•CDC SARS-CoV-2 RT-PCR requires three reactions per sample for N1, N2, and RP targets.•Triplex RT-qPCR assay developed for N1, N2, and RP.•Assay sensitivity of 97.7% and limit of detection of 1000 ...copies/mL of sample.•4. Triplex assay saves time and cost effective alternative to standard CDC protocol.
Several RT-qPCR kits are available for SARS-CoV-2 diagnosis and some have emergency use authorization from the US Food and Drug Administration. In particular, the nCoV19 CDC kit includes two targets for detecting SARS-CoV-2 (N1 and N2) and an RNaseP (RP) target for RNA extraction quality control, all of which are labeled with FAM, and thus three PCR reactions are required per sample.
We designed a triplex RT-qPCR assay based on nCoV19 primers and probes where N1, N2, and RP are labeled with FAM, HEX, and Cy5, respectively, so only a single PCR reaction is required for each sample for SARS-CoV-2 diagnosis.
In total, 172 samples were analyzed in both singleplex and triplex assays, where 86 samples tested SARS-CoV-2 negative with both assays, so the triplex assay specificity was 100%. In addition, 86 samples tested SARS-Co-V 2 positive with the singleplex assay and 84 with the triplex assay, so the sensitivity was 97.7%. The limit of detection for the triplex assay was determined as 1000 copies/mL.
This new triplex RT-qPCR assay based on primers and probes from the CDC protocol is highly reliable for SARS-CoV-2 diagnosis, and it could speed up detection and save reagents during the current SARS-CoV-2 testing supplies shortage.
Dozens of commercial RT-qPCR kits for SARS-CoV-2 detection are available with or without Emergency Use Authorization (EUA) by FDA or other regulatory agencies.
We evaluated the clinical performance ...of two SARS-CoV-2 RT-PCR kits designed and produced in South America, "COVID-19 RT-PCR Real TM FAST (CY5)" (ATGen, Uruguay) and "ECUGEN SARS-CoV-2 RT-qPCR" (UDLA-STARNEWCORP, Ecuador), for RT-qPCR SARS-CoV2 detection using "TaqMan 2019-nCoV Assay Kit v1" (Thermofisher, USA) as a gold standard technique.
We report a great clinical performance and analytical sensitivity for the two South American kits with sensitivity values of 96.4 and 100%, specificity of 100% and limit of detection in the range of 10 copies/uL of RNA extraction.
"COVID-19 RT-PCR Real TM FAST (CY5)" and "ECUGEN SARS-CoV-2 RT-qPCR" kits are reliable SARS-CoV-2 tests made in South America that have been extensively used in Uruguay, Argentina, Brazil, Bolivia and Ecuador. These locally produced SARS-CoV-2 tests have contributed to overcome supply shortages and reduce diagnosis cost, while maintaining the high quality standards of FDA EUA commercially available kits. This approach could be extended for other diagnostic products to improve infectious diseases surveillance at middle and low income countries beyond COVID-19 pandemic.
•“Isopollo COVID19 detection kit” for RT-LAMP based SARS-CoV-2 diagnosis show a poor clinical performance with sensitivity of 61.9.•“Isopollo COVID19 detection kit” presented a good sensitivity only ...for viral loads above 100.000 copies/mL of sample.•Emergency use authorization at country of origin should be mandatory for SARS-CoV-2 RT-PCR diagnosis kits prior to importation to developing countries like Ecuador.
Multiple molecular kits are available for the diagnosis of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) worldwide, with many lacking proper clinical evaluation due to the emergency caused by the coronavirus disease 2019 (COVID-19) pandemic, particularly in developing countries.
This study was conducted to evaluate the clinical performance of the Isopollo COVID-19 detection kit (M Monitor, South Korea) for reverse transcription loop-mediated isothermal amplification (RT-LAMP) SARS-CoV-2 diagnosis, using the SARS-CoV-2 reverse transcription polymerase chain reaction (RT-PCR) protocol as the gold standard.
A total of 220 clinical samples were included in the study; 168 samples were SARS-CoV-2-positive and 52 samples were SARS-CoV-2-negative according to the SARS-CoV-2 RT-PCR protocol. For the Isopollo COVID-19 detection kit, only 104 out of 168 samples were SARS-CoV-2-positive. This result shows a low clinical performance, with sensitivity of 61.9% for the evaluated RT-LAMP assay.
Proper clinical performance evaluation studies by regulatory agencies in developing countries such as Ecuador should be mandatory prior to clinical use authorization of SARS-CoV-2 diagnosis kits, particularly when those kits lack either US Food and Drug Administration or country of origin clinical use authorization.
