Several experiments in fundamental physics demand an environment of very low, homogeneous, and stable magnetic fields. For the magnetic characterization of such environments, we present a portable ...SQUID system that measures the absolute magnetic flux density vector and the gradient tensor. This vector-tensor system contains 13 integrated low-critical temperature (LTc) superconducting quantum interference devices (SQUIDs) inside a small cylindrical liquid helium Dewar with a height of 31 cm and 37 cm in diameter. The achievable resolution depends on the flux density of the field under investigation and its temporal drift. Inside a seven-layer mu-metal shield, an accuracy better than ±23 pT for the components of the static magnetic field vector and ±2 pT/cm for each of the nine components of the gradient tensor is reached by using the shifting method.
► Non-invasive detection of 1kHz SEP components. ► Low-noise amplifier custom-made for high-frequency recordings. ► Time–frequency analysis and spatial characterization of somatosensory evoked ...high-frequency components.
Scalp-derived human somatosensory evoked potentials (SEPs) contain high-frequency oscillations (600Hz; ‘sigma-burst’) reflecting concomitant bursts of spike responses in primary somatosensory cortex that repeat regularly at 600Hz. Notably, recent human intracranial SEP have revealed also 1kHz responses (‘kappa-burst’), possibly reflecting non-rhythmic spiking summed over multiple cells (MUA: multi-unit activity). However, the non-invasive detection of EEG signals at 1kHz typical for spikes has always been limited by noise contributions from both, amplifier and body/electrode interface. Accordingly, we developed a low-noise recording set-up optimised to map non-invasively 1kHz SEP components.
SEP were recorded upon 4Hz left median nerve stimulation in 6 healthy human subjects. Scalp potentials were acquired inside an electrically and magnetically shielded room using low-noise custom-made amplifiers. Furthermore, in order to reduce thermal Johnson noise contributions from the sensor/skin interface, electrode impedances were adjusted to ⩽1kΩ. Responses averaged after repeated presentation of the stimulus (n=4000 trials) were evaluated by spatio-temporal pattern analyses in complementary spectral bands.
Three distinct spectral components were identified: N20 (<100Hz), sigma-burst (450–750Hz), and kappa-burst (850–1200Hz). The two high-frequency bursts (sigma, kappa) exhibited distinct and partially independent spatiotemporal evolutions, indicating subcortical as well as several cortical generators.
Using a dedicated low-noise set-up, human SEP ‘kappa-bursts’ at 1kHz can be non-invasively detected and their scalp distribution be mapped. Their topographies indicate a set of subcortical/cortical generators, at least partially distinct from the topography of the 600Hz sigma-bursts described previously.
The non-invasive detection and surface mapping of 1kHz EEG signals presented here provides an essential step towards non-invasive monitoring of multi-unit spike activity.
Ultrafast electroencephalographic signals, having frequencies above 500 Hz, can be observed in somatosensory evoked potential measurements. Usually, these recordings have a poor signal-to-noise ratio ...(SNR) because weak signals are overlaid by intrinsic noise of much higher amplitude like that generated by biological sources and the amplifier. As an example, recordings at the scalp taken during electrical stimulation of the median nerve show a 600 Hz burst with submicro-volt amplitudes which can be extracted from noise by the use of massive averaging and digital signal processing only. We have investigated this signal by means of a very low noise amplifier made in-house (minimal voltage noise 2.7 nV Hz(-1/2), FET inputs). We examined how the SNR of the data is altered by the bandwidth and the use of amplifiers with different intrinsic amplifier noise levels of 12 and 4.8 nV Hz(-1/2), respectively. By analyzing different frequency contributions of the signal, we found an extremely weak 1 kHz component superimposed onto the well-known 600 Hz burst. Previously such high-frequency electroencephalogram responses around 1 kHz have only been observed by deep brain electrodes implanted for tremor therapy of Parkinson patients. For the non-invasive measurement of such signals, we recommend that amplifier noise should not exceed 4 nV Hz(-1/2).
Permanent electric dipole moments (EDMs) arise due to the breaking of time-reversal or, equivalently, CP-symmetry. Although EDM searches have so far only set upper limits, which are many orders of ...magnitude larger than Standard Model (SM) predictions, the motivation for more sensitive searches is stronger than ever. A new effort at FRM-II incorporating
129
Xe and
3
He as a co-magnetometer can potentially improve the current limit. The noble gas mixture of
129
Xe and
3
He is simultanously polarized by spin-exchange optical pumping and then transferred into a high-performance magnetically shielded room. Inside, both species can freely precess in the presence of applied magnetic and electric fields. The precession signals are detected by LTc SQUID sensors. In EDM cells with silicon electrodes we observed spin lifetimes in excess of 2500 s without and with high-voltage applied. This meets one requirement to achieve our goal of improving the EDM limit on
129
Xe by several orders of magnitude.
We present the result of an experiment to measure the electric dipole moment (EDM) of the neutron at the Paul Scherrer Institute using Ramsey's method of separated oscillating magnetic fields with ...ultracold neutrons. Our measurement stands in the long history of EDM experiments probing physics violating time-reversal invariance. The salient features of this experiment were the use of a ^{199}Hg comagnetometer and an array of optically pumped cesium vapor magnetometers to cancel and correct for magnetic-field changes. The statistical analysis was performed on blinded datasets by two separate groups, while the estimation of systematic effects profited from an unprecedented knowledge of the magnetic field. The measured value of the neutron EDM is d_{n}=(0.0±1.1_{stat}±0.2_{sys})×10^{-26} e.cm.
The authors suggest a procedure for the determination of the natural nuclear magnetic resonance line width Δ
ν of liquids using an air coil system at flux densities from 25
μT to 150
μT. Even if the ...line broadening caused by instrumental field inhomogeneity is much higher than Δ
ν, Δ
ν can be found by extrapolating the measured line width’s field dependency. For pure water this procedure yielded Δ
ν
=
0.125
Hz
±
0.005
Hz. This is shown to be consistent with the smallest line width found for this sample below 25
μT using a superconducting quantum interference device-based spectrometer.