Background
During laparoscopic cholecystectomy, common bile duct (CBD) injury is a rare but severe complication. To reduce the risk of injury, near-infrared (NIR) fluorescent cholangiography using ...indocyanine green (ICG) has recently been introduced as a novel method of visualizing the biliary system during surgery. To date, several studies have shown feasibility of this technique; however, liver background fluorescence remains a major problem during fluorescent cholangiography. The aim of the current study was to optimize ICG dose and timing for NIR cholangiography using a quantitative intraoperative camera system during open hepatopancreatobiliary (HPB) surgery. Subsequently, these results were validated during laparoscopic cholecystectomy using a laparoscopic fluorescence imaging system.
Methods
Twenty-seven patients who underwent NIR imaging using the Mini-FLARE image-guided surgery system during open HPB surgery were analyzed to assess optimal dosage and timing of ICG administration. ICG was intravenously injected preoperatively at doses of 5, 10, and 20 mg, and imaged at either 30 min (early) or 24 h (delayed) post-injection. Next, the optimal doses found for early and delayed imaging were applied to two groups of seven patients (
n
= 14) undergoing laparoscopic NIR fluorescent cholangiography during laparoscopic cholecystectomy.
Results
Median liver-to-background contrast was 23.5 (range 22.1–35.0), 16.8 (range 11.3–25.1), 1.3 (range 0.7–7.8), and 2.5 (range 1.3–3.6) for 5 mg/30 min, 10 mg/30 min, 10 mg/24 h, and 20 mg/24 h, respectively. Fluorescence intensity of the liver was significantly lower in the 10 mg delayed-imaging dose group compared with the early imaging 5 and 10 mg dose groups (
p
= 0.001), which resulted in a significant increase in CBD-to-liver contrast ratio compared with the early administration groups (
p
< 0.002). These findings were qualitatively confirmed during laparoscopic cholecystectomy.
Conclusion
This study shows that a prolonged interval between ICG administration and surgery permits optimal NIR cholangiography with minimal liver background fluorescence.
In ovarian cancer, two of the most important prognostic factors for survival are completeness of staging and completeness of cytoreductive surgery. Therefore, intra-operative visualization of tumor ...lesions is of great importance. Preclinical data already demonstrated tumor visualization in a mouse-model using near-infrared (NIR) fluorescence imaging and indocyanine green (ICG) as a result of enhanced permeability and retention (EPR). The aim of this study was to determine feasibility of intraoperative ovarian cancer metastases imaging using NIR fluorescence imaging and ICG in a clinical setting.
Ten patients suspected of ovarian cancer scheduled for staging or cytoreductive surgery were included. Patients received 20 mg ICG intravenously after opening the abdominal cavity. The mini-FLARE NIR fluorescence imaging system was used to detect NIR fluorescent lesions.
6 out of 10 patients had malignant disease of the ovary or fallopian tube, of which 2 had metastatic disease outside the pelvis. Eight metastatic lesions were detected in these 2 patients, which were all NIR fluorescent. However, 13 non-malignant lesions were also NIR fluorescent, resulting in a false-positive rate of 62%. There was no significant difference in tumor-to-background ratio between malignant and benign lesions (2.0 vs 2.0; P=0.99).
This is the first clinical trial demonstrating intraoperative detection of ovarian cancer metastases using NIR fluorescence imaging and ICG. Despite detection of all malignant lesions, a high false-positive rate was observed. Therefore, NIR fluorescence imaging using ICG based on the EPR effect is not satisfactory for the detection of ovarian cancer metastases. The need for tumor-specific intraoperative agents remains.
ISRCTN Registry ISRCTN16945066.
Background Identification of diseased and normal parathyroid glands during parathyroid surgery can be challenging. The aim of this study was to assess whether near-infrared (NIR) fluorescence imaging ...using administration of a low-dose Methylene Blue (MB) at the start of the operation could provide optical guidance during parathyroid surgery and assist in the detection of parathyroid adenomas. Methods Patients diagnosed with primary hyperparathyroidism planned for parathyroidectomy were included. Patients received 0.5 mg/kg MB intravenously directly after start of anesthesia. During the operation, NIR fluorescence imaging was performed to identify parathyroid adenomas. Imaging results were compared with a previous published feasibility study in which 12 patients received MB after intraoperative identification of the adenoma. Results A total of 13 patients were included in the current study. In 10 of 12 patients with a histologically proven adenoma, the adenoma was fluorescent. Mean signal to background ratio was 3.1 ± 2.8. Mean diameter of the resected lesions was 17 ± 9 mm (range 5–28 mm). Adenomas could be identified up to 145 minutes after administration, which was the longest timespan until resection. Interestingly, in 3 patients, a total of 6 normal parathyroid glands (median diameter 2.5 mm) with a signal to background ratio of 1.8 ± 0.4 were identified using NIR fluorescence imaging. Conclusion Early administration of low-dose MB provided guidance during parathyroidectomy by identifying both parathyroid adenomas and normal parathyroid glands. In patients in whom difficult identification of the parathyroid adenoma is expected or when normal glands have to be identified, the administration of MB may improve surgical outcome.
