The validity of continuous glucose monitoring (CGM) is well established in diabetic patients. CGM is also increasingly used for research purposes in normo-glycemic individuals, but the CGM validity ...in such individuals is unknown. We studied the accuracy of CGM measurements in normo-glycemic individuals by comparing CGM-derived versus venous blood-derived glucose levels and measures of glycemia and glycemic variability.
In 34 healthy participants (mean age 65.7 years), glucose was simultaneously measured every 10 minutes, via both an Enlite® CGM sensor, and in venous blood sampled over a 24-hour period. Validity of CGM-derived individual glucose measurements, calculated measures of glycemia over daytime (09:00h-23:00h) and nighttime (23:00h-09:00h), and calculated measures of glycemic variability (e.g. 24h standard deviation SD) were assessed by Pearson correlation coefficients, mean absolute relative difference (MARD) and paired t-tests.
The median correlation coefficient between CGM and venous glucose measurements per participant was 0.68 (interquartile range: 0.40-0.78), and the MARD was 17.6% (SD = 17%). Compared with venous sampling, the calculated measure of glycemia during daytime was 0.22 mmol/L higher when derived from CGM, but no difference was observed during nighttime. Most measures of glycemic variability were lower with CGM than with venous blood sampling (e.g., 24h SD: 1.07 with CGM and 1.26 with venous blood; p-value = 0.004).
In normo-glycemic individuals, CGM-derived glucose measurements had good agreement with venous glucose levels. However, the measure of glycemia was higher during the day and most measures of glycemic variability were lower when derived from CGM.
Guidelines of management of dyslipidemias and prevention of cardiovascular disease (CVD) are based on firm scientific evidence obtained by randomized controlled trials (RCTs). However, the role of ...elevated low-density lipoprotein-cholesterol (LDL-C)as a risk factor of CVD and therapies to lower LDL-C are frequently disputed by colleagues who disagree with the conclusions of the RCTs published. This review focuses on this dispute, and evaluates the current approach of management of dyslipidemias and CVD prevention to find modern alternatives for more precise diagnosis and therapy of dyslipidemic patients.
Recent interest in lipoprotein(a) (Lp(a)) and remnants lipoproteins and in therapies that do not influence LDL-C levels primarily, such as anti-inflammatory drugs and icosapent ethyl, has revitalized our concern to optimize the care for patients with increased CVD risk without focusing simply on reduction of LDL-C by therapy with statins, ezitemibe, and proprotein convertase subtilisin-kexin type 9 (PCSK9) inhibitors.
The limited characterization of study populations by measurement of total cholesterol (TC), high-density lipoprotein-cholesterol (HDL-C) and triglycerides (TG) followed by measurement or calculation of LDL-C should be extended by a more integral approach in order to realize precision diagnostics and precision medicine, for the sake of personalized patient care.
Elevated serum prostate-specific antigen (PSA) levels in body fluids may indicate prostate cancer (PCa), but it is noted that the clinical performance is rather poor. Specificity and sensitivity ...values of 20 and 94% at a cutoff value of 4.1 ng/mL, respectively, result in overdiagnosis and unnecessary interventions. Previous exploratory studies have indicated that the glycosylation of PSA potentially leads to improved PCa diagnosis based on qualitative analyses. However, the applied methods are not suited for a quantitative evaluation or implementation in a medical laboratory. Therefore, in this proof-of-principle study, we have evaluated the use of hydrophilic interaction liquid chromatography (HILIC) in combination with targeted quantitative mass spectrometry for the sialic acid linkage-specific analysis of PSA glyco-proteoforms based on either trypsin or ArgC peptides. The efficiency of PSA proteolysis was optimized as well as the glycopeptide separation conditions (buffer type, strength, and pH). The HILIC-based analysis of PSA glyco-proteoforms presented here has the potential for the clinical validation of patient cohorts. The method shows the feasibility of the use of a HILIC stationary phase for the separation of isomeric glycopeptides to detect specific glyco-proteoforms. This is the first step toward the development and evaluation of PSA glyco-proteoforms for use in a clinical chemistry setting aiming for improved PCa diagnosis or screening.
