  • Clinical diagnostic perform...
    Yang, Shuo; Yang, Ruifeng; Zhang, Siyu; Liu, Di; Zhou, Jiansuo; Wang, Tiancheng; Cui, Liyan

    Journal of clinical laboratory analysis, September 2019, Letnik: 33, Številka: 7
    Journal Article

    Background Hepatitis C virus antibody (anti‐HCV) test had been approved as a preliminary screening test for HCV infection. Light‐initiated chemiluminescent assay (LiCA) was a homogenous method. We aimed to assess the clinical diagnostic performance of LiCA and compare it with that of chemiluminescence immunoassay (CLIA) which was widely used in clinical laboratories. Methods A total of 10 772 patients from the Peking University Third Hospital were enrolled. The serum samples were detected on the ChIVD LiCA500 and Abbott Architect i2000SR platforms. Recombinant immunoblot assay (RIBA) and HCV RNA assay were used for confirmation. Results The negative agreement rate between ChIVD LiCA anti‐HCV assay and Abbott Architect anti‐HCV assay was 99.91%, the positive agreement rate was 37.31%, the total agreement rate was 98.74%, and the kappa coefficient (κ) was 0.519. The sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of ChIVD LiCA anti‐HCV assay were 96.39%, 99.95%, 89.58%, and 99.97%, respectively, which were superior to those of Abbott Architect anti‐HCV assay (93.98%, 99.25%, 51.90%, and 99.95%, respectively). Conclusion ChIVD LiCA anti‐HCV assay was a highly sensitive, specific homogenous method with good diagnostic performance, and was applicable for the routine screening of HCV infection in clinical laboratories.