Patients underwent chemotherapy, cytoreductive surgery with hyperthermic intraperitoneal chemoperfusion (HIPEC), followed by WART with intensity-modulated radiation or volumetric-modulated arc therapy. Results Median overall survival (OS) was 60 months. After 18 months of follow-up, 20 patients (62.5%) had disease recurrence and median disease-free survival (DFS) was 10 months. Median time to extrahepatic abdominal failure was 19.4 months. Factors affecting time to local progression included liver metastases at diagnosis, and an interval of greater than 5.6 months between diagnosis and HIPEC or greater than 2.1 months between HIPEC and WART. None of these factors altered OS. Grade 3 or higher toxicities occurred in 84% of patients. Conclusions WART following chemotherapy, surgical cytoreduction and HIPEC is an aggressive treatment for DSRCT patients and can result in severe side effects. Our median OS of 5 years is favorable compared to prior studies, despite a median DFS of only 10 months, which may be due to improved salvage therapies.