The aim of this study was to determine reference intervals in an outpatient population from Vall d'Hebron laboratory using an indirect approach previously described in a Dutch population (NUMBER ...project). We used anonymized test results from individuals visiting general practitioners and analysed during 2018. Analytical quality was assured by EQA performance, daily average monitoring and by assessing longitudinal accuracy between 2018 and 2020 (using trueness verifiers from Dutch EQA). Per test, outliers by biochemically related tests were excluded, data were transformed to a normal distribution (if necessary) and means and standard deviations were calculated, stratified by age and sex. In addition, the reference limit estimator method was also used to calculate reference intervals using the same dataset. Finally, for standardized tests reference intervals obtained were compared with the published NUMBER results. Reference intervals were calculated using data from 509,408 clinical requests. For biochemical tests following a normal distribution, similar reference intervals were found between Vall d'Hebron and the Dutch study. For creatinine and urea, reference intervals increased with age in both populations. The upper limits of Gamma-glutamyl transferase were markedly higher in the Dutch study compared to Vall d'Hebron results. Creatine kinase and uric acid reference intervals were higher in both populations compared to conventional reference intervals. Medical test results following a normal distribution showed comparable and consistent reference intervals between studies. Therefore a simple indirect method is a feasible and cost-efficient approach for calculating reference intervals. Yet, for generating standardized calculated reference intervals that are traceable to higher order materials and methods, efforts should also focus on test standardization and bias assessment using commutable trueness verifiers.
Abstract
Current dyslipidemia management in patients with atherosclerotic cardiovascular disease (ASCVD) is based on traditional serum lipids. Yet, there is some indication from basic research that ...serum apolipoproteins A-I, (a), B, C-I, C-II, C-III, and E may give better pathophysiological insight into the root causes of dyslipidemia. To facilitate the future adoption of clinical serum apolipoprotein (apo) profiling for precision medicine, strategies for accurate testing should be developed in advance.
Recent discoveries in basic science and translational medicine set the stage for the IFCC Working Group on Apolipoproteins by Mass Spectrometry. Main drivers were the convergence of unmet clinical needs in cardiovascular disease (CVD) patients with enabling technology and metrology. First, the residual cardiovascular risk after accounting for established risk factors demonstrates that the current lipid panel is too limited to capture the full complexity of lipid metabolism in patients. Second, there is a need for accurate test results in highly polymorphic and atherogenic apolipoproteins such as apo(a). Third, sufficient robustness of mass spectrometry technology allows reproducible protein quantification at the molecular level. Fourth, several calibration hierarchies in the revised ISO 17511:2020 guideline facilitate metrological traceability of test results, the highest achievable standard being traceability to SI.
This article outlines the conceptual approach aimed at achieving a novel, multiplexed Reference Measurement System (RMS) for seven apolipoproteins based on isotope dilution mass spectrometry and peptide-based calibration. This RMS should enable standardization of existing and emerging apolipoprotein assays to SI, within allowable limits of measurement uncertainty, through a sustainable network of Reference Laboratories.
The European Atherosclerosis Society-European Federation of Clinical Chemistry and Laboratory Medicine Consensus Panel aims to provide recommendations to optimize atherogenic lipoprotein ...quantification for cardiovascular risk management.
We critically examined LDL cholesterol, non-HDL cholesterol, apolipoprotein B (apoB), and LDL particle number assays based on key criteria for medical application of biomarkers. (
) Analytical performance: Discordant LDL cholesterol quantification occurs when LDL cholesterol is measured or calculated with different assays, especially in patients with hypertriglyceridemia >175 mg/dL (2 mmol/L) and low LDL cholesterol concentrations <70 mg/dL (1.8 mmol/L). Increased lipoprotein(a) should be excluded in patients not achieving LDL cholesterol goals with treatment. Non-HDL cholesterol includes the atherogenic risk component of remnant cholesterol and can be calculated in a standard nonfasting lipid panel without additional expense. ApoB more accurately reflects LDL particle number. (
) Clinical performance: LDL cholesterol, non-HDL cholesterol, and apoB are comparable predictors of cardiovascular events in prospective population studies and clinical trials; however, discordance analysis of the markers improves risk prediction by adding remnant cholesterol (included in non-HDL cholesterol) and LDL particle number (with apoB) risk components to LDL cholesterol testing. (
) Clinical and cost-effectiveness: There is no consistent evidence yet that non-HDL cholesterol-, apoB-, or LDL particle-targeted treatment reduces the number of cardiovascular events and healthcare-related costs than treatment targeted to LDL cholesterol.