Abstract Background Despite recent developments in preoperative breast cancer imaging, intraoperative localization of tumor tissue can be challenging, resulting in tumor-positive resection margins ...during breast conserving surgery. Based on certain physicochemical similarities between Technetium(99m Tc)-sestamibi (MIBI), an SPECT radiodiagnostic with a sensitivity of 83–90% to detect breast cancer preoperatively, and the near-infrared (NIR) fluorophore Methylene Blue (MB), we hypothesized that MB might detect breast cancer intraoperatively using NIR fluorescence imaging. Methods Twenty-four patients with breast cancer, planned for surgical resection, were included. Patients were divided in 2 administration groups, which differed with respect to the timing of MB administration. N = 12 patients per group were administered 1.0 mg/kg MB intravenously either immediately or 3 h before surgery. The mini-FLARE imaging system was used to identify the NIR fluorescent signal during surgery and on post-resected specimens transferred to the pathology department. Results were confirmed by NIR fluorescence microscopy. Results 20/24 (83%) of breast tumors (carcinoma in N = 21 and ductal carcinoma in situ in N = 3) were identified in the resected specimen using NIR fluorescence imaging. Patients with non-detectable tumors were significantly older. No significant relation to receptor status or tumor grade was seen. Overall tumor-to-background ratio (TBR) was 2.4 ± 0.8. There was no significant difference between TBR and background signal between administration groups. In 2/4 patients with positive resection margins, breast cancer tissue identified in the wound bed during surgery would have changed surgical management. Histology confirmed the concordance of fluorescence signal and tumor tissue. Conclusions This feasibility study demonstrated an overall breast cancer identification rate using MB of 83%, with real-time intraoperative guidance having the potential to alter patient management.
Intraoperative fluorescence imaging of the folate-receptor alpha (FRα) could support completeness of resection in cancer surgery. Feasibility of EC17, a FRα-targeting agent that fluoresces at 500nm, ...was demonstrated in a limited series of ovarian cancer patients. Our objective was to evaluate EC17 in a larger group of ovarian cancer patients. In addition, we assessed the feasibility of EC17 in patients with breast cancer.
Two-to-three hours before surgery 0.1mg/kg EC17 was intravenously administered to 12 patients undergoing surgery for ovarian cancer and to 3 patients undergoing surgery for biopsy-proven FRα-positive breast cancer. The number of lesions/positive margins detected with fluorescence and concordance between fluorescence and tumor- and FRα-status was assessed in addition to safety and pharmacokinetics.
Fluorescence imaging in ovarian cancer patients allowed detection of 57 lesions of which 44 (77%) appeared malignant on histopathology. Seven out of these 44 (16%) were not detected with inspection/palpation. Histopathology demonstrated concordance between fluorescence and FRα- and tumor status. Fluorescence imaging in breast cancer patients, allowed detection of tumor-specific fluorescence signal. At the 500nm wavelength, autofluorescence of normal breast tissue was present to such extent that it interfered with tumor identification.
FRα is a favorable target for fluorescence-guided surgery as EC17 produced a clear fluorescent signal in ovarian and breast cancer tissue. This resulted in resection of ovarian cancer lesions that were otherwise not detected. Notwithstanding, autofluorescence caused false-positive lesions in ovarian cancer and difficulty in discriminating breast cancer-specific fluorescence from background signal. Optimization of the 500nm fluorophore, will minimize autofluorescence and further improve intraoperative tumor detection.
We hypothesized that T-cell immune interaction affects tumor development and thus clinical outcome. Therefore, we examined the clinical impact of human leukocyte antigen (HLA) class I tumor cell ...expression and regulatory T-cell (Treg) infiltration in breast cancer.
