NUK - logo
E-viri
Celotno besedilo
  • Intercomparison of a 10 mH ...
    Gulmez, Y.; Turhan, E.; Gulmez, G.; Eckardt, H.; Rather, P.

    1998 Conference on Precision Electromagnetic Measurements Digest (Cat. No.98CH36254), 1998
    Conference Proceeding

    Turkish and German Metrology Institutes (UME and PTB) performed an intercomparison of 10 mH inductance standards at 1 kHz. The measurements of the thermostatically-controlled transfer standard of PTB were taken by new Maxwell-Wien bridge with a Wagner arm made in UME. Relative difference of UME and PTB measurements on the same standard was found to be 5.10/sup -6/. Results have shown that this type of Maxwell-Wien bridge is capable of producing very precise measurements on standard inductors and promised further developments that will increase the accuracy of the system.