Pancreatic cancer is the fourth leading cause of cancer-related mortality in the United States. The minority of patients can undergo curative-intended surgical therapy due to progressive disease ...stage at time of diagnosis. Nonetheless, tumor involvement of surgical margins is seen in up to 70% of resections, being a strong negative prognostic factor. Real-time intraoperative imaging modalities may aid surgeons to obtain tumor-free resection margins. Full-field optical coherence tomography (FF-OCT) is a promising diagnostic tool using high-resolution white-light interference microscopy without tissue processing. Therefore, we composed an atlas of FF-OCT images of malignant and benign pancreatic tissue, and investigated the accuracy with which the pathologists could distinguish these.
One hundred FF-OCT images were collected from specimens of 29 patients who underwent pancreatic resection for various indications between 2014 and 2016. One experienced gastrointestinal pathologist and one pathologist in training scored independently the FF-OCT images as malignant or benign blinded to the final pathology conclusion. Results were compared to those obtained with standard hematoxylin and eosin (H&E) slides.
Overall, combined test characteristics of both pathologists showed a sensitivity of 72%, specificity of 74%, positive predictive value of 69%, negative predictive value of 79% and an overall accuracy of 73%. In the subset of pancreatic ductal adenocarcinoma patients, 97% of the FF-OCT images (n = 35) were interpreted as tumor by at least one pathologist. Moreover, normal pancreatic tissue was recognised in all cases by at least one pathologist. However, atrophy and fibrosis, serous cystadenoma and neuroendocrine tumors were more often wrongly scored, in 63%, 100% and 25% respectively.
FF-OCT could distinguish normal pancreatic tissue from pathologic pancreatic tissue in both processed as non-processed specimens using architectural features. The accuracy in pancreatic ductal adenocarcinoma is promising and warrants further evaluation using improved assessment criteria.
Targeted real-time imaging during robot-assisted radical prostatectomy provides information on the localisation and extent of prostate cancer. We assessed the safety and feasibility of the ...prostate-specific membrane antigen (PSMA)-targeted fluorescent tracer OTL78 in patients with prostate cancer.
In this single-arm, phase 2a, feasibility trial with an adaptive design was carried out in The Netherlands Cancer Institute, Netherlands. Male patients aged 18 years or older, with PSMA PET-avid prostate cancer with an International Society of Urological Pathology (ISUP) grade group of 2 or more, who were scheduled to undergo robot-assisted radical prostatectomy with or without extended pelvic lymph node dissection were eligible. All patients had a robot-assisted radical prostatectomy using OTL78. Based on timing and dose, patients received a single intravenous infusion of OTL78 (0·06 mg/kg 1–2 h before surgery dose cohort 1, 0·03 mg/kg 1–2 h before surgery dose cohort 2, or 0·03 mg/kg 24 h before surgery dose cohort 3). The primary outcomes, assessed in all enrolled patients, were safety and pharmacokinetics of OTL78. This study is completed and is registered in the European Trial Database, 2019-002393-31, and the International Clinical Trials Registry Platform, NL8552, and is completed.
Between June 29, 2020, and April 1, 2021, 19 patients were screened for eligibility, 18 of whom were enrolled. The median age was 69 years (IQR 64–70) and median prostate-specific antigen concentration was 15 ng/mL (IQR 9·3–22·0). In 16 (89%) of 18 patients, robot-assisted radical prostatectomy was accompanied by an extended pelvic lymph node dissection. Three serious adverse events occurred in one (6%) patient: an infected lymphocele, a urosepsis, and an intraperitoneal haemorrhage. These adverse events were considered unrelated to the administration of OTL78 or intraoperative fluorescence imaging. No patient died, required a dose reduction, or required discontinuation due to drug-related toxicity. The dose-normalised maximum serum concentration (Cmax/dose) in patients was 84·1 ng/mL/mg for the 0·03 mg/kg dose and 79·6 ng/mL/mg for the 0·06 mg/kg dose, the half-life was 5·1 h for the 0·03 mg/kg dose and 4·7 h for the 0·06 mg/kg dose, the volume of distribution was 22·9 L for the 0·03 mg/kg dose and 19·5 L for the 0·06 mg/kg dose, and the clearance was 3·1 L/h for the 0·03 mg/kg dose and 3·0 L/h for the 0·06 mg/kg dose.
