Integral field spectroscopy represents a powerful technique for the detection and characterization of extrasolar planets through high-contrast imaging since it allows us to obtain simultaneously a ...large number of monochromatic images. These can be used to calibrate and then to reduce the impact of speckles, once their chromatic dependence is taken into account. The main concern in designing integral field spectrographs for high-contrast imaging is the impact of the diffraction effects and the noncommon path aberrations together with an efficient use of the detector pixels. We focus our attention on integral field spectrographs based on lenslet arrays, discussing the main features of these designs: the conditions of appropriate spatial and spectral sampling of the resulting spectrograph's slit functions and their related cross-talk terms when the system works at the diffraction limit. We present a new scheme for the integral field unit based on a dual-lenslet device (BIGRE), that solves some of the problems related to the classical Traitement Integral des Galaxies par l'Etude de leurs Rays (TIGER) design when used for such applications. We show that BIGRE provides much lower cross-talk signals than TIGER, allowing a more efficient use of the detector pixels and a considerable saving of the overall cost of a lenslet-based integral field spectrograph.
Context.
Circumstellar disks and self-luminous giant exoplanets or companion brown dwarfs can be characterized through direct-imaging polarimetry at near-infrared wavelengths. SPHERE/IRDIS at the ...Very Large Telescope has the capabilities to perform such measurements, but uncalibrated instrumental polarization effects limit the attainable polarimetric accuracy.
Aims.
We aim to characterize and correct the instrumental polarization effects of the complete optical system, that is, the telescope and SPHERE/IRDIS.
Methods.
We created a detailed Mueller matrix model in the broadband filters
Y, J, H,
and
K
s
and calibrated the model using measurements with SPHERE’s internal light source and observations of two unpolarized stars. We developed a data-reduction method that uses the model to correct for the instrumental polarization effects, and applied it to observations of the circumstellar disk of T Cha.
Results.
The instrumental polarization is almost exclusively produced by the telescope and SPHERE’s first mirror and varies with telescope altitude angle. The crosstalk primarily originates from the image derotator (K-mirror). At some orientations, the derotator causes severe loss of signal (> 90% loss in the
H
- and
K
s
-band) and strongly offsets the angle of linear polarization. With our correction method we reach, in all filters, a total polarimetric accuracy of ≲0.1% in the degree of linear polarization and an accuracy of a few degrees in angle of linear polarization.
Conclusions.
The correction method enables us to accurately measure the polarized intensity and angle of linear polarization of circumstellar disks, and is a vital tool for detecting spatially unresolved (inner) disks and measuring the polarization of substellar companions. We have incorporated the correction method in a highly-automated end-to-end data-reduction pipeline called IRDAP, which we made publicly available online.
Second-generation exoplanet imagers using extreme adaptive optics (ExAO) and coronagraphy have demonstrated their great potential for studying close circumstellar environments and for detecting new ...companions and helping to understand their physical properties. However, at very small angular separation, their performance in contrast is limited by several factors: diffraction by the complex telescope pupil (central obscuration and spiders) not perfectly canceled by the coronagraph, residual dynamic wavefront errors, chromatic wavefront errors, and wavefront errors resulting from noncommon path aberrations (NCPAs). These latter are differential aberrations between the visible wavefront sensing path of the ExAO system and the near-infrared science path in which the coronagraph is located. In a previous work, we demonstrated the use of a Zernike wavefront sensor called ZELDA for sensing NCPAs in the VLT/SPHERE exoplanet imager and their compensation with the high-order deformable mirror of the instrument. These early tests on the internal light source led to encouraging results for the attenuation of the quasi-static speckles at very small separation. In the present work, we move to the next step with the on-sky validation of NCPA compensation with ZELDA. With an improved procedure for the compensation of NCPAs, we start by reproducing previous results on the internal source. We show that the amount of aberration integrated between 1 and 15 cycles/pupil (c/p) is decreased by a factor of approximately five, which translates into a gain in raw contrast of between 2 and 3 at separations below 300 mas. On sky, we demonstrate that NCPA compensation works in closed loop, leading to an attenuation of the amount of aberration by a factor of approximately two. However, we identify a loss of sensitivity for the sensor that is only partly explained by the difference in Strehl ratio between the internal and on-sky measurements. Our simulations show that the impact of ExAO residuals on ZELDA measurements is negligible for integration times beyond a few tenths of a second. Coronagraphic imaging on sky is improved in raw contrast by a factor of 2.5 at most in the ExAO-corrected region. We use coronagraphic image reconstruction based on a detailed model of the instrument to demonstrate that both internal and on-sky raw contrasts can be precisely explained, and we establish that the observed performance after NCPA compensation is no longer limited by an improper compensation for aberration but by the current apodized-pupil Lyot coronagraph design. We finally conclude that a coronagraph upgrade combined to a proper NCPA compensation scheme could easily bring a gain in raw contrast of a factor of two to three below 200 mas.