Background
Multiple RT-qPCR kits are available in the market for SARS-CoV-2 diagnosis, some of them with Emergency Use Authorization (EUA) by FDA or their country of origin agency, but many of them ...lack of proper clinical evaluation.
Objective
We evaluated the clinical performance of two Korean SARS-CoV-2 RT-PCR kits available in South America, AccuPower SARS-CoV-2 Multiplex RT-PCR kit (Bioneer, South Korea) and Allplex 2019-nCoV Assay (Seegene, South Korea), for RT-qPCR SARS-CoV-2 diagnosis using the CDC protocol as a gold standard.
Results
We found strong differences among both kits clinical performance and analytical sensitivity; while the Allplex 2019-nCoV Assay has sensitivity of 96.5% and an estimated limit of detection of 4,000 copies/ml, the AccuPower SARS-CoV-2 Multiplex RT-PCR kit has a sensitivity of 75.5% and limit of detection estimated to be bigger than 20,000 copies/ml.
Conclusions
AccuPower SARS-CoV-2 Multiplex RT-PCR kit and Allplex 2019-nCoV Assay are both made in South Korea but EUA by Korean CDC was only granted to the later. Our results support that Korean CDC EUA should be considered as a quality control proxy for Korean SARS-CoV-2 RT-PCR kits prior to importation by developing countries to guarantee high sensitivity diagnosis.
Several molecular kits are available for SARS-CoV-2 diagnosis, mostly lacking of proper clinical evaluation due to the emergency caused by COVID19 pandemia, particularly at developing countries like ...Ecuador.
We carried out an evaluation of the clinical performance of "AccuPower SARS-CoV-2 Real Time RT-PCR kit" (Bioneer, South Korea) for SARS-CoV-2 diagnosis using 2019-nCoV CDC EUA kit (IDT, USA) as a gold standard.
48 clinical specimens were included on the study, 38 tested SARS-CoV-2 positive and 10 SARS-CoV-2 negative for 2019-nCoV CDC EUA kit. For "AccuPower SARS-CoV-2 Real Time RT-PCR kit", only 30 were SARS-CoV-2 positive, indicating a low clinical performance with sensitivity of 78.9%. Moreover, the limit of detection for "AccuPower SARS-CoV-2 Real Time RT-PCR kit" was estimated to be higher than 40,000 viral RNA copies/mL of sample.
Proper clinical performance evaluation studies from government agencies at developing countries should be mandatory prior to clinical use authorization of SARS-CoV-2 diagnosis kits, particularly when those kits lack of either FDA or its country of origin clinical use authorization, to prevent the distribution of low quality products that may have a negative impact of COVID19 surveillance at developing countries.
Hundreds of RT-qPCR kits are available in the market for SARS-CoV-2 diagnosis, some of them with emergency use authorization (EUA) by the Food Drug Administration (FDA) or their country of origin ...agency, but also many of them without any independent clinical performance evaluation. We performed a clinical evaluation for two Chinese SARS-CoV-2 RT-PCR kits available in South America, COVID-19 Nucleic Acid Test Kit (eDiagnosis Biomedicine, Wuhan, China) and 2019-nCoV Nucleic Acid Diagnostic Kit (Sansure Biotech, Changsha, China), for RT-qPCR SARS-CoV-2 diagnosis using the FDA EUA 2019-nCoV CDC kit (IDT, Coralville, IA) as gold standard. We found an excellent clinical performance and analytical sensitivity for both kits with sensitivity values of 100% and 95.3% and estimated limit of detection of 500 copies/mL and 1,000 copies/mL, for eDiagnosis and Sansure Biotech kits, respectively. COVID-19 Nucleic Acid Test Kit (eDiagnosis) and 2019-nCoV Nucleic Acid Diagnostic Kit (Sansure Biotech) are both made in China and hold EUA by the Chinese CDC. Also, Sansure Biotech kit has EUA by the FDA. In conclusion, our results endorse the use of these two commercially available kits imported to Ecuador for SARS-CoV-2 diagnosis, as they had the similar clinical performance as the gold standard from the CDC.
CDC and WHO guidelines for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) diagnosis only recommend synthetic fiber swabs for nasopharyngeal (NP) sampling. We show that cotton-tipped ...plastic swabs do not inhibit PCR and have equivalent performance to rayon swabs. Cotton-tipped plastic swabs are massively produced worldwide and would prevent swab supply shortages under the current high SARS-CoV-2 testing demands, particularly in developing countries.