Noninvasive Glucose Sensing In Vivo Leung, Ho Man Colman; Forlenza, Gregory P; Prioleau, Temiloluwa O ...
Sensors,
08/2023, Letnik:
23, Številka:
16
Journal Article
Recenzirano
Odprti dostop
Blood glucose monitoring is an essential aspect of disease management for individuals with diabetes. Unfortunately, traditional methods require collecting a blood sample and thus are invasive and ...inconvenient. Recent developments in minimally invasive continuous glucose monitors have provided a more convenient alternative for people with diabetes to track their glucose levels 24/7. Despite this progress, many challenges remain to establish a noninvasive monitoring technique that works accurately and reliably in the wild. This review encompasses the current advancements in noninvasive glucose sensing technology in vivo, delves into the common challenges faced by these systems, and offers an insightful outlook on existing and future solutions.
This study evaluated a new insulin delivery system designed to reduce insulin delivery when trends in continuous glucose monitoring (CGM) glucose concentrations predict future hypoglycemia.
...Individuals with type 1 diabetes (
= 103, age 6-72 years, mean HbA
7.3% 56 mmol/mol) participated in a 6-week randomized crossover trial to evaluate the efficacy and safety of a Tandem Diabetes Care t:slim X2 pump with Basal-IQ integrated with a Dexcom G5 sensor and a predictive low-glucose suspend algorithm (PLGS) compared with sensor-augmented pump (SAP) therapy. The primary outcome was CGM-measured time <70 mg/dL.
Both study periods were completed by 99% of participants; median CGM usage exceeded 90% in both arms. Median time <70 mg/dL was reduced from 3.6% at baseline to 2.6% during the 3-week period in the PLGS arm compared with 3.2% in the SAP arm (difference PLGS - SAP = -0.8%, 95% CI -1.1 to -0.5,
< 0.001). The corresponding mean values were 4.4%, 3.1%, and 4.5%, respectively, represent-ing a 31% reduction in the time <70 mg/dL with PLGS. There was no increase in mean glucose concentration (159 vs. 159 mg/dL,
= 0.40) or percentage of time spent >180 mg/dL (32% vs. 33%,
= 0.12). One severe hypoglycemic event occurred in the SAP arm and none in the PLGS arm. Mean pump suspension time was 104 min/day.
The Tandem Diabetes Care Basal-IQ PLGS system significantly reduced hypoglycemia without rebound hyperglycemia, indicating that the system can benefit adults and youth with type 1 diabetes in improving glycemic control.
Closed-loop control systems of insulin delivery may improve glycemic outcomes in young children with type 1 diabetes. The efficacy and safety of initiating a closed-loop system virtually are unclear.
...In this 13-week, multicenter trial, we randomly assigned, in a 2:1 ratio, children who were at least 2 years of age but younger than 6 years of age who had type 1 diabetes to receive treatment with a closed-loop system of insulin delivery or standard care that included either an insulin pump or multiple daily injections of insulin plus a continuous glucose monitor. The primary outcome was the percentage of time that the glucose level was in the target range of 70 to 180 mg per deciliter, as measured by continuous glucose monitoring. Secondary outcomes included the percentage of time that the glucose level was above 250 mg per deciliter or below 70 mg per deciliter, the mean glucose level, the glycated hemoglobin level, and safety outcomes.
A total of 102 children underwent randomization (68 to the closed-loop group and 34 to the standard-care group); the glycated hemoglobin levels at baseline ranged from 5.2 to 11.5%. Initiation of the closed-loop system was virtual in 55 patients (81%). The mean (±SD) percentage of time that the glucose level was within the target range increased from 56.7±18.0% at baseline to 69.3±11.1% during the 13-week follow-up period in the closed-loop group and from 54.9±14.7% to 55.9±12.6% in the standard-care group (mean adjusted difference, 12.4 percentage points equivalent to approximately 3 hours per day; 95% confidence interval, 9.5 to 15.3; P<0.001). We observed similar treatment effects (favoring the closed-loop system) on the percentage of time that the glucose level was above 250 mg per deciliter, on the mean glucose level, and on the glycated hemoglobin level, with no significant between-group difference in the percentage of time that the glucose level was below 70 mg per deciliter. There were two cases of severe hypoglycemia in the closed-loop group and one case in the standard-care group. One case of diabetic ketoacidosis occurred in the closed-loop group.
In this trial involving young children with type 1 diabetes, the glucose level was in the target range for a greater percentage of time with a closed-loop system than with standard care. (Funded by the National Institute of Diabetes and Digestive and Kidney Diseases; PEDAP ClinicalTrials.gov number, NCT04796779.).
Intense exercise is a major challenge to the management of type 1 diabetes (T1D). Closed-loop control (CLC) systems (artificial pancreas) improve glycemic control during limited intensity and short ...duration of physical activity (PA). However, CLC has not been tested during extended vigorous outdoor exercise common among adolescents.
Skiing presents unique metabolic challenges: intense prolonged PA, cold, altitude, and stress/fear/excitement. In a randomized controlled trial, 32 adolescents with T1D (ages 10-16 years) participated in a 5-day ski camp (∼5 h skiing/day) at two sites: Wintergreen, VA, and Breckenridge, CO. Participants were randomized to the University of Virginia CLC system or remotely monitored sensor-augmented pump (RM-SAP). The CLC and RM-SAP groups were coarsely paired by age and hemoglobin A
(HbA
). All subjects were remotely monitored 24 h per day by the study physicians and clinical team.
