  • Improvement of the long-ter...
    Mumme, T; Gravius, S; Andereya, S; Marx, R; Wirtz, D C; Müller-Rath, R

    Archives of orthopaedic and trauma surgery, 08/2007, Letnik: 127, Številka: 6
    Journal Article

    One of the main causes of cup loosening in cemented total hip replacement is the lack of a hydrolysis stable compound between the hydrophobic bone cement and the hydrophilic acetabular bone stock. Thus, the long-term adhesive strength between bone and bone cement is decreased resulting in premature aseptic loosening. Accordingly, an amphiphilic bonding system was developed to prevent hydrolytic debonding in the interface bone-to-bone cement. Polyethylene cups were cemented in cadaver sheep hips. The hips were prepared with (n = 10) a multilayer bonding system as well as with an improved monolayer bonding system with (n = 10) and without (n = 10) UV-light hardening. The hips were dynamically loaded (10(6) cycles; 3 Hz; upper/lower load 1,500/200 N) and afterwards turned out with a defined strength to determine the interface compound strength. Dynamically loaded cemented cups without using the bonding system (n = 10) function as a control group. The interface adhesive strength improved significantly by a factor of 1.9 (using the multilayer bonding system) compared to the unconditioned ones (P < or = 0.05); for the monolayer bonding system without (with) UV-light hardening the compound strength was 11.9-fold (> or =22.2-fold) higher (P < or = 0.001). The bonding system significantly improves the adhesive strength between bone and bone cement in cemented cup arthroplasty.