This first-in-patient study showed that OTL78 was well tolerated and had the potential to improve prostate cancer detection. Optimal dosing was 0·03 mg/kg, 24 h preoperatively. PSMA-directed fluorescence imaging allowed real-time identification of visually occult prostate cancer and might help to achieve complete oncological resections.
On Target Laboratories.
Molecular fluorescence-guided surgery using near-infrared light has the potential to improve the rate of complete resection of cancer. Typically, monoclonal antibodies are being used as targeting ...moieties, however smaller fragments, such as single-domain antibodies (i.e., Nanobodies
) improve tumor specificity and enable tracer injection on the same day as surgery. In this study, the feasibility of a carcinoembryonic antigen-targeting Nanobody (NbCEA5) conjugated to two zwitterionic dyes (ZW800-1 Forte ZW800F and ZW800-1) for visualization of pancreatic ductal adenocarcinoma (PDAC) was investigated. After site-specific conjugation of NbCEA5 to the zwitterionic dyes, binding specificity was evaluated on human PDAC cell lines with flow cytometry. A dose escalation study was performed for both NbCEA5-ZW800F and NbCEA5-ZW800-1 in mice with subcutaneously implanted pancreatic tumors. Fluorescence imaging was performed up to 24 h after intravenous injection. Furthermore, the optimal dose for NbCEA5-ZW800-1 was injected in mice with orthotopically implanted pancreatic tumors. A dose-escalation study showed superior mean fluorescence intensities for NbCEA5-ZW800-1 compared to NbCEA5-ZW800F. In the orthotopic tumor models, NbCEA5-ZW800-1 accumulated specifically in pancreatic tumors with a mean in vivo tumor-to-background ratio of 2.4 (SD = 0.23). This study demonstrated the feasibility and potential advantages of using a CEA-targeted Nanobody conjugated to ZW800-1 for intraoperative PDAC imaging.
The lack of multi-omics cancer datasets with extensive follow-up information hinders the identification of accurate biomarkers of clinical outcome. In this cohort study, we performed comprehensive ...genomic analyses on fresh-frozen samples from 348 patients affected by primary colon cancer, encompassing RNA, whole-exome, deep T cell receptor and 16S bacterial rRNA gene sequencing on tumor and matched healthy colon tissue, complemented with tumor whole-genome sequencing for further microbiome characterization. A type 1 helper T cell, cytotoxic, gene expression signature, called Immunologic Constant of Rejection, captured the presence of clonally expanded, tumor-enriched T cell clones and outperformed conventional prognostic molecular biomarkers, such as the consensus molecular subtype and the microsatellite instability classifications. Quantification of genetic immunoediting, defined as a lower number of neoantigens than expected, further refined its prognostic value. We identified a microbiome signature, driven by Ruminococcus bromii, associated with a favorable outcome. By combining microbiome signature and Immunologic Constant of Rejection, we developed and validated a composite score (mICRoScore), which identifies a group of patients with excellent survival probability. The publicly available multi-omics dataset provides a resource for better understanding colon cancer biology that could facilitate the discovery of personalized therapeutic approaches.
Photodynamic therapy (PDT) induces cell death through local light activation of a photosensitizer (PS) and has been used to treat head and neck cancers. Yet, common PS lack tumor specificity, which ...leads to collateral damage to normal tissues. Targeted delivery of PS via antibodies has pre-clinically improved tumor selectivity. However, antibodies have long half-lives and relatively poor tissue penetration, which could limit therapeutic efficacy and lead to long photosensitivity. Here, in this feasibility study, we evaluate at the pre-clinical level a recently introduced format of targeted PDT, which employs nanobodies as targeting agents and a water-soluble PS (IRDye700DX) that is traceable through optical imaging. In vitro, the PS solely binds to cells and induces phototoxicity on cells overexpressing the epidermal growth factor receptor (EGFR), when conjugated to the EGFR targeted nanobodies. To investigate whether this new format of targeted PDT is capable of inducing selective tumor cell death in vivo, PDT was applied on an orthotopic mouse tumor model with illumination at 1h post-injection of the nanobody–PS conjugates, as selected from quantitative fluorescence spectroscopy measurements. In parallel, and as a reference, PDT was applied with an antibody–PS conjugate, with illumination performed 24h post-injection. Importantly, EGFR targeted nanobody–PS conjugates led to extensive tumor necrosis (approx. 90%) and almost no toxicity in healthy tissues, as observed through histology 24h after PDT. Overall, results show that these EGFR targeted nanobody–PS conjugates are selective and able to induce tumor cell death in vivo. Additional studies are now needed to assess the full potential of this approach to improving PDT.