Compared with physician-monitored open loop, percent time in range (70-180 mg/dL) improved using CLC: 71.3 vs. 64.7% (+6.6% 95% CI 1-12;
= 0.005), with maximum effect late at night. Hypoglycemia exposure and carbohydrate treatments were improved overall (
= 0.001 and
= 0.007) and during the daytime with strong ski level effects (
= 0.0001 and
= 0.006); ski/snowboard proficiency was balanced between groups but with a very strong site effect: naive in Virginia and experienced in Colorado. There was no adverse event associated with CLC; the participants' feedback was overwhelmingly positive.
CLC in adolescents with T1D improved glycemic control and reduced exposure to hypoglycemia during prolonged intensive winter sport activities, despite the added challenges of cold and altitude.
Insulin pumps are commonly used for intensive insulin therapy to treat type 1 diabetes in adults and youth. Insulin pump technologies have advanced dramatically in the last several years to integrate ...with continuous glucose monitors (CGM) and incorporate control algorithms. These control algorithms automate some insulin delivery in response to the glucose information received from the CGM to reduce the occurrence of hypoglycemia and hyperglycemia and improve overall glycemic control. The t:slim X2 insulin pump system became commercially available in 2016. It is an innovative insulin pump technology that can be updated remotely by the user to install new software onto the pump device as new technologies become available. Currently, the t:slim X2 pairs with the Dexcom G6 CGM and there are two advanced software options available: Basal-IQ, which is a predictive low glucose suspend (PLGS) technology, and Control-IQ, which is a Hybrid Closed Loop (HCL) technology. This paper will describe the different types of advanced insulin pump technologies, review how the t:slim X2 insulin pump works, and summarize the clinical studies leading to FDA approval and commercialization of the Basal-IQ and Control-IQ technologies.
Fear of diabetes and major surgery may prohibit referral of young children severely affected by pancreatitis for total pancreatectomy with islet autotransplant (TPIAT). We evaluated outcomes in our ...youngest TPIAT recipients, 3 to 8 years of age at surgery.
Medical records were reviewed for 17 children (9 girls) ages 8 years or younger undergoing TPIAT from 2000 to 2014. Most (14/17) had genetic risk factors for pancreatitis. Since 2006, TPIAT recipients were followed prospectively with health questionnaires including assessments of pain and narcotic use, and scheduled hemoglobin A1c (HbA1c) and mixed-meal tolerance tests (6 mL/kg Boost HP) before surgery, and at regular intervals after. Patients are 1 to 11 years post-TPIAT (median 2.2 years). Data are reported as median (25th, 75th percentile).
All had relief of pain, with all 17 patients off narcotics at most recent follow-up. Hospitalization rates decreased from 5.0 hospitalization episodes per person-year of follow-up before TPIAT, to 0.35 episodes per person-year of follow-up after TPIAT. Fourteen (82%) discontinued insulin, higher than the observed insulin independence rate of 41% in 399 patients older than 8 years of age undergoing TPIAT over the same interval (P = 0.004). Median post-TPIAT HbA1c was 5.9% (5.6%, 6.3%), and within patient post-TPIAT mean HbA1c was ≤6.5% for all but 2 patients.
Young children with severe refractory chronic pancreatitis may be good candidates for TPIAT, with high rates of pain relief and insulin independence, and excellent glycemic control in the majority.
: Automated insulin delivery for people with type 1 diabetes has been a major goal in the diabetes technology field for many years. While a fully automated system has not yet been accomplished, the ...MiniMed™ 670G artificial pancreas (AP) system is the first commercially available insulin pump that automates basal insulin delivery, while still requiring user input for insulin boluses. Determining the safety and efficacy of this system is essential to the development of future devices striving for more automation.
: This review will provide an overview of how the MiniMed 670G system works including its safety and efficacy, how it compares to similar devices, and anticipated future advances in diabetes technology currently under development.
: The ultimate goal of advanced diabetes technologies is to reduce the burden and amount of management required of patients with diabetes. In addition to reducing patient workload, achieving better glucose control and improving hemoglobin A1c (HbA1c) values are essential for reducing the threat of diabetes-related complications further down the road. Current devices come close to reaching these goals, but understanding the unmet needs of patients with diabetes will allow future technologies to achieve these goals more quickly.
To describe real-world outcomes for youth using the Tandem t:slim X2 insulin pump with Control-IQ technology ("Control-IQ") for 6 months at a large pediatric clinic.
Youth with type 1 diabetes, who ...started Control-IQ for routine care, were prospectively followed. Data on system use and glycemic control were collected before Control-IQ start, and at 1, 3, and 6 months after start. Mixed models assessed change across time; interactions with baseline hemoglobin A1c (HbA1c) and age were tested.
In 191 youth (median age 14, 47% female, and median HbA1c 7.6%), percent time with glucose levels 70-180 mg/dL (time-in-range TIR) improved from 57% at baseline to 66% at 6 months (
< 0.001). The proportion of participants reaching TIR target (>70%) doubled from 23.5% at baseline to 47.8% at 3 months, sustaining at 46.7% at 6 months (
< 0.001). Glucose management indicator (approximation of HbA1c) improved from 7.5% at baseline to 7.1% at 3 months and 7.2% at 6 months (
< 0.001). Those with higher baseline HbA1c experienced the most substantial improvements in glycemic control. Percent time using the Control-IQ feature was 86.4% at 6 months, and <4% of cohort discontinued use.
The Control-IQ system clinically and significantly improved glycemic control in a large sample of youth. System use was high at 6 months, with only a small proportion discontinuing use, indicating potential for sustaining results long term.