Our study population (N = 677) is consisted of all early breast cancer patients primarily treated with surgery in our center between 1985 and 1994. Formalin-fixed, paraffin-embedded tumor tissue was immunohistochemically stained using HCA2, HC10, and Foxp3 monoclonal antibodies.
HLA class I expression was evaluated by combining results from HCA2 and HC10 antibodies and classified into three groups: loss, downregulation, and expression. Remarkably, only in patients who received chemotherapy, both presence of Treg (P = 0.013) and higher HLA class I expression levels (P = 0.002) resulted in less relapses, independently of other variables. Treg and HLA class I were not of influence on clinical outcome in patients who did not receive chemotherapy.
We showed that HLA class I and Treg affect prognosis exclusively in chemotherapy-treated patients and are therefore one of the few predictive factors for chemotherapy response in early breast cancer patients. Chemotherapy may selectively eliminate Treg, thus enabling CTLs to kill tumor cells that have retained HLA class I expression. As a consequence, HLA class I and Treg can predict response to chemotherapy with high discriminative power. These markers could be applied in response prediction to chemotherapy in breast cancer patients.
Background
Irradical tumor resections and iatrogenic ureteral injury remain a significant problem during lower abdominal surgery. The aim of the current study was to intraoperatively identify both ...colorectal tumors and ureters in subcutaneous and orthotopic animal models using cRGD-ZW800-1 and near-infrared (NIR) fluorescence.
Methods
The zwitterionic fluorophore ZW800-1 was conjugated to the tumor specific peptide cRGD (targeting integrins) and to the a-specific peptide cRAD. One nmol cRGD-ZW800-1, cRAD-ZW800-1, or ZW800-1 alone was injected in mice bearing subcutaneous HT-29 human colorectal tumors. Subsequently, cRGD-ZW800-1 was injected at dosages of 0.25 and 1 nmol in mice bearing orthotopic HT-29 tumors transfected with luciferase2. In vivo biodistribution and ureteral visualization were investigated in rats. Fluorescence was measured intraoperatively at several time points after probe administration using the FLARE imaging system.
Results
Both subcutaneous and orthotopic tumors could be clearly identified using cRGD-ZW800-1. A significantly higher signal-to-background ratio was observed in mice injected with cRGD-ZW800-1 (2.42 ± 0.77) compared with mice injected with cRAD-ZW800-1 or ZW800-1 alone (1.21 ± 0.19 and 1.34 ± 0.19, respectively) when measured at 24 h after probe administration. The clearance of cRGD-ZW800-1 permitted visualization of the ureters and also generated minimal background fluorescence in the gastrointestinal tract.
Conclusions
This study appears to be the first to demonstrate both clear tumor demarcation and ureteral visualization after a single intravenous injection of a targeted NIR fluorophore. As a low dose of cRGD-ZW800-1 provided clear tumor identification, clinical translation of these results should be possible.
Near-infrared (NIR) fluorescence imaging using indocyanine green (ICG) has recently been introduced to improve the sentinel lymph node (SLN) procedure. Several optical tracers have been successfully ...tested. However, the optimal tracer formulation is still unknown. This study evaluates the performance of ICG-technetium-99m (99mTc)-nanocolloid in relation to 2 most commonly used ICG-based formulas during SLN biopsy in vulvar cancer.
Twelve women who planned to undergo SLN biopsy for stage I vulvar cancer were prospectively included. Sentinel lymph node mapping was performed using the dual-modality radioactive and NIR fluorescence tracer ICG-99mTc-nanocolloid. All patients underwent combined SLN localization using NIR fluorescence and the (current) gold standard using blue dye and radioactive guidance.
In all 12 patients, at least 1 SLN was detected during surgery. A total of 21 lymph nodes (median 2; range, 1-3) were resected. Median time between skin incision and first SLN detection was 8 (range, 1-22) minutes. All resected SLNs were both radioactive and fluorescent, although only 13 (62%) of 21 SLNs stained blue. Median brightness of exposed SLNs, expressed as signal-to-background ratio, was 5.4 (range, 1.8-11.8). Lymph node metastases were found in 3 patients.
Near-infrared fluorescence-guided SLN mapping is feasible and outperforms blue dye staining. Premixing ICG with 99mTc-nanocolloid provides real-time intraoperative imaging of the SN and seems to be the optimal tracer combination in terms of intraoperative detection rate of the SN (100%). Moreover, ICG-99mTc-nanocolloid allows the administration of a 5-times lower injected dose of ICG (compared with ICG and ICG absorbed to human serum albumin) and can be injected up to 20 hours before surgery.