Display omitted
Near-infrared (NIR) fluorescence imaging using indocyanine green (ICG) has the potential to improve sentinel lymph node (SLN) mapping of breast cancer. We performed a randomized clinical trial to ...assess the value of blue dyes when used in combination with NIR fluorescence. We also preliminarily examined the possibility of performing SLN mapping without radiotracers.
Clinical trial subjects were 24 consecutive breast cancer patients scheduled to undergo SLN biopsy. All patients received standard of care using 99(m) technetium-nanocolloid and received 1.6 mL of 500 μM ICG injected periareolarly. Patients were randomly assigned to undergo SLN biopsy with or without patent blue. To assess the need for radiocolloids to localize the SLN or SLNs, the surgeon did not use the handheld gamma probe during the first 15 min after the axillary skin incision.
SLN mapping was successful in 23 of the 24 patients. No significant difference was found in signal-to-background ratio between the groups with and without patent blue (8.3 ± 3.8 vs. 10.3 ± 5.7, respectively, P = 0.32). In both groups, 100 % of SLNs were radioactive and fluorescent, and in the patent blue group, only 84 % of SLNs were stained blue. In 25 % of patients, the use of the gamma probe was necessary to localize the SLN within the first 15 min.
This study shows that there is no benefit of using patent blue for SLN mapping in breast cancer patients when using NIR fluorescence and 99(m) technetium-nanocolloid. NIR fluorescence imaging outperformed patent blue in all patients.
Incomplete oncologic resections and damage to vital structures during colorectal cancer surgery increases morbidity and mortality. Moreover, neoadjuvant chemoradiotherapy has become the standard ...treatment modality for locally advanced rectal cancer, where subsequent downstaging can make identification of the primary tumor more challenging during surgery. Near-infrared (NIR) fluorescence imaging can aid surgeons by providing real-time visualization of tumors and vital structures during surgery.
We present the first-in-human clinical experience of a novel NIR fluorescent peptide, cRGD-ZW800-1, for the detection of colon cancer. cRGD-ZW800-1 was engineered to have an overall zwitterionic chemical structure and neutral charge to lower nonspecific uptake and thus background fluorescent signal. We performed a phase I study in 11 healthy volunteer as well as a phase II feasibility study in 12 patients undergoing an elective colon resection, assessing 0.005, 0.015, and 0.05 mg/kg cRGD-ZW800-1 for the intraoperative visualization of colon cancer.
cRGD-ZW800-1 appears safe, and exhibited rapid elimination into urine after a single low intravenous dose. Minimal invasive intraoperative visualization of colon cancer through full-thickness bowel wall was possible after an intravenous bolus injection of 0.05 mg/kg at least 2 hours prior to surgery. Longer intervals between injection and imaging improved the tumor-to-background ratio.
cRGD-ZW800-1 enabled fluorescence imaging of colon cancer in both open and minimal invasive surgeries. Further development of cRGD-ZW800-1 for widespread use in cancer surgery may be warranted given the ubiquitous overexpression of various integrins on different types of tumors and their vasculature.
Colorectal cancer is the fourth most diagnosed malignancy worldwide and surgery is one of the cornerstones of the treatment strategy. Near-infrared (NIR) fluorescence imaging is a new and upcoming ...technique, which uses an NIR fluorescent agent combined with a specialised camera that can detect light in the NIR range. It aims for more precise surgery with improved oncological outcomes and a reduction in complications by improving discrimination between different structures.