Context.
Polarimetric imaging is one of the most effective techniques for high-contrast imaging and for the characterization of protoplanetary disks, and it has the potential of becoming instrumental ...in the characterization of exoplanets. The Spectro-Polarimetric High-contrast Exoplanet REsearch (SPHERE) instrument installed on the Very Large Telescope (VLT) contains the InfraRed Dual-band Imager and Spectrograph (IRDIS) with a dual-beam polarimetric imaging (DPI) mode, which offers the capability of obtaining linear polarization images at high contrast and resolution.
Aims.
We aim to provide an overview of the polarimetric imaging mode of VLT/SPHERE/IRDIS and study its optical design to improve observing strategies and data reduction.
Methods.
For
H
-band observations of TW Hydrae, we compared two data reduction methods that correct for instrumental polarization effects in different ways: a minimization of the “noise” image (
U
ϕ
), and a correction method based on a polarimetric model that we have developed, as presented in Paper II of this study.
Results.
We use observations of TW Hydrae to illustrate the data reduction. In the images of the protoplanetary disk around this star, we detect variability in the polarized intensity and angle of linear polarization that depend on the pointing-dependent instrument configuration. We explain these variations as instrumental polarization effects and correct for these effects using our model-based correction method.
Conclusions.
The polarimetric imaging mode of IRDIS has proven to be a very successful and productive high-contrast polarimetric imaging system. However, the instrument performance is strongly dependent on the specific instrument configuration. We suggest adjustments to future observing strategies to optimize polarimetric efficiency in field-tracking mode by avoiding unfavorable derotator angles. We recommend reducing on-sky data with the pipeline called IRDAP, which includes the model-based correction method (described in Paper II) to optimally account for the remaining telescope and instrumental polarization effects and to retrieve the true polarization state of the incident light.
Observations of circumstellar environments that look for the direct signal of exoplanets and the scattered light from disks have significant instrumental implications. In the past 15 years, major ...developments in adaptive optics, coronagraphy, optical manufacturing, wavefront sensing, and data processing, together with a consistent global system analysis have brought about a new generation of high-contrast imagers and spectrographs on large ground-based telescopes with much better performance. One of the most productive imagers is the Spectro-Polarimetic High contrast imager for Exoplanets REsearch (SPHERE), which was designed and built for the ESO Very Large Telescope (VLT) in Chile. SPHERE includes an extreme adaptive optics system, a highly stable common path interface, several types of coronagraphs, and three science instruments. Two of them, the Integral Field Spectrograph (IFS) and the Infra-Red Dual-band Imager and Spectrograph (IRDIS), were designed to efficiently cover the near-infrared range in a single observation for an efficient search of young planets. The third instrument, ZIMPOL, was designed for visible polarimetric observation to look for the reflected light of exoplanets and the light scattered by debris disks. These three scientific instruments enable the study of circumstellar environments at unprecedented angular resolution, both in the visible and the near-infrared. In this work, we thoroughly present SPHERE and its on-sky performance after four years of operations at the VLT.