Follow-up of pre- and on-treatment (measured or calculated) LDL cholesterol concentration in a patient should ideally be performed with the same documented test method. Non-HDL cholesterol (or apoB) should be the secondary treatment target in patients with mild to moderate hypertriglyceridemia, in whom LDL cholesterol measurement or calculation is less accurate and often less predictive of cardiovascular risk. Laboratories should report non-HDL cholesterol in all standard lipid panels.
An elevated low-density lipoprotein cholesterol concentration is a classical risk factor for cardiovascular disease. This has led to pharmacotherapy in patients with atherosclerotic heart disease or ...high heart disease risk with statins to reduce serum low-density lipoprotein cholesterol. Even in patients in whom the target levels of low-density lipoprotein cholesterol are reached, there remains a significant residual cardiovascular risk; this is due, in part, to a focus on low-density lipoprotein cholesterol alone and neglect of other important aspects of lipoprotein metabolism. A more refined lipoprotein analysis will provide additional information on the accumulation of very low-density lipoproteins, intermediate density lipoproteins, chylomicrons, chylomicron-remnants and Lp(a) concentrations. Instead of measuring the cholesterol and triglyceride content of the lipoproteins, measurement of their apolipoproteins (apos) is more informative. Apos are either specific for a particular lipoprotein or for a group of lipoproteins. In particular measurement of apos in atherogenic particles is more biologically meaningful than the measurement of the cholesterol concentration contained in these particles. Applying apo profiling will not only improve characterization of the lipoprotein abnormality, but will also improve definition of therapeutic targets. Apo profiling aligns with the concept of precision medicine by which an individual patient is not treated as ‘average’ patient by the average (dose of) therapy. This concept of precision medicine fits the unmet clinical need for stratified cardiovascular medicine. The requirements for clinical application of proteomics, including apo profiling, can now be met using robust mass spectrometry technology which offers desirable analytical performance and standardization.
In the past decade a remarkable rebirth of serum/plasma lipoprotein(a) (Lp(a)) as an independent risk factor of cardiovascular disease (CVD) occurred. Updated evidence for a causal continuous ...association in different ethnic groups between Lp(a) concentrations and cardiovascular outcomes has been published in the latest European Atherosclerosis Society (EAS) Lp(a) consensus statement. Interest in measuring Lp(a) at least once in a person's lifetime moreover originates from the development of promising new Lp(a) lowering drugs. Accurate and clinically effective Lp(a) tests are of key importance for the timely detection of high-risk individuals and for future evaluation of the therapeutic effects of Lp(a) lowering medication. To this end, it is necessary to improve the performance and standardization of existing Lp(a) tests, as is also noted in the Lp(a) consensus statement. Consequently, a state-of-the-art internationally endorsed reference measurement system (RMS) must be in place that allows for performance evaluation of Lp(a) field tests in order to certify their validity and accuracy. An ELISA-based RMS from Northwest Lipid Research Laboratory (University of Washington, Seattle, USA) has been available since the 1990s. A next-generation apo(a)/Lp(a) RMS is now being developed by a working group from the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC). The envisioned apo(a) RMS is based on the direct measurement of selected proteotypic fragments generated after proteolytic digestion using quantitative protein mass spectrometry (MS). The choice for an MS-based RMS enables selective measurement of the proteotypic peptides and is by design apo(a) isoform insensitive. Clearly, the equimolar conversion of apo(a) into the surrogate peptide measurands is required to obtain accurate Lp(a) results. The completeness of proteolysis under reaction conditions from the candidate reference measurement procedure (RMP) has been demonstrated for the quantifying apo(a) peptides. Currently, the candidate apo(a) RMP is endorsed by the IFCC and recommendations for suitable secondary reference materials have been made in a recent commutability study paper. Ongoing efforts toward a complete apo(a) RMS that is listed by the Joint Committee on Traceability in Laboratory Medicine (JCTLM) are focused on the peptide-based calibration and the establishment of a network of calibration laboratories running the apo(a) RMS in a harmonized way. Once completed, it will be the holy grail for evaluation and certification of Lp(a) field methods.