A systematic search was conducted in the Embase, Medline and Cochrane databases with search terms corresponding to ‘fluorescence-guided surgery’, ‘colorectal surgery’, and ‘colorectal cancer’ to identify all relevant trials.
The following clinical applications of fluorescence guided surgery for colorectal cancer were identified and discussed: (1) tumour imaging, (2) sentinel lymph node imaging, (3) imaging of distant metastases, (4) imaging of vital structures, (5) imaging of perfusion. Both experimental and FDA/EMA approved fluorescent agents are debated. Furthermore, promising future modalities are discussed.
Fluorescence-guided surgery for colorectal cancer is a rapidly evolving field. The first studies show additional value of this technique regarding change in surgical management. Future trials should focus on patient related outcomes such as complication rates, disease free survival, and overall survival.
Near-infrared (NIR) fluorescence imaging using indocyanine green (ICG) has recently been introduced to improve the sentinel lymph node (SLN) procedure. Several optical tracers have been successfully ...tested. However, the optimal tracer formulation is still unknown. This study evaluates the performance of ICG-technetium-99m (99mTc)-nanocolloid in relation to 2 most commonly used ICG-based formulas during SLN biopsy in vulvar cancer.
Twelve women who planned to undergo SLN biopsy for stage I vulvar cancer were prospectively included. Sentinel lymph node mapping was performed using the dual-modality radioactive and NIR fluorescence tracer ICG-99mTc-nanocolloid. All patients underwent combined SLN localization using NIR fluorescence and the (current) gold standard using blue dye and radioactive guidance.
In all 12 patients, at least 1 SLN was detected during surgery. A total of 21 lymph nodes (median 2; range, 1-3) were resected. Median time between skin incision and first SLN detection was 8 (range, 1-22) minutes. All resected SLNs were both radioactive and fluorescent, although only 13 (62%) of 21 SLNs stained blue. Median brightness of exposed SLNs, expressed as signal-to-background ratio, was 5.4 (range, 1.8-11.8). Lymph node metastases were found in 3 patients.
Near-infrared fluorescence-guided SLN mapping is feasible and outperforms blue dye staining. Premixing ICG with 99mTc-nanocolloid provides real-time intraoperative imaging of the SN and seems to be the optimal tracer combination in terms of intraoperative detection rate of the SN (100%). Moreover, ICG-99mTc-nanocolloid allows the administration of a 5-times lower injected dose of ICG (compared with ICG and ICG absorbed to human serum albumin) and can be injected up to 20 hours before surgery.
Background: Fluorescence cholangiography using indocyanine green (ICG) can enhance orientation of bile duct anatomy during laparoscopic cholecystectomy. To ensure clear discrimination between bile ...ducts and liver, the fluorescence ratio between both should be sufficient. This ratio is influenced by the ICG dose and timing of fluorescence imaging. We first systematically identified all strategies for fluorescence cholangiography. Second, we aimed to optimize the dose of ICG and dosing time in a prospective clinical trial. Methods: PubMed was searched for clinical trials studying fluorescence cholangiography. Furthermore, 28 patients planned to undergo laparoscopic cholecystectomy were divided into 7 groups, receiving different intravenous doses (5 or 10 mg ICG) at different time points (0.5, 2, 4, 6, or 24 hours prior to surgery). Results: The systematic review revealed 27 trials including 1057 patients. The majority of studies used 2.5 mg administered within 1 hour before imaging. Imaging 3 to 24 hours after ICG administration was never studied. The clinical trial demonstrated that the highest bile duct-to-liver ratio was achieved 3 to 7 hours after administration of 5 mg and 5 to 25 hours after administration of 10 mg ICG. Up to 3 hours after administration of 5 mg and up to 5 hours after administration of 10 mg ICG, the liver was equally or more fluorescent than the cystic duct, resulting in a ratio ≤1.0. Conclusion: This study shows for the first time that the interval between ICG administration and intraoperative fluorescence cholangiography should be extended. Administering 5 mg ICG at least 3 hours before imaging is easy to implement in everyday clinical practice and results in bile duct-to-liver ratios >1.0.